| SERVICE | PRODUCT | DETERMINANT | TECHNIQUE | PROCEDURE | LIMITATION / RANGE |
DC and low frequency electrical metrology - Electrical measurement and test equipment
| Insulation resistance test equipment;
| Resistance;
| Comparison with a reference standard by substitution measurement; | By the method of PSS WI-11 | At voltages of 50 V to 5 kV
|
|
CAPABILITY with Calibration and Measurement Capability of; 3% from 1 MΩ - 1 GΩ; 5% from 1 GΩ - 10 GΩ; 10% from 10 GΩ - 100 GΩ; |
| Ammeters;
| AC current;
| Indirect measurement against a reference standard; | by method of single or multiple turns | from 50 mA up to 5000 A at 50 Hz |
|
CAPABILITY with Calibration and Measurement Capability of -; 1% of reading; from 50 mA up to 5000 A at 50 Hz;
|
DC and low frequency electrical metrology - High-voltage/high-current standards and equipment
| Current transformers;
| Current error; Phase displacement;
| Comparison with a reference standard by differential measurement; | AS 1675; AS/IEC 60044.1; IEC 61869-1,-2 | |
|
CAPABILITY Accuracy tests at 50 Hz, up to 5000 A applied primary current; for CTs with rated secondaries of 1 A or 5 A; with Calibration and Measurement Capability of -; ±0.025% for current error; ±0.030 crad for phase displacement; |
| Voltage transformers;
| Phase displacement; Voltage error;
| Comparison with a reference standard by differential measurement; | AS 1243; AS/IEC 60044.2 AS/IEC 61869-1,-3; | |
|
CAPABILITY Accuracy tests at 50 Hz, up to rated primary voltage of 132 kV/√3; Single phase measurements of VTs with rated secondaries of 63.5 V or 110 V;with Calibration and Measurement Capability of -; 0.025% for voltage error; 0.025 crad for phase displacement;
|
| High voltage test sets;
| Voltage;
| Direct measurement against a reference standard; | | Measurements performed in the laboratory and in the field ; a.c. test sets from 1 kVrms to 100 kVrms at 50 Hz; from ±1 kVpeak/√2 to ±100 kVpeak/√2, at 50 Hz;
from ±1 kVpeak to ±140 kVpeak, at 50 Hz;
d.c. test sets from - 90 kV to +90 kV |
|
CAPABILITY Measurements performed in the laboratory; with Calibration and Measurement Capability of -;
a.c. test sets
2% of reading from 1 kVrms to 100 kVrms at 50 Hz; 2% of reading from ±1 kVpeak/√2 to ±100 kVpeak/√2, at 50 Hz;
2% of reading from ±1 kVpeak to ±140 kVpeak, at 50 Hz;
d.c. test sets 2% of reading from -90 kV to 90 kV
Measurements performed in the field;
with Calibration and Measurement Capability of;
a.c. test sets 2.5% of reading from 1 kVrms to 100 kVrms at 50 Hz; 2.5% of reading from ±1 kVpeak/√2 to ±100 kVpeak/√2, at 50 Hz; 2.5% of reading from ±1 kVpeak to ±140 kVpeak, at 50 Hz;
d.c. test sets 2.5% of reading from -90 kV to 90 kV |