Back to Search Results


Address Details:
1064 Centre Road
OAKLEIGH SOUTH, VIC 3167
AUSTRALIA Map
www.vicom.com.au

Contact Details:
Mr Rohan Koottala
+61(03) 95750119

Availability:
Services available to external clients

SCOPE OF ACCREDITATION   View and Print Scope of Accreditation Printable version

Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.


  • ISO/IEC 17025 - Calibration The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise (2005)
    Communications, EMR and EMC Equipment
  • SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
    Calibration of communications, electromagnetic field strength and EMC test equipment
    Impedance and reflection measuring equipment;
    Reflection coefficient;
    To be determined;
    VSWR and return loss, 30 kHz to 18 GHz; 
    Directivity and source match measurement, 10 MHz to 18 GHz;
    Source match from 2 to 18 GHz in the range 10 dB match to greater or equal to 30 dB match
    MEASUREMENT UNCERTAINTY
    with Calibration and Measurement Capability of - 
    VSWR and return loss measuring equipment
    0.021+0.08 ρ2 for reflection coefficient from 30 kHz to 250 MHz ;
    0.014+0.021 ρ2 for reflection coefficient from 250 MHz to 18 GHz
    Directivity and source match measurement
    0.015 equivalent reflection coefficient from 10 MHz to 2 GHz ;
    0.006 equivalent reflection coefficient from 2 GHz to 18 GHz
    Source match from 2 to 18 GHz
    0.175 equivalent reflection coefficient for 10 dB match ;
    0.052 equivalent reflection coefficient for 15 dB match ;
    0.023 equivalent reflection coefficient for 20 dB match ;
    0.017 equivalent reflection coefficient for 25 dB match ;
    0.015 equivalent reflection coefficient for ≥ 30 dB match 

    Amplifiers; Attenuators;
    Attenuation;
    To be determined;
    30 kHz to 18 GHz, and
    0 to 120 dB
    MEASUREMENT UNCERTAINTY
    Attenuators and amplifiers
    with Calibration and Measurement Capability of - 
    0.35 dB from 0 to 30 dB and 30 kHz to 100 kHz ;
    0.25 dB from 0 to 30 dB and 100 kHz to 300 kHz ;
    0.2 dB from 0 to 30 dB and 300 kHz to 2.5 MHz ;
    0.15 dB from 0 to 40 dB and 2.5 MHz to 1300 MHz ;
    0.2 dB from 40 to 80 dB and 2.5 MHz to 1300 MHz ;
    0.25 dB from 80 to 120 dB and 2.5 MHz to 1300 MHz ;
    0.25 dB from 0 to 30 dB and 1.3 to 8 GHz ; 
    0.3 dB from 0 to 30 dB and 8 to 10 GHz ; 
    0.4 dB from 0 to 30 dB and 10 to 12 GHz ; 
    0.45 dB from 0 to 30 dB and 12 to 18 GHz 

  • ISO/IEC 17025 - Calibration The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise (2005)
    Time and Frequency Metrology
  • SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
    Calibration of frequency and time standards
    Frequency standards;
    Frequency;
    To be determined;
    at 1, 2, 2.5, 5 and 10 MHz
    MEASUREMENT UNCERTAINTY
    with Calibration and Measurement Capability of - 
    2 in 1011 at 1, 2, 2.5, 5 and 10 MHz

    Signal sources;
    Frequency; Power; Voltage;
    To be determined;
    Frequency characteristics from 10 to 40 GHz ;
    Power measurements from 0 to 120 ; 1mW to 30 W and 30 kHz to 40 GHz ; 
    Pulse characteristics from 10 ns to 1 s ;
    Measurement of internal D.C. voltage and current in accordance with manufacturers' handbook - 
    DC Volts from 10 mV to 1000 V ;
    DC Current from 1 mA to 3 A
    MEASUREMENT UNCERTAINTY
    Frequency characteristics
    with Calibration and Measurement Capability of - 
    5 in 106 from 10 to 100 Hz ;
    5 in 107 from 100 Hz to 1 kHz ;
    5 in 108 from 1 to 10 kHz ;
    5 in 109 from 10 to 100 kHz ;
    5 in 1010 from 100 kHz to 1 MHz ;
    6 in 1011 from 1 to 10 MHz ;
    4 in 1011 from 10 MHz to 3 GHz ;
    3 in 1010 from 3 to 10 GHz ;
    1 in 1010 from 10 to 20 GHz ;
    5 in 1011 from 20 to 30 GHz ;
    3 in 1011 from 30 to 40 GHz
    Power measurements
    with Calibration and Measurement Capability of - 
    0.6% at 1 mW and 50 MHz ;
    2.7% from 1 to 30 W and 1.8 MHz to 1 GHz ;
    3.7% from 0 to 35 dBm and 30 kHz to 50 kHz ;
    2.9% from 0 to 35 dBm and 50 kHz to 4.2 GHz ;
    3.1% from -20 to -10 dBm and 10 to 30 MHz ;
    2.8% from -20 to -10 dBm and 30 MHz to 2 GHz ;
    3.3% from -20 to -10 dBm and 2 to 12.4 GHz ;
    3.7% from -20 to -10 dBm and 12.4 to 18 GHz ;
    2.5% from -10 to +20 dBm and 10 to 30 MHz ;
    2.1% from -10 to +20 dBm and 30 MHz to 2 GHz ;
    2.8% from -10 to +20 dBm and 2 to 12.4 GHz ;
    3.2% from -10 to +20 dBm and 12.4 to 18 GHz ;
    4.3% from -20 to +20 dBm and 18 to 26 GHz ;
    4.8% from -20 to +20 dBm and 26 to 38 GHz ;
    5.6% from -20 to +20 dBm and 38 to 40 GHz ;
    3.7% from -50 to -20 dBm and 10 to 30 MHz ;
    3.6% from -50 to -20 dBm and 30 MHz to 2 GHz ;
    5.5% from -50 to -20 dBm and 2 to 18 GHz ;
    7.3% from -50 to -20 dBm and 18 to 33 GHz ;
    7.7% from -50 to -20 dBm and 33 to 40 GHz ;
    5.1% from -60 to -50 dBm and 10 to 30 MHz ;
    5.0% from -60 to -50 dBm and 30 MHz to 2 GHz ;
    5.6% from -60 to -50 dBm and 2 to 18 GHz ;
    7.3% from -60 to -50 dBm and 18 to 33 GHz ;
    8.0% from -60 to -50 dBm and 33 to 40 GHz ;
    36% from -70 to -60 dBm and 10 MHz to 40 GHz ;
    4.4% from -40 to 0 dBm and 2.5 to 1300 MHz ;
    4.7% from -80 to -40 dBm and 2.5 to 1300 MHz ;
    8.6% from -120 to -80 dBm and 2.5 to 1300 MHz
    Pulse characteristics 
    with Calibration and Measurement Capability of - 
    0.2 ns or 0.002% (whichever is greater) for pulse spacing ;
    1.2 ns or 0.002% (whichever is greater) for pulse width ;
    0.4 ns or 0.004% (whichever is greater) for rise time
    Measurement of internal instrument parameters
    with Calibration and Measurement Capability of - 
    DC Volts 0.015% of reading + 5 μV ;
    DC Current 0.2% of reading + 2 μA
    Calibration of frequency, time and waveform measuring equipment
    Counters;
    Frequency;
    To be determined;
    Counters from 10 to 40 GHz;
    MEASUREMENT UNCERTAINTY
    Counters
    with Calibration and Measurement Capability of -
    5 in 106 from 10 to 100 Hz ;
    5 in 10from 100 Hz to 1 kHz ;
    5 in 108 from 1 to 10 kHz ;
    5 in 109 from 10 to 100 kHz ;
    5 in 1010 from 100 kHz to 1 MHz ;
    6 in 1011 from 1 to 10 MHz ;
    4 in 1011 from 10 MHz to 3 GHz ;
    3 in 1010 from 3 to 10 GHz ;
    1 in 1010 from 10 to 20 GHz ;
    5 in 1011 from 20 to 30 GHz ;
    3 in 1011 from 30 to 40 GHz