NATA


Facilities and Labs

New Search |

Australian Pressure Laboratory Pty Ltd

Accreditation No: 9727
www.auspressurelab.com.au
 
Contact: Mr R A Anderson
Mobile: 0408 312 699
Phone: (03) 9431 3658
Fax: (03) 9431 3668
Email:
 
Address:
Unit 9, 39 Susan Street,
ELTHAM
VIC 3095
 
Facilities: Public testing service
Last Modified: 28-JUN-17


This laboratory complies with the requirements of ISO/IEC 17025 (2005)
The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise


1.19 Barometers
.01 Aneroid barometers
Absolute pressure calibration up to 120 MPa
with least uncertainties of measurement of -
0.003% reading or 5 Pa (whichever is the greater) from 0.5 kPa to 2500 kPa
0.006% reading from 2.5 MPa to 120 MPa
.04 Gauge barometers
.99 Other barometers

1.20 Pressure and vacuum measuring devices
Gauge pressure calibration in the range -100 kPa to 280 MPa
with least uncertainties of measurement of -

0.004% of reading or 0.5 Pa (whichever is the greater) from -100 kPa to -5 kPa
0.3 Pa from -5 kPa to -0.0015 kPa
0.3 Pa from 0.0015 kPa to 5 kPa
0.004% of reading or 0.5 Pa (whichever is the greater) from 5 kPa to 20 kPa
0.003% of reading from 20 kPa to 2500 kPa
0.005% of reading from 2.5 MPa to 80 MPa
0.006% of reading from 80 MPa to 140 MPa
0.009% of reading from 140 MPa to 175 MPa
0.015% of reading from 175 MPa to 280 MPa

Absolute pressure calibration up to 120 MPa
with least uncertainties of measurement of -
0.003% of reading or 5 Pa (whichever is the greater) from 0.5 kPa to 2500 kPa
0.006% of reading from 2.5 MPa to 120 MPa

.01 Pressure gauges
Including test gauges to the methods of -
AS 1349 and MSA Test Method 2 - 2008
.02 Vacuum gauges (bourdon tube)
Including test gauges to the methods of -
AS 1349 and MSA Test Method 2 - 2008
.11 Pressure transducers
Self indicating and electrical transducers with dc voltage
or dc current output
to the methods of MSA Test Method 1 - 2008
.12 Pressure calibrators
.13 Pressure recorders
to the methods of MSA Test Method 1 - 2008
.22 Other liquid manometers
.23 Digital manometers

1.21 Pressure balances
Gauge pressure calibration in the range -100 kPa to 280 MPa
with least uncertainties of measurement of -

0.004% of reading or 0.5 Pa (whichever is the greater) from -100 kPa to -5 kPa
0.3 Pa from -5 kPa to -0.0015 kPa
0.3 Pa from 0.0015 kPa to 5 kPa
0.004% of reading or 0.5 Pa (whichever is the greater) from 5 kPa to 20 kPa
0.003% of reading from 20 kPa to 2500 kPa
0.005% of reading from 2.5 MPa to 80 MPa
0.006% of reading from 80 MPa to 140 MPa
0.009% of reading from 140 MPa to 175 MPa
0.015% of reading from 175 MPa to 280 MPa

.01 Air operated piston gauges
Calibration up to 8 MPa
.02 Oil operated piston gauges
Calibration up to 280 MPa
with a base least uncertainty of
measurement of - 0.05 kPa
.03 Pressure gauge calibrators
.04 Determination of mass for dead weight testers
with least uncertainties of measurement of
6 in 106or 60 μg (whichever is greater) up to 12 kg

1.38 Instrument calibrators
.01 D.C. voltage
0.004% of reading or 0.004 mV (which ever is the greater) from 0 V to 100 V
.11 D.C. current
0.0075% of reading or 0.0006 mA (which ever is the greater) from 0 mA to 100 mA

1.39 Indicating and recording instruments
.01 D.C. voltmeters
With least uncertainties of measurement of -
0.004% of reading or 0.004 mV (which ever is the greater) from 0 V to 100 V
.03 D.C. ammeters
With least uncertainties of measurement of -
0.0075% of reading or 0.0006 mA (which ever is the greater) from 0 mA to 100 mA