ISO/IEC 17025 (2017)
Calibration
SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
Dimensional metrology - Engineering equipment and precision instruments
Centrifuges;
Rotational speed;
Comparison with a reference standard by differential measurement;
In-house method CY 20821
CAPABILITY
With Calibration and measurement capability of -
1.2 rpm from 10 to 900 rpm
2 rpm above 900 to 5000 rpm
0.02% of reading or 3 rpm (whichever is greater) above 5000 to 99 900 rpm
Mass - Determination of mass and calibration of weighing devices
Mass standards;
Mass;
Gravimetric measurement against reference mass;
CAPABILITY
With Calibration and measurement capability of-
3 μg from 1mg to 20 mg
4 μg from above 20 mg up to 50 mg 
5 μg from above 50 mg up to 500 mg 
6 μg from above 500 mg up to 1 g
7 μg from above 1 g up to 2 g
10 μg from above 2 g up to 5 g
25 μg from above 5 g up to 10 g
30 μg above 10 g up to 20 g 
40 μg above 20 g up to 50 g 
60 μg from above 50 g up to 100 g 
0.11 mg from above 100 g up to 200 g 
0.3 mg from above 200 g up to 500 g 
0.6 mg from above 500 g up to 1 kg 
2.5 mg from above 1 kg up to 2 kg 
20 mg from above 2 kg up to 10 kg 
0.2 g from above 10 kg up to 20 kg 
20 g from above 20 kg up to 30 kg
Mass - Determination of mass and calibration of weighing devices
Laboratory weighing devices; Precision laboratory balances;
Mass;
Gravimetric measurement against reference mass;
including on site calibrations



CAPABILITY
Including on site calibrations
With Calibration and measurement capability of-
1 in 106 or 4 ug (whichever is greater)  up to 0.5 g
6 in 106 above 0.5 g up to 1 g
4 in 106 above 1 g up to 2 g
2 in 106 above 2g up to 5g
3 in 106 above 5 g up to 10 g
2 in 106 above 10 g up to 20 g
1 in 106 above 20 g up to 1 kg
4 in 106 above 1 kg up to 10 kg
1 in 105 above 10 kg up to 20 kg
Mass - Determination of mass and calibration of weighing devices
Industrial weighing devices;
Mass;
Gravimetric measurement against reference mass;
including on site calibrations



CAPABILITY
Including on site calibrations
With Calibration and measurement capability of-
1 in 10âμ or 1 mg (whichever is greater) up to 30 kg
Temperature metrology - Ancillary temperature measuring equipment
Multi-channel thermocouple data recorders; Process calibrators;
Volt to temperature conversion;
Direct measurement by electrical input;
CAPABILITY
With Calibration and measurement capability of-
0.2 °C from -100 °C to 500 °C
0.3 °C from 500 °C to 1100 °C
Temperature metrology - Temperature measuring equipment
Digital temperature measuring systems;
Temperature;
Measurement against reference standard;
CAPABILITY
With Calibration and measurement capability of-
0.4 °C from -80 °C to 250 °C
0.6 °C from 250 °C to 500 °C
2.5 °C from 500 °C to 1100 °C
Temperature metrology - Verification of controlled enclosures
Autoclaves and sterilising ovens; Baths; Freezers; Furnaces; Incubators; Ovens;
Gradient; Spatial uniformity;
Direct temperature measurement;
AS 2853
CAPABILITY
With Calibration and measurement capability of -
0.9 °C in the range from 0 °C to 250 °C for Ovens and Furnaces
2.6 °C in the range from 250 °C to 1100 °C for Ovens and Furnaces
0.9 °C in the range from -40 °C to 100 °C for Incubators
0.8 °C in the range from 100 °C to 130 °C for Autoclaves and sterilising ovens
1 °C in the range from 131 °C to 160 °C for Autoclaves and sterilising ovens
1 °C in the range from -80 °C to -16 °C for Industrial freezers
0.9 °C in the range from -15 °C to ambient for Industrial freezers
0.8 °C in the range from 0 °C to 130 °C for Baths
Torque - Torque measuring and testing equipment
Torque wrenches;
Torque;
Comparison with a reference standard;
ISO 6789.2 and in house method 5.23
CAPABILITY
With Calibration and Measurement Capability of -
1.2% from 1.25 Nm to 1500 Nm
Volume and density - Determination of density
Alcohol meters; Apparent density (draft survey) hydrometers; Bark hydrometer; Baume hydrometer; Brix hydrometers; Density hydrometers; Soil (colloidal solution) hydrometers; Specific gravity hydrometers;
Baume; Brix; g/L; Percentage (%) alcohol by volume; Specific gravity;
Gravimetric method;
CAPABILITY
With Calibration and measurement capability of -
1/3 of the smallest scale division in the range 0 to 2.0 g/ml for Density Hydrometers;
1/3 of the smallest scale division in the range 0.990 to 1.040 kg/L for Apparent Density Hydrometers;
1/3 of the smallest scale division in the range -5 to +60 g/l for Soil Hydrometers;
1/3 of the smallest scale division in the range 0 to 90° Bx for Brix Hydrometers;
1/3 of the smallest scale division in the range 0 to 80° Bk for Bark Hydrometers;
1/3 of the smallest scale division in the range 0 to 90° Be for Baume Hydrometers;
1/3 of the smallest scale division in the range 0.6 to 2.0 for Specific Gravity Hydrometers;
1/3 of the smallest scale division in the range 0 to 100% for Alcohol Hydrometers