Back to Search Results


Address Details:
66 Malcolm Road
BRAESIDE , VIC 3195
AUSTRALIA Map
www.aga.asn.au

Contact Details:
Mr B Tabourlos
+61(03) 95804500

Availability:
Services available to external clients

SCOPE OF ACCREDITATION   View and Print Scope of Accreditation Printable version

Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.



ISO/IEC 17025 (2005)
Calibration
SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
Chemical metrology - Gas analysers
Gas analysers; Mine safety and confined space equipment;
Concentration;
Direct measurement by certified reference material;
in-house method QWI 739
CAPABILITY
 With calibration and measurement capability - 
O2 in air: Paramagnetic -  0.4% to 21% Ov/v
2.4% of reading over the range 0.4% to 21% Ov/v


Concentration;
Direct measurement by certified reference material;
in-house method QWI 738 
CAPABILITY
 With calibration and measurement capability - 
CO in air: NDIR  - 3 ppm to 3000 ppm CO v/v
 2.5% of reading over the range 3 ppm to 3000 ppm CO v/v
CO2 in air: NDIR -  0.1% to 20% COv/v
2.5% of reading over the range.of 0.1% to 20% COv/v


Concentration;
Direct measurement by certified reference material;
in-house method QWI 736
CAPABILITY
With calibration and measurement capability - 
NO in air: Chemiluminescent - 0.1 ppm to 10 ppm NO v/v
4.3% of reading over the range 0.1 ppm to 10 ppm NO v/v
NOin air: Chemiluminescent - 0.1 ppm to 10 ppm NOv/v
4.3% of reading over the range 0.1 ppm to 10 ppm NOv/v
NOX in air: Chemiluminescent - 0.1 ppm to 10 ppm NOv/v
4.3% of reading over the range 0.1 ppm to 10 ppm NOv/v


Concentration;
Direct measurement by certified reference material;
in-house method QWI 737 
CAPABILITY
With calibration and measurement capability - 
CO in air: Electrochemical - 10 ppm to 3000 ppm CO v/v
4% of reading over the range of 10 ppm to 3000 ppm CO v/v
O2 in air: Electrochemical - 0.4% to 21% O2 v/v
4% of reading over the range 0.4% to 21% O2 v/v
Pressure metrology - Pressure and vacuum measuring equipment
Manometers; Pressure transducers;
Gauge pressure;
Comparison with reference instrument;
Metrology Society of Australia Test Method 1 
CAPABILITY
with Calibration measurement capability of 

0.03 kPa from 0 kPa to 10 kPa
3 kPa above 10 kPa to 4 MPa

Pressure gauges;
Gauge pressure;
Comparison with reference instrument;
MetrologySociety of Australia Test Method 2 
CAPABILITY
with Calibration measurement capability of

0.03 kPa from 0 kPa to 10 kPa
3 kPa above 10 kPa to 4 MPa