Back to Search Results


Address Details:
720 Springvale Road
MULGRAVE, VIC 3170
AUSTRALIA Map
https://www.kingfisherfiber.com

Contact Details:
Mr Bruce Robertson
+61(03) 85441750

Availability:
Services available to external clients

SCOPE OF ACCREDITATION   View and Print Scope of Accreditation Printable version

Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.



ISO/IEC 17025 (2017)
Calibration
SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
Optical metrology - Optical measuring equipment
Laser energy meters; Laser power meters; Optical fibre systems; Photodiodes; Radiometers;
Linearity;
Direct comparison against a reference meter;
IEC 61315 Ed. 2.0, Methods 1, 2. 3. 4, 6 and 7
CAPABILITY
Photodiodes and radiometers with Fibre Optic interface
with Calibration and Measurement Capability of -
0.01 dB per 10 dB or 0.2% of the reading whichever is greater from -70 dB to 10 dB at 1310 nm & 1550 nm, and from -70 dBm to 0 dBm at 650 nm & 850 nm, non-coherent broadband light,
0.1 dB per 10 dB or 0.7% of the reading whichever is greater from -60 dBm to 10 dBm at 850 nm, 1310 nm & 1550 nm, using coherent laser light.


Responsivity;
Direct comparison against a reference meter;
Methods 1, 2, 6 and 7
CAPABILITY
Photodiodes and radiometers with Fibre Optic interface
with Calibration and Measurement Capability of-
0.06 dB or 1.4% of the reading whichever is greater from 350 nm to 1650 nm in 5 nm steps, -60 dBm to 3 dBm / 10 uW - 1 nW and wavelength accuracy of 0.5 nm, using non-coherent light. In accordance with TIA-455-231 / IEC 61315 / FOTP 231 Calibration of Fibre Optic Power Meters

Optical fibre systems;
Power - Optical fibre; Wavelength;
Direct comparison against a reference meter; Direct comparison against a reference spectrometer;
IEC 61315 Ed. 2.0, Method 1, 2, 3, 4, 5, 6 and 7
CAPABILITY
with Calibration and Measurement Capability of-
Wavelength 
0.5 nm from 350 nm to 1700 nm
Power - Optical fibre
0.06 dB or 1.4% of reading whichever is greater from 350 nm to 1650 nm in steps of 5 nm + 22 to -60 dBm / 10 uW - 1 nW. Maximum power per IEC 60825 Class 1M / 2M

Return loss - Optical fibre
0.04 dB at 850 nm, 1310 nm, 1550 nm in the range from -60 dB to 0 dB