NATA


Facilities and Labs

New Search |

Hinco Instruments Pty Ltd

Accreditation No: 546
www.hinco.com.au
 
Contact: Mr J W French
Phone: (08) 9455 6777
Fax: (08) 9455 7711
Email:
 
Address:
Unit 5/52 Vinnicombe Drive,
CANNING VALE
WA 6155
 
Facilities: Public testing service
Last Modified: 16-DEC-15


This laboratory complies with the requirements of ISO/IEC 17025 (2005)
The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise


1.20 Pressure and vacuum measuring devices
Calibration of gauge pressure (including on-site calibration)
with least uncertainties of measurement of:-
0.2 kPa from -95 to 0 kPa;
0.05% of reading or 0.3 kPa (whichever is greater) from 0 up to 4 MPa;
0.05% of reading or 3 kPa (whichever is greater) from 4 to 100 MPa
.01 Pressure gauges
By the method of AS 1349 or MSA Test Method 2
.02 Vacuum gauges (bourdon tube)
By the method of MSA Test Method 2
.11 Pressure transducers
Industrial grade, self indicating and electrical transducers with dc voltage, dc current
or resistance outputs
By the method of MSA Test Method 1
.13 Pressure recorders
By the method of MSA Test Method 1
.23 Digital manometers
By the method of MSA Test Method 1
with least uncertainties of measurement of
0.02 kPa from -20 kPa to 20 kPa
.31 Pressure control devices
Pressure switches
with least uncertainties of measurement of -
0.2 kPa from -95 to 0 kPa;
0.05% of reading or 0.3 kPa (whichever is greater) up to 4 MPa;
0.05% of reading or 3 kPa (whichever is greater) from 4 to 100 MPa

1.32 Resistors, resistance boxes and potential dividers
.01 Precision resistors, resistance boxes and conductance boxes
with least uncertainties of measurement of -
0.5% + 1 mΩ from 0.001 to 0.1 Ω
0.1% + 1 mΩ from 0.1 Ω to 10 kΩ
0.5% from 10 to 100 kΩ
2% from 100 kΩ to 10 MΩ

1.38 Instrument calibrators
Calibration of Process Calibrators only
.01 D.C. voltage
with least uncertainties of measurement of -
0.05% from 0.01 to 100 V
.11 D.C. current
with least uncertainties of measurement of -
0.05% from 0.1 to 100 mA
.51 Resistance
with least uncertainties of measurement of -
0.1% ± 1 mΩ from 0.1 Ω to 10 kΩ

1.39 Indicating and recording instruments
.01 D.C. voltmeters
with least uncertainties of measurement of -
0.05% from 0.01 up to 1 000 V
.02 A.C. voltmeters
with least uncertainties of measurement of -
0.1% from 0.5 to 700 V at 50 and 60 Hz
.03 D.C. ammeters
with least uncertainties of measurement of -
0.05% from 0.1 mA to 20 A
0.9% from 20 A to 1000 A for Current Clamps [using multiple-turn technique]
.04 A.C. ammeters
with least uncertainties of measurement of -
0.1% from 10 mA to 20 A at 50 and 60 Hz
0.5% from 20 A to 1000 A for Toroidal Current Clamps [using multiple-turn technique]
1.0% from 20 A to 1000 A for Non-Toroidal Current Clamps [using multiple-turn technique]
.09 Ohmmeters
with least uncertainties of measurement of -
0.5% from 1Ω to 100 kΩ
2% from 100 kΩ to 10 MΩ
5% from 10 to 100 MΩ
10% from 100 MΩ to 1 GΩ


1.80 Calibration of temperature measuring equipment
.41 Digital temperature indicator systems
with least uncertainties of measurement of -
0.5 °C from -50 to 0 °C
0.2 °C from 0 to 150 °C
0.3 °C from 150 to 200 °C

1.81 Calibration of ancillary temperature measuring instruments
.04 Indicators, recorders and controllers
Calibration of temperature calibrator instruments
by electrical simulation [measure and source mode]
with least uncertainties of measurement of -
0.8 °C + 0.1% of Temperature for Type K thermocouples from -100 °C to +1100 °C
0.8 °C + 0.1% of Temperature for Type T thermocouples from -100 °C to +390 °C
0.8 °C + 0.1% of Temperature for Type N thermocouples from +100 °C to +1200 °C
0.3 °C for RTD PT100 from -100 °C to +400 °C
Analogue chart recorders
with least uncertainties of measurement of -
0.5 °C from -50 °C to 200 °C

1.84 Testing of controlled enclosures
.01 Ovens and Furnaces
with least uncertainties of measurement of -
1 °C from 0 to 200 °C
2.5 °C from 200 to 1100 °C
.05 Dry block calibrators
with the least uncertainty of measurement of -
0.5 ºC from 0 to 500 ºC
.15 Medical refrigeration equipment
as per AS 3864.2 section 3.7 - Annual tasks
with least uncertainties of measurement of -
0.8 °C from -50 °C to 0 °C
0.3 °C from 0 °C to 10 °C
AS 3864.2 section 3.8.3 - Spatial temperature distribution check
with least uncertainties of measurement of -
0.8 °C from -50 °C to 10 °C