National Association of Testing Authorities, Australia

SCOPE OF ACCREDITATION



Kingfisher International Pty Ltd
OPTICAL CALIBRATION LABORATORY
| Accreditation Number: 20533 | Site Number: 24605 |

Address Details:
720 Springvale Road
MULGRAVE, VIC 3170
AUSTRALIA
Website: https://www.kingfisherfiber.com
Contact Details:
Mr Bruce Robertson
+61(03) 85441750


Availability:Services available to external clients



Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.

 

ISO/IEC 17025 (2017)
Calibration

SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION/RANGE
Optical metrology - Optical measuring equipmentFibre optic systemsPower - Optical fibre; Wavelength; Direct comparison against a reference meter; Direct comparison against a reference spectrometer; IEC 61315 Ed. 2.0, Method 1, 2, 3, 4, 5, 6 and 7
CAPABILITY
with Calibration and Measurement Capability of-
Wavelength 
0.5 nm from 350 nm to 1700 nm
Power - Optical fibre
0.06 dB or 1.4% of reading whichever is greater from 350 nm to 1650 nm in steps of 5 nm + 22 to -60 dBm / 10 uW - 1 nW. Maximum power per IEC 60825 Class 1M / 2M
Return loss - Optical fibre
0.04 dB at 850 nm, 1310 nm, 1550 nm in the range from -60 dB to 0 dB
Fibre optic systems; Laser energy meters; Laser power meters; Photodiodes; Radiometers; LinearityDirect comparison against a reference meterIEC 61315 Ed. 2.0, Methods 1, 2. 3. 4, 6 and 7

CAPABILITY
Photodiodes and radiometers with Fibre Optic interface
with Calibration and Measurement Capability of -
0.01 dB per 10 dB or 0.2% of the reading whichever is greater from -70 dB to 10 dB at 1310 nm & 1550 nm, and from -70 dBm to 0 dBm at 650 nm & 850 nm, non-coherent broadband light,
0.1 dB per 10 dB or 0.7% of the reading whichever is greater from -60 dBm to 10 dBm at 850 nm, 1310 nm & 1550 nm, using coherent laser light.
ResponsivityDirect comparison against a reference meterMethods 1, 2, 6 and 7
CAPABILITY
Photodiodes and radiometers with Fibre Optic interface
with Calibration and Measurement Capability of-
0.06 dB or 1.4% of the reading whichever is greater from 350 nm to 1650 nm in 5 nm steps, -60 dBm to 3 dBm / 10 uW - 1 nW and wavelength accuracy of 0.5 nm, using non-coherent light. In accordance with TIA-455-231 / IEC 61315 / FOTP 231 Calibration of Fibre Optic Power Meters
| Accreditation Number: 20533 | Site Number: 24605 | Printed on : 28-Sep-2020

 

--------------------- END OF SCOPE ----------------------