Back to Search Results


Address Details:
1492 Ferntree Gully Road
KNOXFIELD, VIC 3180
AUSTRALIA Map
www.ecotech.com

Contact Details:
Mr Horacio Fernando Viana Herazo
+61(03) 97307800

Availability:
Services available to external clients

SCOPE OF ACCREDITATION   View and Print Scope of Accreditation Printable version

Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.



ISO/IEC 17025 (2005)
Calibration
SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
Acoustical metrology - Acoustic equipment
Sound level meters;
Electrical response; Sound pressure level;
Direct measurement by electrical input; Direct measurement by reference sound source;
IEC 61672-3 Acoustic response from 125 Hz, 1 kHz, 8 kHz;

Electrical test from 63 Hz to 16 kHz
CAPABILITY
Calibration of 01dB sound level meters
With Calibration and Measurement Capability of -
For electrical tests;
0.42 dB from 63 Hz to 4 kHz
0.58 dB at 8 kHz
0.64 dB at 16 kHz
0.24 dB for Tone Burst response, linearity, overload indication
0.3 dB for C weighted peak sound level
0.1 dB for frequency and time weightings at 1 KHz, long term and high level stability
For acoustic response;
0.35 dB for frequencies at 125 Hz
0.33 dB for frequencies at 1 kHz
0.49 dB at frequency 8 kHz
Flow metrology - Environmental flow measuring devices and systems
Anemometers;
Velocity;
Comparison measurement in wind tunnel;
0.3 m/s to 30 m/s
CAPABILITY
Calibration of Pitot tube, cup, vane, hot wire, propeller, ultra-sonic and Doppler anemometers
With Calibration and Measurement Capability of-
0.05 m/s from 0.3 m/s to 5 m/s
1% of reading from 5 m/s to 30 m/s

Wind direction devices;
Direction;
Comparison measurement in wind tunnel;
CAPABILITY
Calibration of wind vane and sonic anemometer wind direction
With Calibration and Measurement Capability of-
2º at a nominal wind speed of 6.0 m/s
Optical metrology - Optical measuring equipment
Pyranometers;
Irradiance;
Comparison with a reference meter;
ISO 9847  550 W/m2 to 1100 W/m2
including on site calibrations



CAPABILITY
With Calibration and Measurement Capability of-
2.4% of the derived sensitivity factor in μV.m2/W from 550 W/m2 to 1100 W/m2
Temperature metrology - Humidity measuring equipment
Relative humidity sensors;
Relative humidity (RH);
Comparison with a reference standard;
20% RH to 90% RH within temperature range 21 °C and 25 °C
CAPABILITY
With Calibration and Measurement Capability of-
2.1% RH in the range 20% RH to 90% RH within temperature range 21 ºC and 25 ºC
Temperature metrology - Temperature measuring equipment
Digital temperature measuring systems;
Temperature;
Measurement against reference standard;
0 °C to 40 °C
CAPABILITY
With Calibration and Measurement Capability of-
0.2 °C from 0 ºC to 40 ºC