SCOPE OF ACCREDITATION   View and Print Scope of Accreditation Printable version

Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.

  • ISO/IEC 17025 (2005) - The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
    Calibration
  • SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
    High frequency electrical metrology - Communications, electromagnetic field strength and EMC test equipment
    E&H field probes and meters; Electromagnetic field hazard meters and warning indicators;
    Frequency;
    Comparison with a reference standard;
    CAPABILITY
    with Calibration and Measurement Capability of-
    Calibration of E & H-field probes in the range 
    0.01 MHz to 433 MHz at 120 V/m and 2.4 A/m
    433 MHz to 1 GHz at 61.4 V/m and 0.55 A/m
    1 GHz to 3 GHz at 61.4 V/m
    3 GHz to 8 GHz at 61.4 V/m
    8 GHz to 18 GHz at 61.4 V/m
    ± 1.2 dB from 0.01 MHz to 3 GHz
    ± 1.5 dB from 3 GHz to 8 GHz
    ± 1.8 dB from 8 GHz to 18 GHz




    Electrical fast transients (EFT) and surge generators;
    Pulse characteristics;
    Direct measurement by electrical input;
    IEC 61000-4-5, clause 6.1 (1.2/50 μs - 8/20 μs) ;
    CAPABILITY
    Surge Generators
    with Calibration and Measurement Capability of -
    in the range ± 0.1 kV to 7 kV
    ± 5.60% for open circuit voltage
    ± 10.53% for open circuit front time
    ± 4.4% for open circuit duration
    ± 4.4% for short circuit current
    ± 4.29% for short circuit front time
    ± 4.3% for short circuit duration


    Pulse characteristics;
    Direct measurement by electrical input;
    IEC 61000-4-4, clause 6.1 ;
    CAPABILITY
    Electrical fast transients (EFT)
    with Calibration and Measurement Capability of -
    in the range ± 0.5 kV to 20 kV
    ± 5.80% for peak voltage
    ± 4.2% for pulse width
    ± 1.2% for rise time
    ± 3.07% for repetition rate

    Harmonic emission compliance test systems;
    Harmonic currents;
    Comparison with a reference standard by differential measurement;
    CAPABILITY
    with Calibration and Measurement Capability of -
    0.15% of the fundamental of current for Harmonic currents 
    - from 0.7 A to 16 A total rms per phase 
    - from 1st to 40th harmonic with fundamental frequency between 50 to 60 Hz and voltage between 120 to 230 V

    Coupling/decoupling networks (CDN);
    Insertion loss;
    Direct measurement by electrical input;
    IEC 61000-4-6, clause 6.2 and Annex B
    CAPABILITY
    with Calibration and Measurement Capability of -
    in the frequency range 150 kHz to 230 MHz
    ± 10.9% for impedance

    Electrostatic discharge (ESD) generators;
    Pulse characteristics;
    Direct measurement by electrical input;
    IEC 61000-4-2 clause 6.2 ;
    ISO 10605, Annex A
    CAPABILITY
    with Calibration and Measurement Capability of -
    in the range ± 2 kV to ± 30 kV
    ± 14.2% for rise time
    ± 8.10% for peak current
    ± 1.2% for peak voltage

    Artificial mains networks (AMN); Line impedance stabilisation networks (LISN);
    Voltage division factor;
    Direct measurement by electrical input;
    CISPR 16-1-2, clause 4
    CAPABILITY
    with Calibration and Measurement Capability of -
    in the frequency range 9 kHz to 108 MHz
    ± 0.6 dB for voltage division factor  
    ± 2.1% for voltage drop
    ± 10.9% for impedance
    ± 3.8º for phase

    Line impedance stabilisation networks (LISN);
    Voltage division factor;
    Direct measurement by electrical input;
    CISPR 16-1-2 (clause 7.2 - Table 5 excluding Insertion loss ; Annex E - Clause E.2)
    CAPABILITY
    Balanced Telecom ISN (TLISN)
    with Calibration and Measurement Capability of -
    in the frequency range 150 kHz to 30 MHz
    ± 4.3 Ω for common mode impedance  
    ± 2.4º for common mode phase
    ± 0.97 dB for voltage division factor
    ± 1.12 dB for isolation / DCMD
    ± 2.28 dB for Longitudinal conversion loss (LCL)

    Voltage fluctuation and flicker compliance test systems;
    Internal reference impedance resistance; Short term flicker; Voltage deviation;
    Comparison with a reference standard by substitution measurement;
    CAPABILITY
    with Calibration and Measurement Capability of -
    0.5% of Pst in the range from 0.5 Ps to 1.5 Ps for Short term flicker perceptibility
    0.07% of fundamental of the voltage in the range from 0.5 to 2.0% for Voltage deviation, Dmax, with fundamental frequency between 50 Hz to 60 Hz and voltage between 1 V to 230 V
    1.8 MΩ for Internal reference impedance resistance

    Active rod antennas; Biconical antennas; Horn antennas; Log periodic antennas;
    Antenna factor; Frequency;
    Direct measurement by electrical input;
    SAE ARP 958, version D
    CAPABILITY
    Narrowband transducers and indicators
    Radiofrequency antenna
    with Calibration and Measurement Capability of-
    Antenna Factor from -5 dB(1/m) up to 50 dB(1/m)
    ± 1.0 dB from 9kHz to 40 GHz
    Active Rod antennas 9kHz to 30MHz
    Biconical antennas 30MHz to 300MHz
    Log periodic antennas 200MHz to 1000MHz
    Large Horn antennas 200MHz to 2000MHz
    Small Horn antennas 1GHz to 40 GHz