Back to Search Results


Address Details:
41 Enterprise Circuit
PRESTONS, NSW 2170
AUSTRALIA Map
www.trcalibration.com.au

Contact Details:
Mr Mario Cafarelli
+61(02) 87122002

Availability:
Services available to external clients

SCOPE OF ACCREDITATION   View and Print Scope of Accreditation Printable version

Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.



ISO/IEC 17025 (2005)
Calibration
SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
DC and low frequency electrical metrology - Electrical instrument calibrators
Instrument calibrators;
Resistance;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
10 μΩ/Ω from 1 mΩ to 100 mΩ
4 μΩ/Ω above 100 mΩ to 10 Ω
1.5 μΩ/Ω above 10 Ω to 10 kΩ
2.5 μΩ/Ω above 10 kΩ to 100 kΩ
4 μΩ/Ω above 100 kΩ to 1 MΩ
40 μΩ/Ω above 1 MΩ to 100 MΩ
0.17% above 100 MΩ to 10 GΩ
0.27% above 10 GΩ to 1 TΩ


AC voltage;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of -
275 μV/V at 2 mV at 20 Hz
250 μV/V at 2 mV from 40 Hz to 50 kHz
260 μV/V at 2 mV at 100 kHz
290 μV/V at 2 mV at 200 kHz
320 μV/V at 2 mV at 300 kHz
520 μV/V at 2 mV at 500 kHz
140 μV/V from 10 mV to 30 mV at 10 Hz
100 μV/V from 10 mV to 30 mV at 20 Hz
80 μV/V from 10 mV to 30 mV from 40 Hz to 50 kHz
90 μV/V from 10 mV to 30 mV at 100 kHz
130 μV/V from 10 mV to 30 mV at 200 kHz
165 μV/V from 10 mV to 30 mV at 300 kHz
350 μV/V from 10 mV to 30 mV at 500 kHz
1200 μV/V from 10 mV to 30 mV at 1 MHz
55 μV/V above 30 mV to 300 mV at 10 Hz
45 μV/V above 30 mV to 300 mV at 20 Hz
55 μV/V above 30 mV to 300 mV from 40 Hz to 100 kHz
95 μV/V above 30 mV to 300 mV at 200 kHz
150 μV/V above 30 mV to 300 mV at 300 kHz
350 μV/V above 30 mV to 300 mV at 500kHz
1330 μV/V above 30 mV to 300 mV at 1 MHz
55 μV/V above 300 mV to 200 V at 10 Hz
50 μV/V above 300 mV to 300 V from 20 Hz to 50 kHz
45 μV/V above 300 mV to 700m V from 100 kHz to 300 kHz
65 μV/V above 300 mV to 700 mV at 500 kHz
100 μV/V above 300 mV to 700 mV at 1 MHz
60 μV/V above 700 mV to 2.2 V at 100 kHz
290 μV/V above 700 mV to 2.2 V at 200 kHz
350 μV/V above 700 mV to 2.2 V at 300 kHz
1800 μV/V above 700 mV to 2.2 V at 500 kHz
5300 μV/V above 700 mV to 2.2 V at 1 MHz
50 μV/V above 2.2 V to 7 V at 100 kHz
70 μV/V above 2.2 V to 7 V at 200 kHz
75 μV/V above 2.2 V to 7 V at 300 kHz
700 μV/V above 2.2 V to 7 V and from 500 kHz to 1 MHz
50 μV/V above 7 V to 300 V at 100 kHz
65 μV/V above 7 V to 100 V at 200 kHz
70 μV/V above 7 V to 60 V and from 300 kHz to 500 kHz
75 μV/V above 7 V to 22 V at 1 MHz
50 μV/V above 300 V to 1000 V at 40 Hz to 50 kHz
75 μV/V above 300 V to 1000 V at 100 kHz
700 μV/V from 0.2 V to 3 V from 1 MHz to less than 10 MHz
1000 μV/V from 0.2 V to 3 V from 10 MHz to less than 30 MHz
1900 μV/V from 0.2 V to 3 V at 30 MHz


DC current;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
150 pA from 20 pA to 10 μA
15 μA/A above 10 μA to 20 A
65 μA/A above 20 A to 100 A 


AC current;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
290 μA/A from 100 μA to 200 μA from 10 Hz to <1 kHz
310 μA/A from 100 μA to 200 μA at 1 kHz
335 μA/A from 100 μA to 200 μA at 5 kHz
380 μA/A from 100 μA to 200 μA at 10 kHz
285 μA/A above 200 μA to 2 mA from 10 Hz to <1 kHz
290 μA/A above 200 μA to 2 mA at 1 kHz
320 μA/A above 200 μA to 2 mA at 5 kHz
355 μA/A above 200 μA to 2 mA at 10 kHz
70 μA/A above 2 mA to 20 A from 10 Hz to 1 kHz
190 μA/A above 2 mA to 20 A at 5 kHz
580 μA/A above 2 mA to 20 A at 10 kHz
170 μA/A above 20 A to 100 A from 10 Hz to 1 kHz
580 μA/A from 10 mA to 300 mA at 30 kHz


DC voltage;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
0.4 μV up to 100 mV
2.5 μV/V at 1 V and 10 V
4.0 μV/V from 100 mV to 100 V
8.5 μV/V above 100 V to 1100 V
0.2% above 1100 V to 10000 V
DC and low frequency electrical metrology - Electrical measurement and test equipment
LCR meters;
Capacitance;
Comparison with a reference standard;
CAPABILITY
from 50 Hz to 1 kHz
with Calibration and Measurement Capability of - 
0.03% + 0.1 pF from 1 pF to 1 μF
0.045% from 1 μF to 110 μF


Resistance;
Comparison with a reference standard;
CAPABILITY
D.C. Resistance
with Calibration and Measurement Capability of - 
0.035% at 1 Ω
0.02% from 1 Ω to 1 MΩ
0.035% from 1 MΩ to 2 MΩ


Inductance;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
Inductance at 1 kHz
0.05% from 100 μH to 100 mH
0.13% at 10 H 

Galvano meters and null detectors;
AC voltage; DC voltage;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
0.1% + 0.1 μV from 0 V to 1 mV

Current clamps;
AC current;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
0.2% from 1 mA to 1000 A, 50 Hz to 1 kHz
0.64% from 1000 A to 5000 A, 50 Hz to 400 Hz

Appliance testers; Capacitance meters; Data recorders; Digital multimeters (DMM); Galvano meters and null detectors; Insulation resistance test equipment; LCR meters; Ohm meters; Phase angle meters; Power analysers; Power supplies; Voltmeters; Watt meters;
AC voltage;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of -
310 μV/V at 2 mV from 20 Hz to 50 kHz
300 μV/V at 2 mV at 100 kHz
450 μV/V at 2 mV from 200 kHz to 500 kHz
165 μV/V from 10 mV to 30 mV at 10 Hz
140 μV/V from 10 mV to 30 mV at 20 Hz
125 μV/V from 10 mV to 30 mV from 40 Hz to 50 kHz
130 μV/V from 10 mV to 30 mV from 100 kHz to 300 kHz
140 μV/V from 10 mV to 30 mV at 300 kHz
160 μV/V from 10 mV to 30 mV at 500 kHz
200 μV/V from 10 mV to 30 mV at 1 MHz
75 μV/V above 30 mV to 300 mV at 10 Hz
70 μV/V above 30 mV to 300 mV from 20 Hz to 100 kHz
75 μV/V above 30 mV to 300 mV at 200 kHz
85 μV/V above 30 mV to 300 mV at 300 kHz
100 μV/V above 30 mV to 300 mV at 500 kHz
205 μV/V above 30 mV to 300 mV at 1 MHz
55 μV/V above 300 mV to 200 V at 10 Hz
50 μV/V from 30 mV to 300 V from 20 Hz to 50 kHz
50 μV/V from 30 mV to 700 mV from 100 kHz to 500 kHz
110 μV/V from 30 mV to 700 mV at 1 MHz
45 μV/V above 700 mV to 2.2 V from 100 kHz to 200 kHz
50 μV/V above 700 mV to 2.2 V at 300 kHz
100 μV/V above 700 mV to 2.2 V at 500 kHz
110 μV/V above 700 mV to 2.2 V at 1 MHz
50 μV/V above 2.2 V to 7 V from 100 kHz to 300 kHz
100 μV/V above 2.2 V to 7 V at 500 kHz
110 μV/V above 2.2 V to 7 V at 1 MHz
50 μV/V above 7 V to 300 V at 100 kHz
50 μV/V above 7 V to 100 V at 200 kHz
70 μV/V above 7 V to 60 V at 300 kHz
120 μV/V above 7 V to 30 V at 500 kHz
120 μV/V above 7 V to 22 V at 1 MHz
50 μV/V above 300 V to 1000 V from 40 Hz to 50 kHz
75 μV/V above 300 V to 1000 V at 100 kHz 

Data recorders; Power analysers; Power supplies; Watt meters;
Apparent power;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
0.1% from 0.1 V up to 1000 V and from 0.001 A to 20 A on DC (50 Hz at unity power factor)

Data recorders; Digital multimeters (DMM); Galvano meters and null detectors; Insulation resistance test equipment; Power analysers; Power supplies; Residual current circuit breaker testers; Voltmeters; Watt meters;
DC voltage;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
2.4 μV/V at 1.018 V and 10 V
3.7 μV/V + 0.1 μ V from 1 mV to 100 V
6.4 μV/V above 100 to 1100 V

Data recorders; Digital multimeters (DMM); Ohm meters;
Resistance;
Comparison with a reference standard;
Including megohmmeters up to 1 TΩ and up to 5 000 V dc applied voltage


CAPABILITY
with Calibration and Measurement Capability of - 
100 μΩ/Ω from 10 μΩ to 1 mΩ
10 μΩ/Ω from 1 to 100 mΩ
3 μΩ/Ω from 100 mΩ to 1 Ω
1 μΩ/Ω from 1 Ω to 10 kΩ
2 μΩ/Ω from 10 to 100 kΩ
3 μΩ/Ω from 100 kΩ to 1 MΩ
50 μΩ/Ω from 1 to 100 MΩ
0.2% from 100 MΩ to 10 GΩ
0.3% from 10 GΩ to 1 TΩ

Ammeters; Appliance testers; Current clamps; Data recorders; Digital multimeters (DMM); Galvano meters and null detectors; Insulation resistance test equipment; Ohm meters; Phase angle meters; Power analysers; Power supplies; RCD testers; Residual current circuit breaker testers; Watt meters;
AC current;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
165 μA/A from 100 μA to 200 μA from 10 Hz to <1 kHz
380 μA/A from 100 μA to 200 μA at 1 kHz
750 μA/A from 100 μA to 200 μA at 5 kHz
1000 μA/A from 100 μA to 200 μA at 10 kHz
185 μA/A above 200 μA to 2 mA from 10 Hz to <1 kHz
165 μA/A above 200 μA to 2 mA at 1 kHz
425 μA/A above 200 μA to 2 mA at 5 kHz
750 μA/A above 200 μA to 2 mA at 10 kHz
90 μA/A above 2 mA to 20 A from 10 Hz to <1 kHz
120 μA/A above 2 mA to 20 A at 1 kHz
450 μA/A above 2 mA to 20 A at 5 kHz
940 μA/A above 2 mA to 20 A at 10 kHz
2600 μA/A from 10 mA to 300 mA at 30 kHz

Ammeters; Digital multimeters (DMM); Power supplies; Watt meters;
DC current;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
4% + 0.1 fA up to 20 pA
1.5% above 20 pA to 200 pA
5000 μA/A above 200 pA to 20 nA
2000 μA/A above 20 nA to 10 μA
14 μA/A above 10 μA to 1 A
23 μA/A above 1 A to 10 A
45 μA/A above 10 A to 100 A
200 μA/A above 100 A to 200 A
DC and low frequency electrical metrology - Electrical standards
Voltage standards - Standard cells;
Voltage; Voltage AC-DC difference; Voltage ratio;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
1.5 μV at 1.018 V and 1.019 V

Current shunts;
Current;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
DC shunts
with Calibration and Measurement Capability of -
5 μΩ/Ω from 10 μA to 100 mA
9 μΩ/Ω above 100 mA to 1 A
20 μΩ/Ω above 1 A to 10 A
30 μΩ/Ω above 10 A to 100 A
AC shunts from 100 μA to 20 A
with Calibration and Measurement Capability of -
60 μΩ/Ω at 57 Hz
110 μΩ/Ω at 1 kHz
720 μΩ/Ω at 5 kHz and 10 kHz

Voltage standards - E.M.F. reference devices;
Voltage; Voltage AC-DC difference; Voltage ratio;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
1.5 μV/V at 1.0, 1.018 and 10 V

Conductance boxes; Precision resistors; Resistance boxes;
Resistance;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
100 μΩ/Ω from 10 μΩ to 1 mΩ
10 μΩ/Ω above 1 to 100 mΩ
3 μΩ/Ω above  100 mΩ to 1 Ω
1 μΩ/Ω above  1 Ω to 10 kΩ
2 μΩ/Ω above  10 to 100 kΩ
3 μΩ/Ω above  100 kΩ to 1 MΩ
50 μΩ/Ω above  1 to 100 MΩ
0.2% above  100 MΩ to 10 GΩ
0.3% above 10 GΩ to 1 TΩ

Potential dividers; Voltage ratio boxes;
Voltage ratio;
Comparison with a reference standard;
CAPABILITY
Volt ratio boxes with fixed ratios
with Calibration and Measurement Capability of -
2 μV/V of input from 0.1 V to 500 V
4 μV/V of input above 500 V to 1100 V


Variable ratio potential dividers from 10⁻⁷ to 1.1
with Calibration and Measurement Capability of - 
2 μV/V of input from 0.1 V to  500 V
4 μV/V of input above 500 V to 1100 V

AC/DC transfer instruments;
Voltage;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
715 μV/V at 2 mV from 20 Hz to 30 Hz
510 μV/V at 2 mV from 40 Hz to 50 kHz
1100 μV/V at 2 mV and at 100 kHz
1250 μV/V at 2 mV and at 200 kHz
1550 μV/V at 2 mV and at 300 kHz
1550 μV/V at 2 mV and at 500 kHz
2100 μV/V at 2 mV and at 1 MHz
210 μV/V from 2 mV to 10 mV and 40 Hz to 50 kHz
510 μV/V from 2 mV to 10 mV at 100 kHz
620 μV/V from 2 mV to 10 mV at 200 kHz
820 μV/V from 2 mV to 100 mV and 300 kHz to 500 kHz
1000 μV/V from 2 mV to 600 mV at 1 MHz
820 μV/V at 10 mV and at 10 Hz
310 μV/V above 10 mV to 600 mV at 10 Hz
620 μV/V from 10 mV to 300 mV and 20 Hz to 30 Hz
100 μV/V above 300 mV to 600 mV and 20 Hz to 30 Hz
85 μV/V from 10 mV to 100 mV and 40 Hz to 50 kHz
50 μV/V above 100 mV to 600 mV and 40 Hz to 50 kHz
410 μV/V from 10 mV to 100 mV and 100 kHz to 200 kHz
205 μV/V above 100 mV to 600 mV and 100 kHz to 200 kHz
205 μV/V from 100 mV to 600 mV and 300 kHz to 500 kHz
310 μV/V from 0.6 V to 1000 V and at 10 Hz
100 μV/V from 0.6 V to 1000 V from 20 Hz to 40 Hz
20 μV/V from 0.6 V to 400 V and 50 Hz to 10 kHz
20 μV/V from 0.6 V to 60 V and at 20 kHz
30 μV/V from 0.6 V to 400 V and at 50 kHz
45 μV/V from 0.6 V to 60 V and at 100 kHz
175 μV/V from 0.6 V to 60 V and 200 kHz to 500 kHz
205 μV/V from 0.6 V to 20 V and 500 kHz to 1 MHz
25 μV/V above 60 V to 400 V and at 20 kHz
55 μV/V from 60 V to 400 V and at 100 kHz
30 μV/V from 400 V to 1000 V and at 50 Hz
40 μV/V from 400 V to 1000 V and 400 Hz to 20 kHz
85 μV/V from 400 V to 1000 V and at 50 kHz
125 μV/V from 400 V to 1000 V and at 100 kHz
600 μV/V from 0.2 V to 3 V and at 3 MHz
900 μV/V from 0.2 V to 3 V and at 10 MHz
1500 μV/V from 0.2 V to 3 V and at 30 MHz
3000 μV/V from 0.2 V to 3 V and at 50 MHz
9000 μV/V from 0.2 V  to 3 V and at 100 MHz
3000 μV/V at 0.27 V and at 70 MHz


Current;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
50 μA/A from 100 μA to 20 A at 57 Hz
85 μA/A from 100 μA to 20 A and 100 Hz to 1 kHz
750 μA/A from 100 μA to 20 A and above 1 kHz to 10 kHz
300 μA/A from 10 mA to 300 mA at 30 kHz

Inductors;
Inductance;
Comparison with a reference standard;
CAPABILITY
Two terminal inductors at 1 kHz 
with Calibration and Measurement Capability of - 
0.1% from 0.1 mH to 1 mH
0.05% above 1 mH to 100 mH
0.13% at 10 H

Precision capacitors;
Capacitance;
Comparison with a reference standard;
CAPABILITY
Three terminal capacitors from 40 Hz up to 1 kHz
with Calibration and Measurement Capability of -
30 μF/F + 0.1 fF from 0.1 pF up to 1000 pF
0.01% from 1000 pF to 1.1 μF
Two terminal capacitors from 50 Hz to 1 kHz
with Calibration and Measurement Capability of -
(0.1% + 0.02 pF) from 1 pF to 110 μF
DC and low frequency electrical metrology - High-voltage/high-current standards and equipment
High voltage measurement systems;
Voltage;
Comparison with a reference standard;
CAPABILITY
with Calibration and Measurement Capability of -
2500 μV/V from 1.1 kV to 10 kV
High frequency electrical metrology - Communications, electromagnetic field strength and EMC test equipment
Amplifiers; Artificial mains networks (AMN); Attenuators; Coupling/decoupling networks (CDN); Data transmission measuring equipment; Electrical fast transients (EFT) and surge generators; Electrical noise and interference measuring equipment; Flicker harmonics load unit; Harmonic emission compliance test systems; Impedance and reflection measuring equipment; Line impedance stabilisation networks (LISN); Line transmission measuring equipment; Power measuring equipment; Radio transmission measuring equipment; Short interruptions and voltage variations; Spectrum analysers; Voltage dips; Voltage fluctuation and flicker compliance test systems; Waveguide and coaxial components;
Reflection coefficient;
Comparison with a reference standard; Direct measurement by electrical input;
CAPABILITY
with Calibration and Measurement Capability of -
0.01 for reflection coefficient

Amplifiers; Artificial mains networks (AMN); Coupling/decoupling networks (CDN); Data transmission measuring equipment; Electrical fast transients (EFT) and surge generators; Electrical noise and interference measuring equipment; Flicker harmonics load unit; Harmonic emission compliance test systems; Impedance and reflection measuring equipment; Line impedance stabilisation networks (LISN); Line transmission measuring equipment; Power measuring equipment; Radio transmission measuring equipment; Spectrum analysers; Voltage dips; Voltage fluctuation and flicker compliance test systems; Waveguide and coaxial components;
Attenuation;
Comparison with a reference standard; Direct measurement by electrical input;
CAPABILITY
with Calibration and Measurement Capability of - 
0.003 dB + 0.0005 dB/ 10 dB from 0 to 60 dB at D.C and 10 Hz to 10 kHz
0.04 dB + 0.005 dB/ 10 dB above 60 dB to 110 dB at D.C. and 10 Hz to 10 kHz
0.007 dB + 0.001 dB/ 10 dB from 0 to 70 dB at 10 kHz to 30 MHz
0.014 dB + 0.02 dB/ 10 dB above 70 dB to 120 dB at 10 kHz to 30 MHz
0.02 dB + 0.02 dB/ 10 dB from 0 to 110 dB at 30 MHz to 18 GHz
0.24 dB + 0.3 dB/ 10 dB above 110 dB to 120 dB at 30 MHz to 18 GHz

Waveguide and coaxial components;
Insertion loss;
Comparison with a reference standard; Direct measurement by electrical input;
CAPABILITY
with Calibration and Measurement Capability of -
0.005 dB at 8.2 GHz to 18 GHz ("X" and "P" band waveguides) coupling factor, directivity and insertion loss of directional couplers

Power measuring equipment; Spectrum analysers;
Power;
Comparison with a reference standard; Direct measurement by electrical input;
CAPABILITY
with Calibration and Measurement Capability of - 
0.65% from 1 μW to 100 mW from D.C. to 30 MHz;
0.6% from 1 μW to 2 W from 30 to 500 MHz;
0.6% + 0.07% per GHz from 1 μW to 2 W from 500 MHz to 18 GHz (50 Ω coaxial);
1.5% + 0.07% per GHz from 1 μW to 2 W from 8.2 GHz to 18 GHz ("X" and "P" band waveguides);
10% from 2 W to 50 W from 30 MHz to 250 MHz
4.5% from 2 W to 15 W from 800 MHz to 2.5 GHz 
Pressure metrology - Pressure and vacuum measuring equipment
Pressure gauges; Pressure recorders; Pressure transducers; Vacuum gauges;
Gauge pressure; High vacuum;
By pressure calibrator; Comparison with dead weight tester; Comparison with reference instrument;
MSA Test Method 1;
MSA Test Method 2
CAPABILITY
Pneumatic devices
with Calibration and Measurement Capability of -
0.03 kPa from -100 kPa to 40 kPa
0.04% + 0.001 kPa above 40 kPa to 200 kPa
0.1 kPa above 200 kPa to 800 kPa
0.015% above 800 kPa to 83.5 MPa

Hydraulic devices
with Calibration and Measurement Capability of -
0.07 kPa from 45 kPa to 300 kPa;
0.01% above 300 kPa to 83.5 MPa
Temperature metrology - Ancillary temperature measuring equipment
Multi-channel thermocouple data recorders; Process calibrators;
Resistance to temperature conversion; Volt to temperature conversion;
Direct measurement by electrical input;
CAPABILITY
with Calibration and Measurement Capability of -
0.25% of range plus 2 digits from -270 °C to 1370 °C
Temperature metrology - Temperature measuring equipment
Base metal thermocouples;
Temperature;
Measurement against reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
0.3 °C + 0.1%T from -50 °C to 300 °C
0.6 °C + 0.1%T above 300 °C to 500 °C
1 °C above 500 °C to 650 °C 

Digital temperature measuring systems;
Temperature;
Measurement against reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
0.3 °C + 0.1%T from -50 °C to 300 °C
0.6 °C + 0.1%T above 300 °C to 500 °C
1 °C above 500 °C to 650 °C 
Temperature metrology - Temperature standards and reference equipment
Dry block calibrators;
Temperature;
Direct measurement using a reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
0.05 °C at 0 °C
0.15 °C + 0.05%T from -50 °C to 300 °C
0.35 °C + 0.05%T above 300 °C to 650 °C 
Time and frequency metrology - Frequency and time standards
Frequency standards;
Frequency;
Measurement against reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
1 in 1011 up to 10 MHz

Signal sources;
Voltage;
Measurement against reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
275 μV/V at 2 mV at 20 Hz
250 μV/V at 2 mV from 40 Hz to 50 kHz
260 μV/V at 2 mV at 100 kHz
290 μV/V at 2 mV at 200 kHz
320 μV/V at 2 mV at 300 kHz
520 μV/V at 2 mV at 500 kHz
140 μV/V from 10 mV to 30 mV at 10 Hz
100 μV/V from 10 mV to 30 mV at 20 Hz
80 μV/V from 10 mV to 30 mV from 40 Hz to 50 kHz
90 μV/V from 10 mV to 30 mV at 100 kHz
130 μV/V from 10 mV to 30 mV at 200 kHz
165 μV/V from 10 mV to 30 mV at 300 kHz
350 μV/V from 10 mV to 30 mV at 500 kHz
1200 μV/V from 10 mV to 30 mV at 1 MHz
55 μV/V above 30 mV to 300 mV at 10 Hz
45 μV/V above  30 mV to 300 mV at 20 Hz
55 μV/V above 30 mV to 300 mV from 40 Hz to 100 kHz
95 μV/V above 30 mV to 300 mV at 200 kHz
150 μV/V above 30 mV to 300 mV at 300 kHz
350 μV/V above 30 mV to 300 mV at 500 kHz
1330 μV/V above 30 mV to 300 mV at 1 MHz
55 μV/V above 300 mV to 200 V at 10 Hz
50 μV/V above 30 mV to 300 V from 20 Hz to 50 kHz
45 μV/V above 30 mV to 700 mV from 100 kHz to 300 kHz
65 μV/V above 30 mV to 700 mV at 500 kHz
100 μV/V above 30 mV to 700 mV at 1 MHz
60 μV/V above 700 mV to 2.2 V at 100 kHz
290 μV/V above 700 mV to 2.2 V at 200 kHz
350 μV/V above 700 mV to 2.2 V at 300 kHz
1800 μV/V above 700 mV to 2.2 V at 500 kHz
5300 μV/V above 700 mV to 2.2 V at 1 MHz
50 μV/V above 2.2 V to 7 V at 100 kHz
70 μV/V above 2.2 V to 7 V at 200 kHz
75 μV/V above 2.2 V to 7 V at 300 kHz
700 μV/V above 2.2 V to 7 V from 500 kHz to 1 MHz
50 μV/V above 7 V to 300 V at 100 kHz
65 μV/V above 7 V to 100 V at 200 kHz
70 μV/V above 7 V to 60 V from 300 kHz to 500 kHz
75 μV/V above 7 V to 22 V at 1 MHz
50 μV/V from 300 V to 1000 V at 40 Hz to 50 kHz
75 μV/V from 300 V to 1000 V at 100 kHz
700 μV/V from 0.2 V to 3 V from 1 MHz to less than 10 MHz
1000 μV/V from 0.2 V to 3 V from 10 MHz to less than 30 MHz
1900 μV/V from 0.2 V to 3 V at 30 MHz


Frequency;
Measurement against reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
5 in 1011 up to 10 MHz;
5 in 1010 above 10 MHz to 18 GHz


Modulation;
Measurement against reference standard;
CAPABILITY
Amplitude and frequency modulation
with Calibration and Measurement Capability of - 
1% from 1 MHz to 18 GHz
Time and frequency metrology - Frequency, time and waveform measuring equipment
Frequency meters;
Frequency;
Measurement against reference standard;
CAPABILITY
with Calibration and Measurement Capability of -
1 in 107 from 0.1 MHz to 18 GHz

Clocks and timers;
Time interval;
Measurement against reference standard;
CAPABILITY
Stopwatches
with Calibration and Measurement Capability of -
0.040 s from 1 s to 30 min
0.30 s from 30 min to 3 h
Stop clock
with Calibration and Measurement Capability of -
0.025 s from 0.5 s to 1 h

Frequency analysers; Oscilloscopes; Waveform measuring instruments;
Rise/fall time;
Measurement against reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
9 ps from a minimum of 25 ps


Distortion;
Measurement against reference standard;
CAPABILITY
with an input voltage range 50 mV to 300 V and 10 Hz to 1 MHz
with Calibration and Measurement Capability of - 
0.1 dB from -10 dB to -60 dB
0.3 dB from -60 dB to -80 dB

Frequency analysers; Oscilloscopes; Time interval meters; Waveform measuring instruments;
Time interval;
Measurement against reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
5 in 1011 + 5 ns from 10 ns to 20 000 s

Counters; Frequency analysers; Oscilloscopes; Stroboscopes; Waveform measuring instruments;
Frequency;
Measurement against reference standard;
CAPABILITY
with Calibration and Measurement Capability of - 
5 in 1011 up to 10 MHz;
5 in 1010 from 10 MHz to 18 GHz

Tachometers;
Rotational speed;
Comparison with a reference standard;
CAPABILITY
Optical tachometers
with Calibration and Measurement Capability of - 
17 μHz from 0.017 Hz to 1.67 Hz
(0.001 rpm from 1 rpm to 100 rpm)
10 μHz/Hz from 1.67 Hz to 16666.7 Hz
(0.001% from 100 rpm to 1000000 rpm)

ISO/IEC 17025 (2005)
Manufactured Goods
SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
Evaluation of electrical appliances and devices - Performance
External power supplies, stabilisers and related items;
Drift; Electrical parameters; Frequency and time interval; Input impedance; Insulation resistance; Line regulation; Load regulation; Meter accuracy; Power; Power factor; Recovery time - Transient response; Ripple - Noise;
Not applicable;
Performance tests on power supplies by the methods of Calibration Procedure CP910125 or Manufacturers Performance Verification
A.C/D.C power, 
A.C/D.C voltage
A.C/D.C current