Facilities: Public testing service
Last Modified: 27-MAR-17


This laboratory complies with the requirements of ISO/IEC 17025 (2005)
The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise


1.32 Resistors, resistance boxes and potential dividers
.01 Precision resistors, resistance boxes and conductance boxes
with least uncertainties of measurement of -
0.01% reading + 10 μΩ from 1 mΩ to 1 MΩ
0.1% reading from 1 MΩ to 100 MΩ

1.36 Voltage standards
.01 Standard cells
with least uncertainties of measurement of -
5 μV at 1V, 1.018 V and 10 V
.11 Electronic e.m.f. reference devices
with least uncertainties of measurement of -
5 μV/V in the range 1 V to 10 V

1.37 Precision transfer instruments
.01 A.C./D.C. transfer instruments
with least uncertainties of measurement of -
0.02% reading from 0.25 V to 300 V at 50 Hz to 20 kHz
0.02% reading above 300 V to 1 000 V at 50 Hz to 200 Hz
0.04% reading above 300 V to 1 000 V at 200 Hz to 20 kHz
0.04% reading above 0.25 V to 300 V at 20 kHz to 50 kHz
0.08% reading above 300 V to 1 000 V at 20 kHz to 50 kHz
0.2% reading above 0.25 V to 10 V at 10 Hz
0.08% reading above 0.25 V to 1 000 V at 20 Hz to 50 Hz
0.08% reading above 0.25 V to 50 V at 50 kHz to 100 kHz
0.2% reading above 50 V to 300 V at 50 kHz to 100 kHz
0.4% reading above 300 V to 1 000 V at 50 kHz to 100 kHz
0.2% reading above 0.25 V to 50 V at 100 kHz to 1 MHz
0.1% reading from 2.5 mA to 20 A at 50 Hz to 1 kHz

1.38 Instrument calibrators
.01 D.C. voltage
with least uncertainties of measurement of -
0.7 μV at 0 V
10 μV/V or 1.5 μV whichever is the greater in the range 1 mV to 1.1 kV
.02 A.C. voltage
with least uncertainties of measurement of -
0.05% reading + 20 μV from 0.01 V to 0.25 V at 50 Hz to 20 kHz
0.08% reading + 30 μV from 0.01 V to 0.25 V at 20 kHz to 50 kHz
0.1% reading + 30 μV from 0.01 t V to 0.25 V at 50 kHz to 100 kHz
0.02% reading from 0.25 V to 300 V at 50 Hz to 20 kHz
0.02% reading above 300 V to 1 000 V at 50 Hz to 200 Hz
0.04% reading above 300 V to 1 000 V at 200 Hz to 20 kHz
0.04% reading above 0.25 V to 300 V at 20 kHz to 50 kHz
0.08% reading above 300 V to 1 000 V at 20 kHz to 50 kHz
0.2% reading above 0.25 V to 10 V at 10 Hz
0.08% reading above 0.25 V to 1 000 V at 20 Hz to 50 Hz
0.08% reading above 0.25 V to 50 V at 50 kHz to 100 kHz
0.2% reading above 50 V to 300 V at 50 kHz to 100 kHz
0.4% reading above 300 V to 1 000 V at 50 kHz to 100 kHz
0.2% reading above 0.25 V to 50 V at 100 kHz to 1 MHz
0.1% reading from 2.5 mA to 20 A at 50 Hz to 1 kHz
.11 D.C. current
with least uncertainties of measurement of -
1.3 nA at 0 A
0.02% reading + 0.02 μA up to 2 A
0.02% reading from 2 A to 20 A
.12 A.C. current
with least uncertainties of measurement of -
0.1% reading above from 2.5 mA to 20 A at 50 Hz to 1kHz
.51 Resistance
with least uncertainties of measurement of -
27 μΩ at 0 Ω
0.01% reading + 10 μΩ from 1 mΩ to 1 MΩ
0.1% reading from 1 MΩ to 100 MΩ

1.39 Indicating and recording instruments
.01 D.C. voltmeters
with least uncertainties of measurement of -
0.7 μV at 0 V
10 μV/V or 1.5 μV whichever is the greater in the range 1 mV to 1.1 kV
.02 A.C. voltmeters
with least uncertainties of measurement of -
0.05% reading + 20 μV from 0.01 V to 0.25 V at 50 Hz to 20 kHz
0.08% reading + 30 μV from 0.01 V to 0.25 V at 20 kHz to 50 kHz
0.1% reading + 30 μV from 0.01 V to 0.25 V at 50 kHz to 100 kHz
0.02% reading from 0.25 V to 300 V at 50 Hz to 20 kHz
0.02% reading above 300 V to 1 000 V at 50 Hz to 200 Hz
0.04% reading above 300 V to 1 000 V at 200 Hz to 20 kHz
0.04% reading above 0.25 V to 300 V at 20 kHz to 50 kHz
0.08% reading above 300 V to 1 000 V at 20 kHz to 50 kHz
0.2% reading above 0.25 V to 10 V at 10 Hz
0.08% reading above 0.25 V to 1 000 V at 20 Hz to 50 Hz
0.08% reading above 0.25 V to 50 V at 50 kHz to 100 kHz
0.2% reading above 50 V to 300 V at 50 kHz to 100 kHz
0.4% reading above 300 V to 1 000 V at 50 kHz to 100 kHz
0.2% reading above 0.25 V to 50 V at 100 kHz to 1 MHz
0.1% reading from 2.5 mA to 20 A at 50 Hz to 1 kHz
.03 D.C. ammeters
with least uncertainties of measurement of -
1.3 nA at 0 A
0.02% reading + 0.02 μA up to 2 A
0.02% reading from 2 A to 20 A
1% from 20 A up to 1 000 A for clamp-on ammeters using multiple turns
.04 A.C. ammeters
with least uncertainties of measurement of -
0.1% reading above from 2.5 mA to 20 A at 50 Hz to 1kHz
1% reading above 20 A to 900 A at 50 Hz for clamp-on ammeters using multiple turns
.09 Ohmmeters
with least uncertainties of measurement of -
27 μΩ at 0 Ω
0.01% reading + 10 μΩ from 10 mΩ to 1 MΩ
0.1% reading above 1 MΩ to 100 MΩ
.81 Graphic recording instruments
D.C.voltage
with least uncertainties of measurement of -
0.7 μV at 0 V
10 μV/V or 1.5 μV whichever is the greater in the range 1 mV to 1.1 kV

A.C. voltage
with least uncertainties of measurement of -
0.05% reading + 20 μV from 0.01 V to 0.25 V at 50 Hz to 20 kHz
0.08% reading + 30 μV from 0.01 V to 0.25 V at 20 kHz to 50 kHz
0.1% reading + 30 μV from 0.01 V to 0.25 V at 50 kHz to 100 kHz
0.02% reading from 0.25 V to 300 V at 50 Hz to 20 kHz
0.02% reading above 300 V to 1 000 V at 50 Hz to 200 Hz
0.04% reading above 300 V to 1 000 V at 200 Hz to 20 kHz
0.04% reading above 0.25 V to 300 V at 20 kHz to 50 kHz
0.08% reading above 300 V to 1 000 V at 20 kHz to 50 kHz
0.2% reading above 0.25 V to 10 V at 10 Hz
0.08% reading above 0.25 V to 1 000 V at 20 Hz to 50 Hz
0.08% reading above 0.25 V to 50 V at 50 kHz to 100 kHz
0.2% reading above 50 V to 300 V at 50 kHz to 100 kHz
0.4% reading above 300 V to 1 000 V at 50 kHz to 100 kHz
0.2% reading above 0.25 V to 50 V at 100 kHz to 1 MHz
0.1% reading from 2.5 mA to 20 A at 50 Hz to 1 kHz

D.C. current
with least uncertainties of measurement of -
1.3 nA at 0 A
0.02% reading + 0.02 μA up to 2 A
0.02% reading from 2 A to 20 A

A.C. current
with least uncertainties of measurement of -
0.1% reading above from 2.5 mA to 20 A at 50 Hz to 1kHz

Ohmmeters
with least uncertainties of measurement of -
27 μΩ at 0 Ω
0.01% reading + 10 μΩ from 10 mΩ to 1 MΩ
0.1% reading above 1 MΩ to 100 MΩ

1.41 Frequency and time measuring instruments and standards
.11 Counters
with least uncertainties of measurement of -
3 parts in 109 (time base only)
.14 Stroboscopes
In the range 2 pulses to 1000 pulses per sec
with least uncertainties of measurement of -
0.1% reading
.15 Tachometers
Hand held tachometers in the range 100 r/min to 15000 r/min
with least uncertainties of measurement of -
0.1% reading
Optically coupled tachometers in the range 1000 r/min to 60,000 r/min
with least uncertainties of measurement of -
0.1% reading

1.42 Frequency analysers and waveform measuring instruments
.01 Frequency characteristics
Timebase measurements
with least uncertainties of measurement of -
3 parts in 109
.02 Input characteristics
Amplitude Measurements
D.C.voltage
with least uncertainties of measurement of -
0.7 μV at 0 V
10 μV/V or 1.5 μV whichever is the greater in the range 1 mV to 1.1 kV

A.C. voltage
with least uncertainties of measurement of -
0.05% reading + 20 μV from 0.01 V to 0.25 V at 50 Hz to 20 kHz
0.08% reading + 30 μV from 0.01 V to 0.25 V at 20 kHz to 50 kHz
0.1% reading + 30 μV from 0.01 V to 0.25 V at 50 kHz to 100 kHz
0.02% reading from 0.25 V to 300 V at 50 Hz to 20 kHz
0.02% reading above 300 V to 1 000 V at 50 Hz to 200 Hz
0.04% reading above 300 V to 1 000 V at 200 Hz to 20 kHz
0.04% reading above 0.25 V to 300 V at 20 kHz to 50 kHz
0.08% reading above 300 V to 1 000 V at 20 kHz to 50 kHz
0.2% reading above 0.25 V to 10 V at 10 Hz
0.08% reading above 0.25 V to 1 000 V at 20 Hz to 50 Hz
0.08% reading above 0.25 V to 50 V at 50 kHz to 100 kHz
0.2% reading above 50 V to 300 V at 50 kHz to 100 kHz
0.4% reading above 300 V to 1 000 V at 50 kHz to 100 kHz
0.2% reading above 0.25 V to 50 V at 100 kHz to 1 MHz
0.1% reading from 2.5 mA to 20 A at 50 Hz to 1 kHz

1.81 Calibration of ancillary temperature measuring instruments
.04 Indicators, recorders and controllers
Including portable temperature measuring instruments by electrical simulation
with least uncertainties of measurements of -

Type K Thermocouple
0.10 ºC from -200 ºC to 0 ºC
0.04 ºC from 0 ºC to 1000 ºC
0.05 ºC from 1000 ºC to 1372 ºC

Type T Thermocouple
0.07 ºC from -200 ºC to 0 ºC
0.04 ºC from 0 ºC to 400 ºC

Type B Thermocouple
0.12 ºC from 800 ºC to 1820 ºC

Type E Thermocouple
0.07 ºC from -200 ºC to 0 ºC
0.04 ºC from 0 ºC to 1000 ºC

Type J Thermocouple
0.08 ºC from -200 ºC to 0 ºC
0.04 ºC from 0 ºC to 1200 ºC

Type N Thermocouple
0.07 ºC from -100 ºC to 0 ºC
0.06 ºC from 0 ºC to 800 ºC
0.04 ºC from 800 ºC to 1300 ºC

Type R Thermocouple
0.10 ºC from 400 ºC to 1768 ºC

Type S Thermocouple
0.11 ºC from 300 ºC to 1768 ºC

Cold Junction
0.13 ºC at 0 ºC
0.11 ºC at 21 ºC

RTD Indicators
0.008 ºC from -200 ºC to 800 ºC