NATA


Facilities and Labs

New Search |

Australian Pressure Laboratory Pty Ltd

Accreditation No: 9727
www.auspressurelab.com.au
 
Contact: Mr R A Anderson
Mobile: 0408 312 699
Phone: (03) 9431 3658
Fax: (03) 9431 3668
Email:
 
Address:
Unit 9, 39 Susan Street,
ELTHAM
VIC 3095
 
Facilities: Public testing service
Last Modified: 18-AUG-17


This laboratory complies with the requirements of ISO/IEC 17025 (2005)
The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise


1.19 Barometers
.01 Aneroid barometers
Absolute pressure calibration up to 120 MPa
with least uncertainties of measurement of -
0.003% reading or 5 Pa (whichever is the greater) from 0.5 kPa to 2500 kPa
0.006% reading from 2.5 MPa to 120 MPa
.04 Gauge barometers
.11 Altimeters
Including the calibration of Air Data Test Sets (ADTS) pitot and static ranges to the methods of MSA Test Method 1 - 2008 and based on pressure to altitude conversion as per ISO 2533:1997 with uncertainty of measurement of;
0.003% reading or 5 Pa (whichever is the greater) from 2.7 kPa to 375 kPa absolute
Equivalent feet altitude range of up to 80478 feet and equivalent uncertainty of measurement in feet altitude as per the pressure capability as converted

Including the calibration of Air Data Test Sets (ADTS) airspeed ranges to the methods of MSA Test Method 1 2008 and based on and pressure to airspeed as per NASA Technical Note D-822 with uncertainty of measurement of;
0.004% of reading or 0.5 Pa (whichever is the greater) from 0 kPa to 375 kPa gauge
Equivalent Air Speed range of 0 to 1195 knots and equivalent uncertainty in air speed as per the pressure capability as converted
.99 Other barometers

1.20 Pressure and vacuum measuring devices
Gauge pressure calibration in the range -100 kPa to 280 MPa
with least uncertainties of measurement of -

0.004% of reading or 0.5 Pa (whichever is the greater) from -100 kPa to -5 kPa
0.3 Pa from -5 kPa to -0.0015 kPa
0.3 Pa from 0.0015 kPa to 5 kPa
0.004% of reading or 0.5 Pa (whichever is the greater) from 5 kPa to 20 kPa
0.003% of reading from 20 kPa to 2500 kPa
0.005% of reading from 2.5 MPa to 80 MPa
0.006% of reading from 80 MPa to 140 MPa
0.009% of reading from 140 MPa to 175 MPa
0.015% of reading from 175 MPa to 280 MPa

Absolute pressure calibration up to 120 MPa
with least uncertainties of measurement of -
0.003% of reading or 5 Pa (whichever is the greater) from 0.5 kPa to 2500 kPa
0.006% of reading from 2.5 MPa to 120 MPa

.01 Pressure gauges
Including test gauges to the methods of -
AS 1349 and MSA Test Method 2 - 2008
.02 Vacuum gauges (bourdon tube)
Including test gauges to the methods of -
AS 1349 and MSA Test Method 2 - 2008
.11 Pressure transducers
Self indicating and electrical transducers with dc voltage
or dc current output
to the methods of MSA Test Method 1 - 2008
.12 Pressure calibrators
.13 Pressure recorders
to the methods of MSA Test Method 1 - 2008
.22 Other liquid manometers
.23 Digital manometers

1.21 Pressure balances
Gauge pressure calibration in the range -100 kPa to 280 MPa
with least uncertainties of measurement of -

0.004% of reading or 0.5 Pa (whichever is the greater) from -100 kPa to -5 kPa
0.3 Pa from -5 kPa to -0.0015 kPa
0.3 Pa from 0.0015 kPa to 5 kPa
0.004% of reading or 0.5 Pa (whichever is the greater) from 5 kPa to 20 kPa
0.003% of reading from 20 kPa to 2500 kPa
0.005% of reading from 2.5 MPa to 80 MPa
0.006% of reading from 80 MPa to 140 MPa
0.009% of reading from 140 MPa to 175 MPa
0.015% of reading from 175 MPa to 280 MPa

.01 Air operated piston gauges
Calibration up to 8 MPa
.02 Oil operated piston gauges
Calibration up to 280 MPa
with a base least uncertainty of
measurement of - 0.05 kPa
.03 Pressure gauge calibrators
.04 Determination of mass for dead weight testers
with least uncertainties of measurement of
6 in 106or 60 μg (whichever is greater) up to 12 kg

1.38 Instrument calibrators
.01 D.C. voltage
0.004% of reading or 0.004 mV (which ever is the greater) from 0 V to 100 V
.11 D.C. current
0.0075% of reading or 0.0006 mA (which ever is the greater) from 0 mA to 100 mA

1.39 Indicating and recording instruments
.01 D.C. voltmeters
With least uncertainties of measurement of -
0.004% of reading or 0.004 mV (which ever is the greater) from 0 V to 100 V
.03 D.C. ammeters
With least uncertainties of measurement of -
0.0075% of reading or 0.0006 mA (which ever is the greater) from 0 mA to 100 mA

13.65 Pipes, hoses, valves and fittings
.11 Hydrostatic pressure tests
Operating pressure tests on pressure relief valves and pressure switches in the range 0 to 70,000 kPa
to section 7.5 Pressure control devices of Q1 Laboratory Quality Manual
.12 Pneumatic pressure tests
Operating pressure tests on pressure relief valves and pressure switches in the range 0 to 70,000 kPa
to section 7.5 Pressure control devices of Q1 Laboratory Quality Manual