  • ISO/IEC 17025 (2005) - The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
    Calibration
  • SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
    DC and low frequency electrical metrology - Electrical instrument calibrators
    Instrument calibrators;
    AC voltage;
    Comparison with a reference standard;
    CAPABILITY
    0.15% + 34μV of reading at 0.01 mV to 10 mV and from 10 Hz to 20 kHz 
    0.25% + 70μV of reading at 0.01 mV to 10 mV and from 20 kHz to 50 kHz 
    1.1% + 110μV of reading at 0.01 mV to 10 mV and from 50 kHz to 100 kHz 
    0.02% of reading from 0.01 V to 10 V and from 10 Hz to 20 kHz 
    0.05% of reading from 0.01 V to 10 V and from 20 kHz to 50 kHz 
    0.12% of reading from 0.01 V to 10.0 V and from 50 kHz to 100 kHz 
    0.04% of reading from 10 V to 100 V at 10 Hz 
    0.03% of reading from 10 V to 100 V and from 10 Hz to 20 kHz 
    0.06% of reading from 10 V to 100 V and from 20 kHz to 50 kHz 
    0.18% of reading from 10 V to 100 V and from 50 kHz to 100 kHz 
    0.07% of reading from 100 V to 700 V at 40 Hz 
    0.09% of reading from 100 V to 700 V and from 40 Hz to 10 kHz 
    0.12 % of reading from 700 V to 1000 V and from 40 Hz to 10 kHz 


    AC current;
    Comparison with a reference standard;
    CAPABILITY
    0.6% of reading from 10 μA to 10 A at 10 Hz 
    0.3% of reading from 10 μA to 100 μA and from 10 Hz to 5 kHz 
    1.6% of reading from 10 μA to 100 μA and from 5 kHz to 10 kHz 
    0.1% of reading from 0.1 mA to 100 mA and from 10 Hz to 20 kHz 
    0.7% of reading from 0.1 mA to 100 mA and from 20 kHz to 30 kHz 
    0.2% of reading from 100 mA to 1 A and from 10 Hz to 5 kHz 
    0.5% of reading from 100 mA to 1 A and from 5 kHz to 20 kHz 
    1.6% of reading from 100 mA to 1 A and from 20 kHz to 30 kHz 
    0.3% of reading from 1 A to 10 A and from 10 Hz to 5 kHz 
    1.6% of reading from 1 A to 10 A and from 5 kHz to 10 kHz 


    Resistance;
    Comparison with a reference standard;
    0 Ω to 100 MΩ
    CAPABILITY
    0.0033% of reading + 0.2 mΩ from 0 Ω to 10 Ω 
    0.003% of reading + 2 mΩ from 10 Ω to 100 Ω 
    0.002% of reading + 1.5 Ω from 100 Ω to 100 kΩ 
    0.003% of reading + 24 Ω from 100 kΩ to 1 MΩ 
    0.01% of reading + 730 Ω from 1 MΩ to 10 MΩ 
    0.085% of reading + 0.06 MΩ from 10 MΩ to 100 MΩ 


    DC current;
    Comparison with a reference standard;
    0 μA to 10 A
    CAPABILITY
    0.004% of reading + 0.002 μA from 0 μA to 100 μA 
    0.003% of reading + 0.14 μA from 100 μA to 10 mA 
    0.005% of reading + 1.9 μA from 10 mA to 100 mA 
    0.014% of reading + 30 μA from 100 mA to 1 A 
    0.14% of reading + 0.75 mA from 1 A to 3A 
    0.18% of reading + 0.22 mA from 3 A to 10 A 


    DC voltage;
    Comparison with a reference standard;
    0 V to 1000 V
    CAPABILITY
    0.0017% of reading + 1.4 μV from 0 V to 0.1 V 
    0.0011% of reading + 11 μV from 0.1 V to 1 V 
    0.001% of reading + 50 μV from 1 V to 10 V 
    0.0014% of reading + 0.8 mV from 10 V to 100 V 
    0.002% of reading + 16 mV from 100 V to 1000 V 
    DC and low frequency electrical metrology - Electrical measurement and test equipment
    Ammeters;
    DC current;
    Comparison with a reference standard;
    0 μA to 10 A
    CAPABILITY
    0.004% of reading + 0.002 μA from 0 μA to 100 μA 
    0.003% of reading + 0.14 μA from 100 μA to 10 mA 
    0.005% of reading + 1.9 μA from 10 mA to 100 mA 
    0.014% of reading + 30 μA from 100 mA to 1 A 
    0.14% of reading + 0.75 mA from 1 A to 3A 
    0.18% of reading + 0.22 mA from 3 A to 10 A 


    AC current;
    Comparison with a reference standard;
    CAPABILITY
    0.6% of reading from 10 μA to 10 A at 10 Hz 
    0.3% of reading from 10 μA to 100 μA and from 10 Hz to 5 kHz 
    1.6% of reading from 10 μA to 100 μA and from 5 kHz to 10 kHz 
    0.1% of reading from 0.1 mA to 100 mA and from 10 Hz to 20 kHz 
    0.7% of reading from 0.1 mA to 100 mA and from 20 kHz to 30 kHz 
    0.2% of reading from 100 mA to 1 A and from 10 Hz to 5 kHz 
    0.5% of reading from 100 mA to 1 A and from 5 kHz to 20 kHz 
    1.6% of reading from 100 mA to 1 A and from 20 kHz to 30 kHz 
    0.3% of reading from 1 A to 10 A and from 10 Hz to 5 kHz 
    1.6% of reading from 1 A to 10 A and from 5 kHz to 10 kHz 

    Ohm meters;
    Resistance;
    Comparison with a reference standard;
    0 Ω to 100 MΩ
    CAPABILITY
    0.0033% of reading + 0.2 mΩ from 0 Ω to 10 Ω 
    0.003% of reading + 2 mΩ from 10 Ω to 100 Ω 
    0.002% of reading + 1.5 Ω from 100 Ω to 100 kΩ 
    0.003% of reading + 24 Ω from 100 kΩ to 1 MΩ 
    0.01% of reading + 730 Ω from 1 MΩ to 10 MΩ 
    0.085% of reading + 0.06 MΩ from 10 MΩ to 100 MΩ 

    Voltmeters;
    DC voltage;
    Comparison with a reference standard;
    0 V to 1000 V
    CAPABILITY
    0.0017% of reading + 1.4 μV from 0 V to 0.1 V 
    0.0011% of reading + 11 μV from 0.1 V to 1 V 
    0.001% of reading + 50 μV from 1 V to 10 V 
    0.0014% of reading + 0.8 mV from 10 V to 100 V 
    0.002% of reading + 16 mV from 100 V to 1000 V 


    AC voltage;
    Comparison with a reference standard;
    CAPABILITY
    0.15% + 34μV of reading at 0.01 mV to 10 mV and from 10 Hz to 20 kHz 
    0.25% + 70μV of reading at 0.01 mV to 10 mV and from 20 kHz to 50 kHz 
    1.1% + 110μV of reading at 0.01 mV to 10 mV and from 50 kHz to 100 kHz 
    0.02% of reading from 0.01 V to 10 V and from 10 Hz to 20 kHz 
    0.05% of reading from 0.01 V to 10 V and from 20 kHz to 50 kHz 
    0.12% of reading from 0.01 V to 10.0 V and from 50 kHz to 100 kHz 
    0.04% of reading from 10 V to 100 V at 10 Hz 
    0.03% of reading from 10 V to 100 V and from 10 Hz to 20 kHz 
    0.06% of reading from 10 V to 100 V and from 20 kHz to 50 kHz 
    0.18% of reading from 10 V to 100 V and from 50 kHz to 100 kHz 
    0.07% of reading from 100 V to 700 V at 40 Hz 
    0.09% of reading from 100 V to 700 V and from 40 Hz to 10 kHz 
    0.12 % of reading from 700 V to 1000 V and from 40 Hz to 10 kHz 
    DC and low frequency electrical metrology - Electrical standards
    Conductance boxes; Precision resistors; Resistance boxes;
    Resistance;
    Comparison with a reference standard;
    0 Ω to 100 MΩ
    CAPABILITY
    0.0033% of reading + 0.2 mΩ from 0 Ω to 10 Ω 
    0.003% of reading + 2 mΩ from 10 Ω to 100 Ω 
    0.002% of reading + 1.5 Ω from 100 Ω to 100 kΩ 
    0.003% of reading + 24 Ω from 100 kΩ to 1 MΩ 
    0.01% of reading + 730 Ω from 1 MΩ to 10 MΩ 
    0.085% of reading + 0.06 MΩ from 10 MΩ to 100 MΩ 
    Pressure metrology - Pressure and vacuum measuring equipment
    Digital pressure gauges; Mechanical pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges;
    Gauge pressure; High vacuum;
    Comparison with reference instrument;
     -95 to 120 MPa
    Electrical transducers with dc voltage, dc current or resistance outputs
    Including on site calibrations



    CAPABILITY
    Gauge pressure calibration
    With Calibration and Measurement Capability of-
    0.016% reading or 0.016 kPa (whichever is greater) in the range -95 to 100 kPa
    0.016% reading in the range 100 kPa to 120 MPa

    For on site calibrations
    Calibration and Measurement Capability of-
    0.13% reading from -95 kPa to 120 MPa



    Pressure metrology - Pressure standards
    Pressure calibrators - Non-dead weight;
    Gauge pressure;
    Comparison with reference instrument;
    Gauge pressure calibration in the range -95 to 120 MPa
    On-site gauge pressure calibration (industrial grade only) in the range -95 kPa to 120 MPa
    CAPABILITY
    Gauge pressure calibration in the range -95 to 100 kPa
    0.016% reading or 0.016 kPa (whichever is greater)  
    Gauge pressure calibration in the range 100 kPa to 120 MPa
    0.016% reading
    On-site gauge pressure calibration
    0.13% reading
    Time and frequency metrology - Frequency, time and waveform measuring equipment
    Counters;
    Frequency;

    1 Hz to  10 MHz 
    CAPABILITY
    0.045% of reading from 1 Hz to 40 Hz 
    0.015% of reading from 40 Hz to 10 MHz 

    Tachometers;
    Rotational speed;

     60 rpm to 90,000 rpm
    CAPABILITY
    0.045% of reading from 60 rpm to 2400 rpm 
    0.015% of reading from 2400 rpm to 90,000 rpm

    Stroboscopes;
    Frequency;

     1 Hz  to 1.5 kHz 
    CAPABILITY
    0.045% of reading from 1 Hz to 40 Hz 
    0.015% of reading from 40 Hz to 1.5 kHz 

    Time interval meters;
    Time interval;

    100 ns to 1 s 
    CAPABILITY
    0.015 % of reading from 100 ns to 1 s