This laboratory complies with the requirements of ISO/IEC 17025 (2005)
The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise


1.11 Masses
.01 Mass standards
with least uncertainties of measurement of -
15 μg + 1 in 105 from 1 mg to 2 g
35 μg + 1 in 106 from 5 to 50 g
1 in 106 from 100 g to 1000 g
3 mg above 1000 g to 2000 g
9 mg above 2000 g to 5000 g
60 mg above 5000 g to 10000 g
.02 Industrial mass standards
with least uncertainties of measurement of -
13 mg from 500 g to 5 kg
0.16 g above 5 kg to 10 kg
0.22 g above 10 kg to 20 kg
0.3 g above 20 kg to 30 kg
3.0 g above 30 kg to 100 kg
8.0 g above 100 kg to 500 kg
16.0 g above 500 kg to 1000 kg
.03 Determination of mass
with least uncertainties of measurement of -
as under classes 1.11.01 and 1.11.02

1.12 Weighing devices
.01 Precision laboratory balances
with least uncertainties of measurement of -
2 in 106 or 20 μg (whichever is greater) up to 300 g
.02 Industrial balances
with least uncertainties of measurement of -
1 in 105 or 1 mg (whichever is greater) up to 60 kg
.03 Industrial weighing appliances
Calibration up to 3 t including the examination of Class III and Class IV weighing instruments to the procedures and requirements of the NMI NITP 6.1-6.4 National Instrument Test Procedures

1.13 Volumetric equipment
.03 Industrial volumetric proving measures
with least uncertainties of measurement of -
0.013% from 1 mL to 500 L
.15 Piston Operated Volumetric Apparatus, pipettes (POVAS)
POVA calibration from 10 μL to 10 mL
By the methods of AS 2162.2
with least uncertainties of measurement of -
0.07 μL at 10 μL
0.08 μL at 20 μL
0.16 μL at 100 μL
0.32 μL at 200 μL
1.6 μL at 1 mL
8.2 μL at 5 mL
16.4 μL at 10mL

1.26 Testing machines
.99 Other testing machines
Rotational speed of centrifuges
with least uncertainties of measurement of -
0.1% of reading or 2 rpm (whichever is greater)
from 10 rpm to 99900 rpm

1.39 Indicating and recording instruments
.01 D.C. voltmeters
0.007% + 3 μV up to 0.3 V
0.006% + 5 μV from 0.3 V to 3.3 V
0.006% + 50 μV from 3.3 V to 33 V
0.006% + 500 μV from 33 V to 330 V
0.007% + 1500 μV from 330 V to 1020 V
.02 A.C. voltmeters
0.36% + 20 μV up to 33 mV and from 10 to 45 Hz
0.16% + 20 μV up to 33 mV and from 45 Hz to 10 kHz
0.21% + 20 μV up to 33 mV and from 10 to 20 kHz
0.27% + 20 μV up to 33 mV and from 20 to 50 kHz
0.37% + 33 μV up to 33 mV and from 50 to 100 kHz
1.0% + 60 μV up to 33 mV and from 100 to 500 kHz

0.26% + 50 μV from 33 mV to 330 mV and 10 to 45 Hz
0.06% + 20 μV from 33 mV to 330 mV and 45 Hz to 10 kHz
0.11% + 20 μV from 33 mV to 330 mV and 10 to 20 kHz
0.17% + 40 μV from 33 mV to 330 mV and 20 to 50 kHz
0.25% + 170 μV from 33 mV to 330 mV and 50 to 100 kHz
0.72% + 330 μV from 33 mV to 330 mV and 100 to 500 kHz

0.16% + 250 μV from 0.33 V to 3.3 V and 10 to 45 Hz
0.04% + 60 μV from 0.33 V to 3.3 V and 45 Hz to 10 kHz
0.09% + 60 μV from 0.33 V to 3.3 V and 10 to 20 kHz
0.15% + 300 μV from 0.33 V to 3.3 V and 20 to 50 kHz
0.25% + 1700 μV from 0.33 V to 3.3 V and 50 to 100 kHz
0.52% + 3300 μV from 0.33 V to 3.3 V and 100 to 500 kHz

0.16% + 2.5 mV from 3.3 V to 33 V and 10 to 45 Hz
0.05% + 0.6 mV from 3.3 V to 33 V and 45 Hz to 10 kHz
0.09% + 2.6 mV from 3.3 V to 33 V and 10 to 20 kHz
0.20% + 5 mV from 3.3 V to 33 V and 20 to 50 kHz
0.25% + 17 mV from 3.3 V to 33 V and 50 to 100 kHz

0.06% + 6.6 mV from 33 V to 330 V and 45 Hz to 1 kHz
0.09% + 15 mV from 33 V to 330 V and 1 to 10 kHz
0.10% + 33 mV from 33 V to 330 V and 10 to 20 kHz

0.06% + 80 mV from 330 V to 1020 V and 45 Hz to 1 kHz
0.21% + 100 mV from 330 V to 1020 V and 1 to 5 kHz
0.21% + 500 mV from 330 V to 1020 V and 5 to 10 kHz
.03 D.C. ammeters
0.02% + 0.05 μA up to 3.3 mA
0.02% + 0.25 μA from 3.3 mA to 33 mA
0.02% + 3.3 μA from 33 mA to 330 mA
0.04% + 44 μA from 0.33 A to 2.2 A
0.07% + 330 μA from 2.2 A to 11 A
.04 A.C. ammeters
0.26% + 0.15 μA from 29 μA to 330 μA and 10 Hz to 20 Hz
0.13% + 0.15 μA from 29 μA to 330 μA and 20 Hz to 45 Hz
0.13% + 0.25 μA from 29 μA to 330 μA and 45 Hz to 1 kHz
0.45% + 0.15 μA from 29 μA to 330 μA and 1 kHz to 5 kHz
1.29% + 0.15 μA from 29 μA to 330 μA and 5 kHz to 10 kHz

0.21% + 0.3 μA from 0.33 to 3.3 mA and 10 Hz to 20 Hz
0.11% + 0.3 μA from 0.33 to 3.3 mA and 20 Hz to 45 Hz
0.11% + 0.3 μA from 0.33 to 3.3 mA and 45 Hz to 1 kHz
0.22% + 0.3 μA from 0.33 to 3.3 mA and 1 kHz to 5 kHz
0.62% + 0.3 μA from 0.33 to 3.3 mA and 5 kHz to 10 kHz

0.21% + 3 μA from 3.3 to 33 mA and 10 Hz to 20 Hz
0.11% + 3 μA from 3.3 to 33 mA and 20 Hz to 45 Hz
0.10% + 3 μA from 3.3 to 33 mA and 45 Hz to 1 kHz
0.22% + 3 μA from 3.3 to 33 mA and 1 kHz to 5 kHz
0.62% + 3 μA from 3.3 to 33 mA and 5 kHz to 10 kHz

0.21% + 30 μA from 33 to 330 mA and 10 Hz to 20 Hz
0.11% + 30 μA from 33 to 330 mA and 20 Hz to 45 Hz
0.10% + 30 μA from 33 to 330 mA and 45 Hz to 1 kHz
0.22% + 30 μA from 33 to 330 mA and 1 kHz to 5 kHz
0.62% + 30 μA from 33 to 330 mA and 5 kHz to 10 kHz

0.21% + 300 μA from 0.33 to 2.2 A and 20 Hz to 45 Hz
0.11% + 300 μA from 0.33 to 2.2 A and 45 Hz to 1 kHz
0.77% + 300 μA from 0.33 to 2.2 A and 1 kHz to 5 kHz

0.07% + 2000 μA from 2.2 to 11A and 45Hz to 65Hz
0.11% + 2000 μA from 2.2 to 11A and 65Hz to 500Hz
0.34% + 2000 μA from 2.2 to 11A and 500Hz to 1kHz
.09 Ohmmeters
0.016% + 0.008 Ω from 0 to 11 Ω
0.014% + 0.015 Ω from 11 to 33 Ω
0.010% + 0.015 Ω from 33 to 110 Ω
0.011% + 0.015 Ω from 110 to 330 Ω
0.010% + 0.06 Ω from 330 Ω to 1.1 kΩ
0.010% + 0.06 Ω from 1.1 kΩ to 3.3 kΩ
0.010% + 0.6 Ω from 3.3 kΩ to 11 kΩ
0.011% + 0.6 Ω from 11 to 33 kΩ
0.012% + 6Ω from 33 to 110 kΩ
0.014% + 6 Ω from 110 to 330 kΩ
0.016% + 55 Ω from 330 kΩ to 1.1 MΩ
0.020% + 55 Ω from 1.1 MΩ to 3.3 MΩ
0.063% + 550 Ω from 3.3 to 11 MΩ
0.10% + 550 Ω from 11 to 33 MΩ
0.51% + 5500 Ω from 33 to 110 MΩ
0.51% + 16500 Ω from 110 to 330 MΩ
.81 Graphic recording instruments
with least uncertainties of measurement of -
As listed in 1.39.01, 1.39.02, 1.39.03, 1.39.04 and 1.39.09
.82 Digital storage recorders
with least uncertainties of measurement of -
As listed in 1.39.01, 1.39.02, 1.39.03, 1.39.04 and 1.39.09


1.41 Frequency and time measuring instruments and standards
.01 Frequency meters
0.006% + 1 mHz from 0.01 to 1200 Hz
0.007% + 1 mHz from 1.2 to 10 kHz
0.006% + 15 mHz from 10 to 120 kHz
.11 Counters
with least uncertainties of measurement of -
As listed in 1.41.01
.15 Tachometers
Optical Tachometers with least uncertainties of measurement of -
0.050% from 10 rpm to 1000 rpm
0.030% from 1000 rpm to 7200 rpm
0.024% from 7200 rpm to 72000 rpm
0.041% from 7200 rpm to 99900 rpm