Back to Search Results


Address Details:
Dock 2 (Building 2)
480 Weerona Road
LIDCOMBE, NSW 2141
AUSTRALIA Map
www.environment.nsw.gov.au

Contact Details:
Mr M Riley
+61(02) 99956302

Availability:
Services conditionally available to external clients

SCOPE OF ACCREDITATION   View and Print Scope of Accreditation Printable version

Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.



ISO/IEC 17025 (2017)
Calibration
SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
Chemical metrology - Dynamic gas blenders
Dynamic gas blenders;
Mass flow;
Absolute method;
in-house method CAL-001
5 to 200 cm3/min (at STP)  and 500 to 10,000 cm3/min (at STP) 
CAPABILITY
Calibration of gas diluters (mass flow controllers)
by in-house method CAL-001
in the ranges 5 to 200 cm3/min (at STP) and 500 to 10,000 cm3/min (at STP)
with calibration and measurement capability of -
0.3 % of full scale
Chemical metrology - Gas analysers
Gas analysers;
Concentration;
Direct measurement by certified reference material;
AS 3580 6.1 and in-house CAL 004 20 ppb to 520 ppb
CAPABILITY
With calibration and  measurement capability of-
1.4 ppb + 0.4 % of the reading  in the range of 20 ppb to 520 ppb for ozone
Flow metrology - Flow measuring devices and systems
Orifice meters;
Flow rate;
Direct measurement by critical flow venturi nozzle
;
in-house methods CAL-002 and CAL-006
CAPABILITY
 with calibration and measurement capability of -
 0.5 % of reading from 1 L/min to 20 L/min
Flow metrology - Environmental flow measuring devices and systems
Air sampling pumps and meters;
Flow rate;
Comparison with a reference meter; Volumetric transfer;
in-house method CAL-003
CAPABILITY
Calibration of piston meters
with calibration and measurement capability of -
0.3 % of reading from 1 L/min to 20 L/min

ISO/IEC 17025 (2017)
Environment
SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
Analysis for physical and chemical characteristics
Air - Ambient;
Particulate matter - PM10;
Tapered element oscillating microbalance (TEOM);
AS/NZS 3580.9.16 (TEOM with FDMS)


Particulate matter - PM10;
Tapered element oscillating microbalance (TEOM);
AS 3580 9.8


Ozone;
Direct reading instrument;
AS 3580 6.1


Particulate matter - PM2.5;
Beta attenuation monitors;
AS/NZS 3580 9.12 


Particulate matter - PM2.5; Particulate matter - PM10;
Tapered element oscillating microbalance (TEOM);
AS/NZS 3580.9.13 (Dichot TEOM with FDMS)


Sulfur dioxide;
Direct reading instrument;
AS 3580 4.1


Light scattering;
Nephelometry;
AS/NZS 3580 12.1


Nitrogen - Oxidised;
Calculation; Chemiluminescence;
AS 3580 5.1



Carbon monoxide;
Direct reading instrument;
AS 3580 7.1


Particulate matter - PM10;
Beta attenuation monitors;
AS/NZS 3580 9.11
Meteorological monitoring
Atmosphere;
Wind direction; Wind speed - Horizontal;
Ultrasonic anemometer;
AS/NZS 3580 14
Sample collection
Air - Ambient;
Not applicable;
Filter paper;
AS 3580 9.10 (Particulates PM2.5); AS/NZS 3580 1.1 (Site selection for ambient station)


Not applicable;
High volume air sampler (HVAS);
TSP using AS/NZS 3580.9.3