ISO/IEC 17025 (2017)
Calibration
SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
Optical metrology - Optical measuring equipment
Photodiodes; Photometric detectors;
Linearity; Responsivity;
Comparison with a reference meter;
Response linearity
from 0.5 to 150 000 lux
CAPABILITY
With Calibration and measurement capability of-
Broadband responsivity
1.5%

Response linearity
1.5% in the range 0.5 to 10 lux;
0.5% in the range 10.1 to 3000 lux;
1.0% in the range 3001 to 30,000 lux;
1.5% in the range 30 001 to 150 000 lux;

Luminance meters;
Luminance;
Direct comparison against a reference meter;
from 1 to 12 000 cd/m2 ;
CAPABILITY
With Calibration and measurement capability of-
1.5% from 1 to 4000 cd/m2 ;
 3% from 4001 to 12 000 cd/m2 ;

Illuminance meters; Light meters; Luxmeters; Photometers;
Cosine correction; Illuminance; Linearity; Responsivity;
Comparison with a reference meter;
CAPABILITY
With Calibration and measurement capability of-
Illuminance;
0.6% from 0.1 to 5 lux;
0.5% from 5 to 300 lux;
0.8% from 300 to 3000 lux;
0.9% in the range 3000 to 150 000 lux

Broadband responsivity;
1.5%

Response linearity;
1.5% in the range 0.5 to 10 lux;
0.5% in the range 10.1 to 3000 lux;
1.0% in the range 3001 to 30,000 lux;
1.5% in the range 30 001 to 150 000 lux

Cosine correction;
1.0% in range 0 to 30 º;
3.0% in range 30 to 75º;

Tricolorimeters;
Chromaticity; Luminance;
Direct measurement using a reference standard;
CAPABILITY
Calibration of incident light tristimulus colorimeters;
With Calibration and measurement capability of-
1.5% from 1 to 4000 cd/m2 ;
3% from 4001 to 12,000 cd/m2 ;
0.002 in CIE (1931) x and y coordinates;

Blue light radiometers; Ultraviolet radiometers; UVA meters; UVC meters;
Cosine correction; Irradiance; Linearity; Responsivity;
Comparison with a reference meter;
CAPABILITY
With Calibration and measurement capability of-
Irradiance
6.5% from 10 to 500 μW/cm2 for UVA;
7.0% from 501 to 1200 μW/cm2 for UVA;
7.5% from 1201 to 2000 μW/cm2 for UVA;
8.0% from 10 μW/cm2 to 2000 μW/cm2 for UVC;

Broadband responsivity
1.5%

Response linearity
1.5% in the range 0.5 to 10 lux;
 0.5% in the range 10.1 to 3000 lux;
 1.0% in the range 3001 to 30,000 lux;
 1.5% in the range 30 001 to 150 000 lux

Cosine correction
1.0% in range 0 to 30 º;
3.0% in range 30 to 75º;
Optical metrology - Optical measuring equipment
Spectroradiometers;
Responsivity; Wavelength;
Direct comparison against a reference spectrometer; Direct measurement using a reference standard;
from 360 nm to 830 nm;
including on site calibrations



CAPABILITY
Calibration in the laboratory and on-site;
With a Calibration and Measurement Capability of-
0.25 nm in wavelength scale;
3 % in spectral response over the wavelength range from 360 nm to 380 nm;
2 % in spectral response over the wavelength range from 380 nm to 780 nm;
3 % in spectral response over the wavelength range from 780 nm to 830 nm;
Optical metrology - Optical standards and reference equipment
Integrating sphere sources; Luminance standard sources;
Luminance;
Direct measurement using a reference meter;
from 1 to 4000 cd/m2
CAPABILITY
With Calibration and measurement capability of-
2% from 1 to 4000 cd/m2

Filters;
Chromaticity; Transmittance;
Comparison with a reference meter;
CAPABILITY
With Calibration and measurement capability of-
Broadband Transmittance;
0.5% of measured value;
Spectral Transmittance;
3.0% of measured value in the range 0.200 to 1.000 from 280 to 315 nm;
0.25% of measured value in the range 0.200 to 1.000 from 315 to 800 nm;
0.006 in the range 0 to 0.200 from 280 to 315 nm;
0.0006 in the range 0 to 0.200 from 315 to 950 nm;
Chromaticity of filters and surfaces:
0.001 in x
0.001 in y

Incandescent lamps; LED sources; Solid state lighting (SSL) sources;
Illuminance;
Direct measurement using a reference meter;
From 0.5 to 150 000 lux;
CAPABILITY
With Calibration and measurement capability of-
0.7% in the range 0.5 to 3000 lux;
1.5% in the range greater than 3000 to 150 000 lux;


Luminous flux;
Direct measurement by goniophotometer; Direct measurement by integrating sphere;
CAPABILITY
With Calibration and Measurement Capability of-
Incandescent lamps;
1.5% above 10 lm;
 2.0% below 10 lm;
LED and other Solid State Lighting sources;
1.5% above 10 lm;
 2.0% below 10 lm;


Luminous intensity;
Direct measurement by goniophotometer; Direct measurement using a reference meter;
CAPABILITY
With Calibration and Measurement Capability of-
Incandescent lamps;
0.7%;
LED and other Solid State Lighting sources;
1.2%;

Diffuse reflectance plates;
Reflectance;
Direct measurement using a reference meter; Indirect measurement using a spectrophotometer and software integration;
CAPABILITY
With Calibration and measurement capability of-
Spectral reflectance;
2.5% of measured value or 0.0025 (whichever is greater) from 380 to 410 nm;
2.0% of measured value or 0.0025 (whichever is greater) from 410 to 700 nm;
2.5% of measured value or 0.0025 (whichever is greater) from 700 to 780 nm;

Discharge sources - Lamps; LED sources; Solid state lighting (SSL) sources; UVC sources;
Broadband irradiance;
Direct measurement using a reference meter;
from 50 to 2000 μW/cm2 for UVC sources
CAPABILITY
With Calibration and measurement capability of-
4.0% from 50 to 500 μW/cm2 for UVC sources;
4.5% from 501 to 1200 μW/cm2 for UVC sources;
5.5% from 1201 to 2000 μW/cm2 for UVC sources;

Discharge sources - Lamps; Incandescent lamps; LED sources; Solid state lighting (SSL) sources;
Spectral irradiance;
Direct measurement using a reference spectrophotometer;
from 250 nm to 1000 nm;
CAPABILITY
With Calibration and measurement capability of-
5.5% of measured value from 250 to 280 nm;
4.5% of measured value from 280 to 380 nm;
3.0% of measured value from 380 to 830 nm
4.0% of measured value from 830 to 1000 nm;


Correlated colour temperature;
Direct measurement using a reference meter;
CAPABILITY
With Calibration and Measurement Capability of-
for incandescent sources;
20 K for temperatures near 2856 K;
10 K to 30 K in the range 1700 to 3500 K;
for other sources:
20 K at 1700 K rising to 40 K at 3000 K;
90 K at 5000 K;
and 430 K at 12 000 K


Chromaticity;
Direct measurement using a reference colorimeter; Direct measurement using a reference spectrophotometer;
CAPABILITY
With Calibration and Measurement Capability of-
0.001 in x;
0.001 in y;