Back to Search Results


Address Details:
11 River Road
Dinmore, QLD 4303
AUSTRALIA Map
www.mobiletestncal.com.au

Contact Details:
Mr Harvey Wharton
+61(07) 32820007

Availability:
Services available to external clients

SCOPE OF ACCREDITATION   View and Print Scope of Accreditation Printable version

Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.


  • ISO/IEC 17025 (2005) - The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
    Calibration
  • SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
    DC and low frequency electrical metrology - Electrical measurement and test equipment
    Digital multimeters (DMM); Ohm meters;
    Resistance;
    Comparison with a reference standard;
    up to 100 MΩ
    CAPABILITY
    With Calibration and Measurement Capability of - 
    7 mΩ at 0 Ω, 1 Ω, 10 Ω
    0.011 Ω at 100 Ω
    0.083 Ω at 1 kΩ
    0.82 Ω at 10 kΩ
    0.004 kΩ at 100 kΩ
    0.00015 MΩ at 1 MΩ
    0.0039 MΩ at 10 MΩ
    0.52 MΩ at 100 MΩ

    Digital multimeters (DMM); Voltmeters;
    AC voltage; DC voltage;
    Comparison with a reference standard;
    from 0 to 1000 V
    from DC to 100 kHz
    CAPABILITY
    With Calibration and Measurement Capability of -

    DC Voltage
    0.004 % + 3 μV from 0 to 200 mV
     0.003 % + 4 μV from 200 mV to 2 V
     0.003 % + 40 μV from 2 V to 20 V
     0.003 % + 0.4 mV from 20 V to 200 V
     0.003 % + 4 mV from 200 V to 1000 V

    AC Voltage
    0.032 % + 20 μV from 0.1 mV to 200 mV and from 45 Hz to 1 kHz,
     0.033 % + 90 μV from 200 mV to 2 V and from 45 Hz to 1 kHz,
     0.031 % + 1 mV from 2 V to 20 V and from 45 Hz to 1 kHz,
     0.035 % + 8 mV from 20 V to 200 V and from 45 Hz to 1 kHz,
     0.036 % + 75 mV from 200 V to 750 V and from 45 Hz to 1 kHz.

     0.07 % + 35 μV from 0.1 mV to 200 mV and from 1 kHz to 20 kHz,
     0.06 % + 135 μV from 200 mV to 2 V and from 1 kHz to 20 kHz,
     0.05 % + 2 mV from 2 V to 20 V and from 1 kHz to 20 kHz.
     0.09 % + 40 mV from 20 V to 200 V and from 1 kHz to 20 kHz,
    0.13 % + 400 mV from 200 V to 750 V and from 1 kHz to 10 kHz.

     0.3 % + 70 μV from 0.4 mV to 200 mV and from 20 kHz to 100 kHz,
     0.22 % + 2 mV from 200 mV to 2 V and from 20 kHz to 100 kHz.
     0.18 % + 33 mV from 2 V to 20 V and from 20 kHz to 100 kHz.




    Ammeters; Digital multimeters (DMM);
    AC current; DC current;
    Comparison with a reference standard;
    from 0 to 10 A
    from DC to 2 kHz

    CAPABILITY
    With Calibration and Measurement Capability of -


    D.C. current
     0.01 % + 30 nA from 0 to 200 μA,
     0.0009 % + 40 nA from 200 μA to 2 mA,
     0.007 % + 0.3 μA from 2 mA to 20 mA,
     0.008 % + 3 μA from 20 mA to 200 mA,
     0.014 % + 35 μA from 200 mA to 2 A,
    0.032 % + 0.35 mA from 2 A to 10 A


    A.C. current
     0.10% + 0.25 μA from 0.1 μA to 100 μA and from 45 Hz to 1 kHz,
     0.09% + 0.5 μA from 100 μA to 1 mA and from 45 Hz to 1 kHz,
     0.08% + 4 μA from 1 mA to 10 mA and from 45 Hz to 1 kHz,
     0.09% + 40 μA from 10 mA to 100 mA and from 45 Hz to 1 kHz,
     0.10% + 400 uA from 100 mA to 1 A and from 45 Hz to 1 kHz,
     0.06% + 2 mA from 1 A to 10 A and from 45 Hz to 100 Hz,

    1.30% + 0.25 μA from 1.5 μA to 100 μA and from 1 kHz to 2 kHz,
    0.70% + 0.7 μA from 100 μA to 1 mA and from 1 kHz to 2 kHz,
    0.46 % + 7 μA from 1 mA to 10 mA and from 1 kHz to 2 kHz,
     0.46 % + 70 μA from 10 mA to 100 mA and from 1 kHz to 2 kHz,
     0.60% + 700 μA from 100 mA to 1 A and from 1 kHz to 2 kHz,
     0.30 % + 4 mA from 1 A to 10 A and from 100 Hz to 1 kHz,