PRINT SCOPE


National Association of Testing Authorities, Australia

SCOPE OF ACCREDITATION



Power System Support Pty Ltd
POWER SYSTEM SUPPORT PTY LTD
| Accreditation Number: 19724 | Site Number: 22965 |

Address Details:
Unit 4
380 Marion Street
CONDELL PARK, NSW 2200
AUSTRALIA
Website: www.powersystemsupport.com.au
Contact Details:
Mr Paul Edward Pokorny
+61 0422906007


Availability:Services available to external clients



Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

 

ISO/IEC 17025 (2005) - The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Calibration

SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION/RANGE
DC and low frequency electrical metrology - Electrical measurement and test equipmentInsulation resistance test equipmentResistanceComparison with a reference standard by substitution measurementBy the method of PSS WI-11At voltages of 50 V to 5 kV

CAPABILITY
with Calibration and Measurement Capability of;
3% from 1 MΩ - 1 GΩ;
5% from 1 GΩ - 10 GΩ;                    
10% from 10 GΩ - 100 GΩ;
AmmetersAC currentIndirect measurement against a reference standardby method of single or multiple turnsfrom 50 mA up to 5000 A at 50 Hz 
CAPABILITY
with Calibration and Measurement Capability of -;
1% of reading;
from 50 mA up to 5000 A at 50 Hz;
DC and low frequency electrical metrology - High-voltage/high-current standards and equipmentCurrent transformersCurrent error; Phase displacement; Comparison with a reference standard by differential measurementAS 1675;
AS/IEC 60044.1;
IEC 61869-1,-2

CAPABILITY
Accuracy tests at 50 Hz, up to 5000 A applied primary current;
for CTs with rated secondaries of 1 A or 5 A;
with Calibration and Measurement Capability of -;
±0.025% for current error;
±0.030 crad for phase displacement;
Voltage transformersPhase displacement; Voltage error; Comparison with a reference standard by differential measurementAS 1243;
AS/IEC 60044.2
AS/IEC 61869-1,-3;

CAPABILITY
Accuracy tests at 50 Hz, up to rated primary voltage of 132 kV/√3;
Single phase measurements of VTs with rated secondaries of 63.5 V or 110 V;with Calibration and Measurement Capability of -;
0.025% for voltage error;
0.025 crad for phase displacement;
High voltage test setsVoltageDirect measurement against a reference standard

Measurements performed in the laboratory and in the field ;
a.c. test sets
from 1 kVrms to 100 kVrms at 50 Hz;
from ±1 kVpeak/√2 to ±100 kVpeak/√2, at 50 Hz;

from ±1 kVpeak to ±140 kVpeak, at 50 Hz;

d.c. test sets
from - 90 kV to +90 kV

CAPABILITY
Measurements performed in the laboratory;
with Calibration and Measurement Capability of -;  

a.c. test sets

2% of reading from 1 kVrms to 100 kVrms at 50 Hz;
2% of reading from ±1 kVpeak/√2 to ±100 kVpeak/√2, at 50 Hz;

2% of reading from ±1 kVpeak to ±140 kVpeak, at 50 Hz;

d.c. test sets
2% of reading from -90 kV to 90 kV

Measurements performed in the field;

with Calibration and Measurement Capability of;

a.c. test sets
2.5% of reading from 1 kVrms to 100 kVrms at 50 Hz;
2.5% of reading from ±1 kVpeak/√2 to ±100 kVpeak/√2, at 50 Hz;
2.5% of reading from ±1 kVpeak to ±140 kVpeak, at 50 Hz;

d.c. test sets
2.5% of reading from -90 kV to 90 kV


ISO/IEC 17025 (2005)
Infrastructure and Asset Integrity

SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION/RANGE
Evaluation of high voltage/power equipment and components - PerformanceClothing materials, gloves and tools for use in high voltage/power applicationsAC voltage; Applied high voltage withstand; Electrical performance of cables and components; Workplace electrical safety; High voltage; Surface leakage;  ISSC14 Guide to electrical workers' safety equipment chapter 13;
 PSS WI-12
At 50 Hz up to 45 kV
       


High voltage operating sticks


Transformers and inductorsAccuracy testsPhase displacement; Voltage error; IEC and AS 60044.2;
AS1243:1982 Superseded;
IEC 61869-1; and IEC 61869-3; Instrument transformers - Voltage transformers
Accuracy tests at 50 Hz up to rated primary voltage of 132 kV/√3;
Single phase measurements of voltage transformers with rated secondaries of 63.5 V or 110 V 
Accuracy testsCurrent error; Phase displacement; IEC and AS 60044.1: Instrument Transformers-Current Transformers;
IEC 61869-1 and IEC 61869-1 -2: Instrument Transformers-Current Transformers;
AS 1675: 1986 Withdrawn Instrument Transformers-Measurement & Protection
Accuracy tests at 50 Hz up to 5000 A applied primary current
for current transformers with rated secondaries of 1 A or 5 A
Bushings; Cable feeders; Cables; Clothing materials, gloves and tools for use in high voltage/power applications; Insulated systems; Insulation products; Insulator and conductor fittings; Insulators; Three phase cable installations; Transformers and inductors; Insulation resistanceInsulation resistanceBy the method of PSS WI-10
Tests performed in the laboratory and in the field;
From 1 MΩ to 100 GΩ at voltages of 50 Vdc to 5,000 Vdc

| Accreditation Number: 19724 | Site Number: 22965 | Printed on : 18-Feb-2019

 

--------------------- END OF SCOPE ----------------------