SCOPE OF ACCREDITATION   View and Print Scope of Accreditation Printable version

Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.



ISO/IEC 17025 (2005)
Calibration
SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
DC and low frequency electrical metrology - Electrical instrument calibrators
Instrument calibrators;
AC current; AC voltage; DC current; DC voltage; Resistance;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of -
DC Voltage
0.0017% of reading + 1.4 μV from 0 V to 0.1 V
0.0011% of reading + 11 μV from 0.1 V to 1 V
0.001% of reading + 50 μV from 1 V to 10 V
0.0014% of reading + 0.8 mV from 10 V to 100 V
0.002% of reading + 16 mV from 100 V to 1000 V 
AC Voltage
0.15% + 34μV of reading at 0.01 mV to 10 mV and from 10 Hz to 20 kHz
0.25% + 70μV of reading at 0.01 mV to 10 mV and from 20 kHz to 50 kHz
1.1% + 110μV of reading at 0.01 mV to 10 mV and from 50 kHz to 100 kHz
0.02% of reading from 0.01 V to 10 V and from 10 Hz to 20 kHz
0.05% of reading from 0.01 V to 10 V and from 20 kHz to 50 kHz
0.12% of reading from 0.01 V to 10.0 V and from 50 kHz to 100 kHz
0.04% of reading from 10 V to 100 V at 10 Hz
0.03% of reading from 10 V to 100 V and from 10 Hz to 20 kHz
0.06% of reading from 10 V to 100 V and from 20 kHz to 50 kHz
0.18% of reading from 10 V to 100 V and from 50 kHz to 100 kHz
0.07% of reading from 100 V to 700 V at 40 Hz
0.09% of reading from 100 V to 700 V and from 40 Hz to 10 kHz
0.12 % of reading from 700 V to 1000 V and from 40 Hz to 10 kHz
DC Current
0.004% of reading + 0.002 μA from 0 μA to 100 μA
0.003% of reading + 0.14 μA from 100 μA to 10 mA
0.005% of reading + 1.9 μA from 10 mA to 100 mA
0.014% of reading + 30 μA from 100 mA to 1 A
0.14% of reading + 0.75 mA from 1 A to 3A
0.18% of reading + 0.22 mA from 3 A to 10 A
AC Current
0.6% of reading from 10 μA to 10 A at 10 Hz 
0.3% of reading from 10 μA to 100 μA and from 10 Hz to 5 kHz 
1.6% of reading from 10 μA to 100 μA and from 5 kHz to 10 kHz 
0.1% of reading from 0.1 mA to 100 mA and from 10 Hz to 20 kHz 
0.7% of reading from 0.1 mA to 100 mA and from 20 kHz to 30 kHz 
0.2% of reading from 100 mA to 1 A and from 10 Hz to 5 kHz 
0.5% of reading from 100 mA to 1 A and from 5 kHz to 20 kHz 
1.6% of reading from 100 mA to 1 A and from 20 kHz to 30 kHz 
0.3% of reading from 1 A to 10 A and from 10 Hz to 5 kHz 
1.6% of reading from 1 A to 10 A and from 5 kHz to 10 kHz 
Resistance
0.0033% of reading + 0.2 mΩ from 0 Ω to 10 Ω
0.003% of reading + 2 mΩ from 10 Ω to 100 Ω
0.002% of reading + 1.5 Ω from 100 Ω to 100 kΩ
0.003% of reading + 24 Ω from 100 kΩ to 1 MΩ
0.01% of reading + 730 Ω from 1 MΩ to 10 MΩ
0.085% of reading + 0.06 MΩ from 10 MΩ to 100 MΩ
DC and low frequency electrical metrology - Electrical measurement and test equipment
Ammeters; Digital multimeters (DMM); Ohm meters; Voltmeters;
AC current; AC voltage; DC current; DC voltage; Resistance;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of -
DC Voltage
0.0017% of reading + 1.4 μV from 0 V to 0.1 V
0.0011% of reading + 11 μV from 0.1 V to 1 V
0.001% of reading + 50 μV from 1 V to 10 V
0.0014% of reading + 0.8 mV from 10 V to 100 V
0.002% of reading + 16 mV from 100 V to 1000 V 
AC Voltage
0.15% + 34μV of reading at 0.01 mV to 10 mV and from 10 Hz to 20 kHz
0.25% + 70μV of reading at 0.01 mV to 10 mV and from 20 kHz to 50 kHz
1.1% + 110μV of reading at 0.01 mV to 10 mV and from 50 kHz to 100 kHz
0.02% of reading from 0.01 V to 10 V and from 10 Hz to 20 kHz
0.05% of reading from 0.01 V to 10 V and from 20 kHz to 50 kHz
0.12% of reading from 0.01 V to 10.0 V and from 50 kHz to 100 kHz
0.04% of reading from 10 V to 100 V at 10 Hz
0.03% of reading from 10 V to 100 V and from 10 Hz to 20 kHz
0.06% of reading from 10 V to 100 V and from 20 kHz to 50 kHz
0.18% of reading from 10 V to 100 V and from 50 kHz to 100 kHz
0.07% of reading from 100 V to 700 V at 40 Hz
0.09% of reading from 100 V to 700 V and from 40 Hz to 10 kHz
0.12 % of reading from 700 V to 1000 V and from 40 Hz to 10 kHz
DC Current
0.004% of reading + 0.002 μA from 0 μA to 100 μA
0.003% of reading + 0.14 μA from 100 μA to 10 mA
0.005% of reading + 1.9 μA from 10 mA to 100 mA
0.014% of reading + 30 μA from 100 mA to 1 A
0.14% of reading + 0.75 mA from 1 A to 3A
0.18% of reading + 0.22 mA from 3 A to 10 A
AC Current
0.6% of reading from 10 μA to 10 A at 10 Hz
0.3% of reading from 10 μA to 100 μA and from 10 Hz to 5 kHz
1.6% of reading from 10 μA to 100 μA and from 5 kHz to 10 kHz
0.1% of reading from 0.1 mA to 100 mA and from 10 Hz to 20 kHz
0.7% of reading from 0.1 mA to 100 mA and from 20 kHz to 30 kHz
0.2% of reading from 100 mA to 1 A and from 10 Hz to 5 kHz
0.5% of reading from 100 mA to 1 A and from 5 kHz to 20 kHz
1.6% of reading from 100 mA to 1 A and from 20 kHz to 30 kHz
0.3% of reading from 1 A to 10 A and from 10 Hz to 5 kHz
1.6% of reading from 1 A to 10 A and from 5 kHz to 10 kHz
Resistance
0.0033% of reading + 0.2 mΩ from 0 Ω to 10 Ω
0.003% of reading + 2 mΩ from 10 Ω to 100 Ω
0.002% of reading + 1.5 Ω from 100 Ω to 100 kΩ
0.003% of reading + 24 Ω from 100 kΩ to 1 MΩ
0.01% of reading + 730 Ω from 1 MΩ to 10 MΩ
0.085% of reading + 0.06 MΩ from 10 MΩ to 100 MΩ

DC and low frequency electrical metrology - Electrical standards
Conductance boxes; Precision resistors; Resistance boxes;
Resistance;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of -
0.0033% of reading + 0.2 mΩ from 0 Ω to 10 Ω 
0.003% of reading + 2 mΩ from 10 Ω to 100 Ω 
0.002% of reading + 1.5 Ω from 100 Ω to 100 kΩ 
0.003% of reading + 24 Ω from 100 kΩ to 1 MΩ 
0.01% of reading + 730 Ω from 1 MΩ to 10 MΩ 
0.085% of reading + 0.06 MΩ from 10 MΩ to 100 MΩ 
Pressure metrology - Pressure and vacuum measuring equipment
Digital pressure gauges; Mechanical pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges;
Gauge pressure; High vacuum;
Comparison with dead weight tester; Comparison with reference instrument;


Electrical transducers with dc voltage, dc current or resistance outputs
including on site calibrations



CAPABILITY
Gauge pressure calibration
with Calibration and Measurement Capability of-
0.016% reading or 0.016 kPa (whichever is greater) from  -95 kPa to 100 kPa
0.016% reading from  100 kPa to 120 MPa

On-site calibration
with Calibration and Measurement Capability of-
0.13% reading from -95 kPa to 120 MPa



Pressure metrology - Pressure standards
Pressure calibrators - Non-dead weight;
Gauge pressure;
Comparison with dead weight tester; Comparison with reference instrument;
CAPABILITY
with Calibration and Measurement Capability of -
0.016% reading or 0.016 kPa (whichever is greater) from -95 kPa to 100 kPa
0.016% of reading from 100 kPa to 120 MPa
Time and frequency metrology - Frequency, time and waveform measuring equipment
Counters; Stroboscopes; Tachometers; Time interval meters;
Frequency; Rotational speed; Time interval;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of -
Counters
0.045% of reading from 60 rpm to 2400 rpm 
0.015% of reading from 2400 rpm to 90,000 rpm
Tachometers
0.045% of reading from 60 rpm to 2400 rpm
0.015% of reading from 2400 rpm to 90,000 rpm
Stroboscopes
0.045% of reading from 1 Hz to 40 Hz
0.015% of reading from 40 Hz to 1.5 kHz
Time interval meters
0.015 % of reading from 100 ns to 1 s