Back to Search Results


Address Details:
26 Newton Road
PERTH AIRPORT, WA 6105
AUSTRALIA Map
www.airflite.com.au

Contact Details:
Mr W van Staden
+61(08) 94789000

Availability:
Services available to external clients

SCOPE OF ACCREDITATION   View and Print Scope of Accreditation Printable version

Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.



ISO/IEC 17025 (2005)
Calibration
SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
DC and low frequency electrical metrology - Electrical measurement and test equipment
Insulation resistance test equipment;
Resistance;
Direct measurement against a reference standard;
CAPABILITY
with Calibration and Measurement Capability of -
0.04% from 30 Ω to 199.9 Ω
0.12% from 200 Ω to 499 Ω
0.04% from 500 Ω to 1.999 kΩ
0.12% from 2.00 kΩ to 4.99 kΩ
0.06% from 5.00 kΩ to 10 kΩ
0.03% from 10 kΩ to 39.99 kΩ at 55 V
0.02% from 40 kΩ to 99.99 kΩ at 400 V
0.03% from 100 kΩ to 199.99 kΩ at 800 V
0.02% from 200 kΩ to 999. 9 kΩ at 1100 V
0.2% from 1 MΩ to 9.999 MΩ at 1150 V
1.0% from 10 MΩ to 99.99 MΩ at 1570 V
2.1% from 100 MΩ to 999.9 MΩ at 1570 V


Ammeters; Ohm meters; Voltmeters;
AC current; AC voltage; DC current; DC voltage; Resistance;
Direct measurement against a reference standard;
by in -house methods of
LMS-P-ELC-09 Calibration of clamp meters/Tong Testers
LMS-P-ELC-01 Calibration of Digital Multimeters
CAPABILITY
with Calibration and Measurement Capability of-
D.C. ammeters
0.003 % from 1μA to 100 μA
0.02% from 100 μA to 1 A
0.03% from 1 A to 10 A
0.5% plus 0.1 A from 10 A to 50 A
0.5% plus 0.05 A from 50 A to 100 A
0.5% plus 0.2 A from 100 A to 300 A
0.5% from 300 A to 500 A
A.C. ammeters
0.06% from 1 mA to 10 mA at %0 Hz and 1 kHz
0.02% from 10 mA to 10 A at 50 Hz and 1 KHz
Non Toroidal clamps using multiple turn technique
0.5% plus 0.2 A at 50 Hz from 10 A to 100 A
0.8% at 50 Hz from 100 A to 200 A
0.7% at 50 Hz from 200 A to 300 A
0.65% at 50 Hz from 300 A to 400 A
0.6% at 50 Hz from 400 A to 500 A
0.5% plus 0.3 A at 400 Hz from 10 A to 30 A
0.5% plus 0.35 A at 400 Hz from 30 A to 50 A
0.5% plus 0.4 A at 400 Hz from 50 A to 70 A
0.8% plus 0.2 A at 400 Hz from 70 A to 100 A
1.0% plus 0.3 A at 400 Hz from 100 A to 200 A
1.0% plus 0.1 A at 400 Hz from 200 A to 300 A
0.95% at 400 Hz from 300 A to 500 A
DC Voltmeters
0.08% from 1 mV to 10 mV
0.03% from 10 mV to 100 mV
0.001% from 100 mV to 100 V
0.002% from 100 V to 1000 V
AC voltmeters
0.5% from 1 mV to 10 mV at 50 Hz and 1 kHz
0.2 % from 10 mV to 100 mV at 50 Hz and 1 kHz
0.02% from 100 mV to 100V at 50 Hz and 1 kHz
0.01% from 100 V to 1000 V at 50 Hz and 1 kHz
Ohm Meters
1% from 0.001 Ω to 0.009 Ω
0.1% from 0.01 Ω to 0.09 Ω
0.02% from 0.1 Ω to 0.9 Ω
0.006% from 1 Ω to 9 Ω
0.02% from 10 Ω to 1 MΩ
0.2% from 10 MΩ to 100 MΩ
DC and low frequency electrical metrology - Electrical standards
Resistance boxes;
Resistance;
Direct measurement against a reference standard;
By the methods of-
in-house method LMS-P-ELC-11 Calibration of resistance boxes
Resistance boxes only from 0.001 Ω to 100 MΩ
CAPABILITY
with Calibration and Measurement Capability of -
1% from 0.001 Ω to 0.009 Ω
0.1% from 0.01 Ω to 0.09 Ω
0.02% from 0.1 Ω to 0.9 Ω
0.006% from 1 Ω to 9 Ω
0.02% from 10 Ω to 1 MΩ
0.2% from 10 MΩ to 100 MΩ 
Pressure metrology - Pressure and vacuum measuring equipment
Digital pressure gauges; Mechanical pressure gauges; Vacuum gauges;
Gauge pressure;
Comparison with reference instrument;
By the methods of –
AS 1349 and the Metrology Society of Australia (MSA) Test Methods 1 and 2
Gauge pressure calibration including test gauges and digital indicators in the range -80 kPa to 70 MPa
CAPABILITY
with Calibration and Measurement Capability of -
0.06 kPa from -80 kPa to 700 kPa
 1.9 kPa from 700 kPa to 7,000 kPa
 2.0 kPa from 7,000 kPa to 14,000 kPa
 0.01 MPa from 14 MPa to 70 MPa
Temperature metrology - Non-contact temperature measuring equipment
Radiation pyrometers; Thermal imaging systems;
Temperature;
Comparison with a reference standard; To be determined;
CAPABILITY
with Calibration and Measurement Capability of -
Calibration of non-contact thermometers and thermal imaging cameras (8-14 μm)
 0.7 °C at 0 °C
 0.7 °C + 0.20%T from 35 °C to 100 °C
 0.7 °C + 0.35%T from 100 °C to 200 °C
 0.7 °C + 0.4%T from 200 °C to 350 °C
 0.7 °C + 0.36%T from 350 °C to 500 °C