SCOPE OF ACCREDITATION



ASG EXICAL Pty Ltd
ASG-X-TECHNOLOGIES
| Accreditation Number: 15464 | Site Number: 16363 |

Address Details:
2/70 Barrie Road
TULLAMARINE, VIC 3043
AUSTRALIA
Website: www.asgxtech.com.au
Contact Details:
Mr Dario Lodkowski
+61 1300880414


Availability:Services available to external clients



Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.

 

ISO/IEC 17025 (2005) - The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
Calibration

SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION/RANGE
DC and low frequency electrical metrology - Electrical instrument calibratorsInstrument calibratorsAC voltageComparison with a reference standard
CAPABILITY
0.15% + 34μV of reading at 0.01 mV to 10 mV and from 10 Hz to 20 kHz 
0.25% + 70μV of reading at 0.01 mV to 10 mV and from 20 kHz to 50 kHz 
1.1% + 110μV of reading at 0.01 mV to 10 mV and from 50 kHz to 100 kHz 
0.02% of reading from 0.01 V to 10 V and from 10 Hz to 20 kHz 
0.05% of reading from 0.01 V to 10 V and from 20 kHz to 50 kHz 
0.12% of reading from 0.01 V to 10.0 V and from 50 kHz to 100 kHz 
0.04% of reading from 10 V to 100 V at 10 Hz 
0.03% of reading from 10 V to 100 V and from 10 Hz to 20 kHz 
0.06% of reading from 10 V to 100 V and from 20 kHz to 50 kHz 
0.18% of reading from 10 V to 100 V and from 50 kHz to 100 kHz 
0.07% of reading from 100 V to 700 V at 40 Hz 
0.09% of reading from 100 V to 700 V and from 40 Hz to 10 kHz 
0.12 % of reading from 700 V to 1000 V and from 40 Hz to 10 kHz 
AC currentComparison with a reference standard
CAPABILITY
0.6% of reading from 10 μA to 10 A at 10 Hz 
0.3% of reading from 10 μA to 100 μA and from 10 Hz to 5 kHz 
1.6% of reading from 10 μA to 100 μA and from 5 kHz to 10 kHz 
0.1% of reading from 0.1 mA to 100 mA and from 10 Hz to 20 kHz 
0.7% of reading from 0.1 mA to 100 mA and from 20 kHz to 30 kHz 
0.2% of reading from 100 mA to 1 A and from 10 Hz to 5 kHz 
0.5% of reading from 100 mA to 1 A and from 5 kHz to 20 kHz 
1.6% of reading from 100 mA to 1 A and from 20 kHz to 30 kHz 
0.3% of reading from 1 A to 10 A and from 10 Hz to 5 kHz 
1.6% of reading from 1 A to 10 A and from 5 kHz to 10 kHz 
ResistanceComparison with a reference standard0 Ω to 100 MΩ

CAPABILITY
0.0033% of reading + 0.2 mΩ from 0 Ω to 10 Ω 
0.003% of reading + 2 mΩ from 10 Ω to 100 Ω 
0.002% of reading + 1.5 Ω from 100 Ω to 100 kΩ 
0.003% of reading + 24 Ω from 100 kΩ to 1 MΩ 
0.01% of reading + 730 Ω from 1 MΩ to 10 MΩ 
0.085% of reading + 0.06 MΩ from 10 MΩ to 100 MΩ 
DC currentComparison with a reference standard0 μA to 10 A

CAPABILITY
0.004% of reading + 0.002 μA from 0 μA to 100 μA 
0.003% of reading + 0.14 μA from 100 μA to 10 mA 
0.005% of reading + 1.9 μA from 10 mA to 100 mA 
0.014% of reading + 30 μA from 100 mA to 1 A 
0.14% of reading + 0.75 mA from 1 A to 3A 
0.18% of reading + 0.22 mA from 3 A to 10 A 
DC voltageComparison with a reference standard0 V to 1000 V

CAPABILITY
0.0017% of reading + 1.4 μV from 0 V to 0.1 V 
0.0011% of reading + 11 μV from 0.1 V to 1 V 
0.001% of reading + 50 μV from 1 V to 10 V 
0.0014% of reading + 0.8 mV from 10 V to 100 V 
0.002% of reading + 16 mV from 100 V to 1000 V 
DC and low frequency electrical metrology - Electrical measurement and test equipmentAmmetersDC currentComparison with a reference standard0 μA to 10 A

CAPABILITY
0.004% of reading + 0.002 μA from 0 μA to 100 μA 
0.003% of reading + 0.14 μA from 100 μA to 10 mA 
0.005% of reading + 1.9 μA from 10 mA to 100 mA 
0.014% of reading + 30 μA from 100 mA to 1 A 
0.14% of reading + 0.75 mA from 1 A to 3A 
0.18% of reading + 0.22 mA from 3 A to 10 A 
AC currentComparison with a reference standard
CAPABILITY
0.6% of reading from 10 μA to 10 A at 10 Hz 
0.3% of reading from 10 μA to 100 μA and from 10 Hz to 5 kHz 
1.6% of reading from 10 μA to 100 μA and from 5 kHz to 10 kHz 
0.1% of reading from 0.1 mA to 100 mA and from 10 Hz to 20 kHz 
0.7% of reading from 0.1 mA to 100 mA and from 20 kHz to 30 kHz 
0.2% of reading from 100 mA to 1 A and from 10 Hz to 5 kHz 
0.5% of reading from 100 mA to 1 A and from 5 kHz to 20 kHz 
1.6% of reading from 100 mA to 1 A and from 20 kHz to 30 kHz 
0.3% of reading from 1 A to 10 A and from 10 Hz to 5 kHz 
1.6% of reading from 1 A to 10 A and from 5 kHz to 10 kHz 
Ohm metersResistanceComparison with a reference standard0 Ω to 100 MΩ

CAPABILITY
0.0033% of reading + 0.2 mΩ from 0 Ω to 10 Ω 
0.003% of reading + 2 mΩ from 10 Ω to 100 Ω 
0.002% of reading + 1.5 Ω from 100 Ω to 100 kΩ 
0.003% of reading + 24 Ω from 100 kΩ to 1 MΩ 
0.01% of reading + 730 Ω from 1 MΩ to 10 MΩ 
0.085% of reading + 0.06 MΩ from 10 MΩ to 100 MΩ 
VoltmetersDC voltageComparison with a reference standard0 V to 1000 V

CAPABILITY
0.0017% of reading + 1.4 μV from 0 V to 0.1 V 
0.0011% of reading + 11 μV from 0.1 V to 1 V 
0.001% of reading + 50 μV from 1 V to 10 V 
0.0014% of reading + 0.8 mV from 10 V to 100 V 
0.002% of reading + 16 mV from 100 V to 1000 V 
AC voltageComparison with a reference standard
CAPABILITY
0.15% + 34μV of reading at 0.01 mV to 10 mV and from 10 Hz to 20 kHz 
0.25% + 70μV of reading at 0.01 mV to 10 mV and from 20 kHz to 50 kHz 
1.1% + 110μV of reading at 0.01 mV to 10 mV and from 50 kHz to 100 kHz 
0.02% of reading from 0.01 V to 10 V and from 10 Hz to 20 kHz 
0.05% of reading from 0.01 V to 10 V and from 20 kHz to 50 kHz 
0.12% of reading from 0.01 V to 10.0 V and from 50 kHz to 100 kHz 
0.04% of reading from 10 V to 100 V at 10 Hz 
0.03% of reading from 10 V to 100 V and from 10 Hz to 20 kHz 
0.06% of reading from 10 V to 100 V and from 20 kHz to 50 kHz 
0.18% of reading from 10 V to 100 V and from 50 kHz to 100 kHz 
0.07% of reading from 100 V to 700 V at 40 Hz 
0.09% of reading from 100 V to 700 V and from 40 Hz to 10 kHz 
0.12 % of reading from 700 V to 1000 V and from 40 Hz to 10 kHz 
DC and low frequency electrical metrology - Electrical standardsConductance boxes; Precision resistors; Resistance boxes; ResistanceComparison with a reference standard0 Ω to 100 MΩ

CAPABILITY
0.0033% of reading + 0.2 mΩ from 0 Ω to 10 Ω 
0.003% of reading + 2 mΩ from 10 Ω to 100 Ω 
0.002% of reading + 1.5 Ω from 100 Ω to 100 kΩ 
0.003% of reading + 24 Ω from 100 kΩ to 1 MΩ 
0.01% of reading + 730 Ω from 1 MΩ to 10 MΩ 
0.085% of reading + 0.06 MΩ from 10 MΩ to 100 MΩ 
Pressure metrology - Pressure and vacuum measuring equipmentDigital pressure gauges; Mechanical pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges; Gauge pressure; High vacuum; Comparison with reference instrument -95 to 120 MPa
Electrical transducers with dc voltage, dc current or resistance outputs

Including on site calibrations



CAPABILITY
Gauge pressure calibration
With Calibration and Measurement Capability of-
0.016% reading or 0.016 kPa (whichever is greater) in the range -95 to 100 kPa
0.016% reading in the range 100 kPa to 120 MPa

For on site calibrations
Calibration and Measurement Capability of-
0.13% reading from -95 kPa to 120 MPa



Pressure metrology - Pressure standardsPressure calibrators - Non-dead weightGauge pressureComparison with reference instrumentGauge pressure calibration in the range -95 to 120 MPa
On-site gauge pressure calibration (industrial grade only) in the range -95 kPa to 120 MPa

CAPABILITY
Gauge pressure calibration in the range -95 to 100 kPa
0.016% reading or 0.016 kPa (whichever is greater)  
Gauge pressure calibration in the range 100 kPa to 120 MPa
0.016% reading
On-site gauge pressure calibration
0.13% reading
Time and frequency metrology - Frequency, time and waveform measuring equipmentCountersFrequency1 Hz to  10 MHz 

CAPABILITY
0.045% of reading from 1 Hz to 40 Hz 
0.015% of reading from 40 Hz to 10 MHz 
TachometersRotational speed 60 rpm to 90,000 rpm

CAPABILITY
0.045% of reading from 60 rpm to 2400 rpm 
0.015% of reading from 2400 rpm to 90,000 rpm
StroboscopesFrequency 1 Hz  to 1.5 kHz 

CAPABILITY
0.045% of reading from 1 Hz to 40 Hz 
0.015% of reading from 40 Hz to 1.5 kHz 
Time interval metersTime interval100 ns to 1 s 

CAPABILITY
0.015 % of reading from 100 ns to 1 s 
| Accreditation Number: 15464 | Site Number: 16363 | Printed on : 28-Feb-2019

 

--------------------- END OF SCOPE ----------------------