SCOPE OF ACCREDITATION   View and Print Scope of Accreditation Printable version

Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.



ISO/IEC 17025 (2017)
Calibration
SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
Flow metrology - Environmental flow measuring devices and systems
Air sampling pumps and meters;
Flow rate;
Comparison with a reference meter;
EP0140
CAPABILITY

Calibration of air sampling pumps using air at atmospheric conditions
with Calibration and Measurement Capability of -
Sampling pump flow rate: 1.2% of flow from 1 L/min to 3 L/min
Sampling pump timer: (0.05% reading + 0.5 minutes) over a minimum of 2 hours

Temperature metrology - Verification of controlled enclosures
Baths; Furnaces; Ovens;
Spatial uniformity; Temperature; Temporal uniformity;
Direct temperature measurement;
AS 2853- 1986
From -40 °C to 850 °C
 
 
CAPABILITY
Including on site calibrations.
With Calibration and Measurement Capability of -
± 1.3 °C  from -40 °C to 300 °C
± 5.4 °C  above 300 °C to 850 °C

Environmental chambers - Humidity;
Fluctuation; Gradient; Humidity;
By temperature and humidity measurement;
AS 60068.3.6; IEC 60068.3.6

for humidity, relative humidity gradient tests, relative humidity fluctuation tests
CAPABILITY
Including on site calibrations.
With Calibration and Measurement Capability of -
Humidity and Relative humidity gradient tests
3.0 %RH in the range 10% to 95%RH  and from 23 °C to 55 °C
5.0 %RH in the range 10% to 93% RH and above 55 °C to 80 °C

Environmental chambers - Temperature;
Spatial uniformity; Temperature; Temporal uniformity;
Direct temperature measurement;
AS 60068.3.5, IEC 60068.3.5
From -40 °C to 300 °C
 
 
CAPABILITY
Including on site calibrations.
With Calibration and Measurement Capability of -
± 1.3 °C from -40 °C to 300 °C