ISO/IEC 17025 (2005)
Calibration
SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
DC and low frequency electrical metrology - Electrical instrument calibrators
Instrument calibrators;
DC current; DC voltage; Resistance;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of -
Process calibrators only
DC current
0.05% reading from 0.1 mA to 100 mA  
DC voltage
0.05% reading from 0.01 V to 100 V
Resistance
0.1% reading ± 1 mΩ from 0.1 Ω to 10 kΩ 

DC and low frequency electrical metrology - Electrical measurement and test equipment
Ammeters; Current clamps; Ohm meters; Voltmeters;
AC current; AC voltage; DC current; DC voltage; Resistance;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of -
AC current
0.1% reading from 10 mA to 20 A at 50 and 60 Hz
0.5% reading from 20 A to 1000 A for Toroidal Current Clamps [using multiple-turn technique]
1.0% reading from 20 A to 1000 A for Non-Toroidal Current Clamps [using multiple-turn technique]
DC current
0.05% of reading from 0.1 mA to 20 A
0.9% of reading from 20 A to 1000 A for Current Clamps [using multiple-turn technique]
AC voltage
0.1% of reading  from 0.5 V to 700 V at 50 Hz and 60 Hz
DC voltage
0.05% of reading from 0.01 V to 1000 V
Resistance
0.5% of reading from 1 Ω to 100 kΩ
2% of reading above 100 kΩ to 10 MΩ
5% of reading above 10 MΩ to 100 MΩ
10% of reading above 100 MΩ to 1 GΩ
DC and low frequency electrical metrology - Electrical standards
Conductance boxes; Precision resistors; Resistance boxes;
Resistance;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of -
0.5% of reading + 1 mΩ from 0.001 Ω to 0.1 Ω
0.1% of reading + 1 mΩ from 0.1 Ω to 10 kΩ
0.5% of reading from 10 kΩ  to 100 kΩ
2% of reading from 100 kΩ to 10 MΩ
Pressure metrology - Pressure and vacuum measuring equipment
Manometers;
Gauge pressure;
By pressure calibrator;
MSA Test Method 1
including on site calibrations



CAPABILITY
Including on site calibrations
with Calibration and Measurement Capability of-
0.02 kPa from -20 kPa to 20 kPa
Pressure metrology - Pressure and vacuum measuring equipment
Digital pressure gauges; Mechanical pressure gauges; Pressure control devices; Pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges;
Gauge pressure;
By pressure calibrator;
AS 1349 (pressure and vacuum gauges)
MSA Test Method 2 (pressure and vacuum gauges)
MSA Test Method 1 (pressure transducers, pressure recorders and industrial grade, self indicating and electrical transducers with dc voltage, dc current or resistance outputs)


including on site calibrations



CAPABILITY
Including on site calibrations
with Calibration and Measurement Capability of-
0.2 kPa from -95 to 0 kPa;
0.05% of reading or 0.3 kPa (whichever is greater) from 0 kPa up to 4 MPa;
0.05% of reading or 3 kPa (whichever is greater) from 4 MPa to 100 MPa
Temperature metrology - Ancillary temperature measuring equipment
Digital voltmeters (DVM) and digital multimeters (DMM); Process calibrators;
Resistance to temperature conversion; Volt to temperature conversion;
Direct measurement by electrical input;
CAPABILITY
with Calibration and Measurement Capability of -
Temperature calibrator instruments
by electrical simulation [measure and source mode]

 0.8 °C + 0.1% of Temperature for Type K thermocouples from -100 °C to +1100 °C
 0.8 °C + 0.1% of Temperature for Type T thermocouples from -100 °C to +390 °C
 0.8 °C + 0.1% of Temperature for Type N thermocouples from +100 °C to +1200 °C
 0.3 °C for RTD PT100 from -100 °C to 400 °C
Temperature metrology - Temperature measuring equipment
Digital temperature measuring systems; Temperature loggers - Single and multi-channel;
Temperature;
Measurement against reference standard;
CAPABILITY
with Calibration and Measurement Capability of -
0.5°C from -50°C to 0°C
0.2 °C above 0°C to 150°C
0.3 °C above 150°C to 200°C
Analogue chart recorders
0.5°C from -50°C to 200°C
Temperature metrology - Temperature standards and reference equipment
Dry block calibrators;
Temperature;
Direct measurement using a reference standard;
CAPABILITY
with Calibration and Measurement Capability of -
0.5 °C from 0°C to 500°C
Temperature metrology - Verification of controlled enclosures
Furnaces; Ovens;
Spatial uniformity; Temperature; Temporal uniformity;
Direct temperature measurement;
including on site calibrations



CAPABILITY
including on site calibrations
with Calibration and Measurement Capability of-
1°C from 0°C to 200 °C
 2.5°C above 200°C to 400°C
 5°C above 400°C to 1100°C
Temperature metrology - Verification of controlled enclosures
Medical refrigeration equipment;
Spatial uniformity; Temperature;
Direct temperature measurement;
AS 3864.2 sections 3.7 and 3.8.3
including on site calibrations



CAPABILITY
Including on site calibrations
with Calibration and Measurement Capability of-
AS 3864.2 section 3.7 - Annual tasks
0.8°C from -50°C to 0°C
0.3°C above 0°C to 10°C
AS 3864.2 section 3.8.3 - Spatial temperature distribution check
0.8°C from -50°C to 10°C