Address Details:
Unit 5/52 Vinnicombe Drive
CANNING VALE, WA 6155
AUSTRALIA Map
www.hinco.com.au

Contact Details:
Mr John French
+61(08) 94556777

Availability:
Services available to external clients

SCOPE OF ACCREDITATION   View and Print Scope of Accreditation Printable version

Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.


  • ISO/IEC 17025 - Calibration The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise (2005)
    DC and Low Frequency Electrical Metrology
  • SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
    Calibration of electrical instrument calibrators
    Instrument calibrators;
    DC current;

    0.1 mA to 100 mA
    Process calibrators only
    MEASUREMENT UNCERTAINTY
    0.05% reading from 0.1 to 100 mA


    Resistance;

    Process calibrators only
    MEASUREMENT UNCERTAINTY
    0.1% reading ± 1 mΩ from 0.1 Ω to 10 kΩ


    DC voltage;

    0.01 V to 100 V
    Process calibrators only
    MEASUREMENT UNCERTAINTY
    0.05% reading  
    Calibration of electrical measurement and test equipment
    Ammeters;
    AC current;

    10 mA to 100 A
    MEASUREMENT UNCERTAINTY
    0.1% reading from 10 mA to 20 A at 50 and 60 Hz
    0.5% reading from 20 A to 1000 A for Toroidal Current Clamps [using multiple-turn technique]
    1.0% reading from 20 A to 1000 A for Non-Toroidal Current Clamps [using multiple-turn technique]


    DC current;

    0.1 mA to 1000 A
    MEASUREMENT UNCERTAINTY
    0.05% of reading from 0.1 mA to 20 A
    0.9% of reading from 20 A to 1000 A for Current Clamps [using multiple-turn technique]

    Ohm meters;
    Resistance;

    1Ω to 1 GΩ
    MEASUREMENT UNCERTAINTY
    0.5% of reading from 1Ω to 100 kΩ
    2% of reading above 100 kΩ to 10 MΩ
    5% reading above 10 MΩ to 100 MΩ
    10%reading above 100 MΩ to 1 GΩ

    Voltmeters;
    AC voltage;

    0.5 V to 700 V at 50 and 60 Hz
    MEASUREMENT UNCERTAINTY
    0.1% of reading


    DC voltage;

    0.01 V to 1000 V
    MEASUREMENT UNCERTAINTY
    0.05% of reading
    Calibration of electrical standards
    Conductance boxes; Precision resistors; Resistance boxes;
    Resistance;

    0.001 Ω to 10 MΩ
    MEASUREMENT UNCERTAINTY
    0.5% of reading + 1 mΩ from 0.001 Ω to 0.1 Ω
    0.1% of reading + 1 mΩ from 0.1 Ω to 10 kΩ
    0.5% of reading from 10 to 100 kΩ
    2% of reading from 100 kΩ to 10 MΩ

  • ISO/IEC 17025 - Calibration The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise (2005)
    Pressure Metrology
  • SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
    Calibration of pressure and vacuum measuring equipment
    Manometers;
    Gauge pressure;
    By pressure calibrator;
    MSA Test Method 1
    -20 kPa to 20 kPa
    Including on site calibrations



    MEASUREMENT UNCERTAINTY
    Including on site calibrations
    With Calibration and Measurement Capability of-
    0.02 kPa from -20 kPa to 20 kPa
    Calibration of pressure and vacuum measuring equipment
    Digital pressure gauges; Mechanical pressure gauges; Pressure control devices; Pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges;
    Gauge pressure;
    By pressure calibrator;
    AS 1349 (pressure and vacuum gauges)
    MSA Test Method 2 (pressure and vacuum gauges)
    MSA Test Method 1 (pressure transducers, pressure recorders and industrial grade, self indicating and electrical transducers with dc voltage, dc current or resistance outputs)


    -95 kPa to 100 MPa
    Including on site calibrations



    MEASUREMENT UNCERTAINTY
    Including on site calibrations
    With Calibration and Measurement Capability of-
    0.2 kPa from -95 to 0 kPa;
    0.05% of reading or 0.3 kPa (whichever is greater) from 0 kPa up to 4 MPa;
    0.05% of reading or 3 kPa (whichever is greater) from 4 MPa to 100 MPa

  • ISO/IEC 17025 - Calibration The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise (2005)
    Temperature Metrology
  • SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
    Calibration of temperature measuring equipment
    Base metal thermocouples; Rare metal thermocouples;
    Temperature;

    -100°C to +1200°C
    MEASUREMENT UNCERTAINTY
    Temperature calibrator instruments
    by electrical simulation [measure and source mode]

     0.8°C + 0.1% of Temperature for Type K thermocouples from -100°C to +1100°C
     0.8°C + 0.1% of Temperature for Type T thermocouples from -100°C to +390°C
     0.8°C + 0.1% of Temperature for Type N thermocouples from +100°C to +1200°C
     0.3°C for RTD PT100 from -100°C to +400°

    Digital temperature measuring systems; Temperature loggers - Single and multi-channel;
    Temperature;

    -50°C to 200°C
    MEASUREMENT UNCERTAINTY
    0.5°C from -50°C to 0°C
    0.2 °C above 0°C to 150°C
    0.3 °C above 150°C to 200°C
    Analogue chart recorders
    0.5°C from -50°C to 200°C
    Calibration of temperature standards and reference equipment
    Dry block calibrators;
    Temperature;

    0ºC to 500ºC
    MEASUREMENT UNCERTAINTY
    0.5ºC
    Verification of controlled enclosures
    Furnaces; Ovens;
    Spatial uniformity; Temperature; Temporal uniformity;
    Direct temperature measurement;
    Including on site calibrations



    MEASUREMENT UNCERTAINTY
    Including on site calibrations
    With Calibration and Measurement Capability of-
    1°C from 0°C to 200 °C
     2.5°C above 200°C to 400°C
     5°C above 400°C to 1100°C
    Verification of controlled enclosures
    Medical refrigeration equipment;
    Spatial uniformity; Temperature;
    Direct temperature measurement;
    AS 3864.2 sections 3.7 and 3.8.3
    -50°C to 10°C
    Including on site calibrations



    MEASUREMENT UNCERTAINTY
    Including on site calibrations
    With Calibration and Measurement Capability of-
    AS 3864.2 section 3.7 - Annual tasks
    0.8°C from -50°C to 0°C
    0.3°C above 0°C to 10°C
    AS 3864.2 section 3.8.3 - Spatial temperature distribution check
    0.8°C from -50°C to 10°C