SCOPE OF ACCREDITATION   View and Print Scope of Accreditation Printable version

Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.



ISO/IEC 17025 (2017)
Calibration
SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
Flow metrology - Flow measuring devices and systems
Air sampling pumps and meters;
Flow rate;
Comparison with a reference meter;
EP0140 Flow rate: 1 L/min to 3 L/min
Timer: minimum 2 hours
CAPABILITY
Calibration of air sampling pumps using air at atmospheric conditions
with Calibration and Measurement Capability of -
Sampling pump flow rate: 1.2% of flow from 1 L/min to 3 L/min
Sampling pump timer: (0.05% reading + 0.5 minutes) over a minimum of 2 hours

Temperature metrology - Verification of controlled enclosures
Baths; Furnaces; Ovens;
Spatial uniformity; Temperature; Temporal uniformity;
Direct temperature measurement;
AS 2853- 1986
From -40 ºC to 850 ºC
 
 
CAPABILITY
Including on site calibrations.
With Calibration and Measurement Capability of -
± 1.3 ºC  from -40 ºC to 300 ºC
± 5.4 ºC  above 300 ºC to 850 ºC

Environmental chambers - Humidity;
Fluctuation; Gradient; Humidity;
By temperature and humidity measurement;
AS 60068.3.6; IEC 60068.3.6

for humidity, relative humidity gradient tests, relative humidity fluctuation tests
CAPABILITY
Including on site calibrations.
With Calibration and Measurement Capability of -
Humidity and Relative humidity gradient tests
3.0 %RH in the range 10% to 95%RH  and from 23 °C to 55 °C
5.0 %RH in the range 10% to 93% RH and above 55 °C to 80 °C

Environmental chambers - Temperature;
Spatial uniformity; Temperature; Temporal uniformity;
Direct temperature measurement;
AS 60068.3.5, IEC 60068.3.5
From -40 ºC to 300 ºC
 
 
CAPABILITY
Including on site calibrations.
With Calibration and Measurement Capability of -
± 1.3 ºC from -40 ºC to 300 ºC