SCOPE OF ACCREDITATION   View and Print Scope of Accreditation Printable version

Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.



ISO/IEC 17025 (2017)
Calibration
SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
DC and low frequency electrical metrology - Electrical instrument calibrators
Instrument calibrators;
AC current; AC voltage; DC current; DC voltage; Resistance;
Comparison with a primary realisation; Comparison with a reference standard;
DC Voltage up to 1100 V
AC Voltage from 0.6 mV to 1000 V and 1 Hz to 1 MHz
DC Current from 50 μA up to 10 A
AC Current from 10 μA to 100 A and 10 Hz to 30 kHz
Resistance from 0 mΩ to 1 GΩ
CAPABILITY
with Calibration and Measurement Capability of -

DC Voltage


Expanded Uncertainty
VoltageReference standard: NanovoltmeterReference standard: CalibratorReference standard: JAVS
10 nV0.1 μV-0.05 μV
> 10 nV to 100 nV0.2 μV-0.05 μV
> 100 nV to < 100 mV0.3 μV-0.05 μV
100 mV to 220 mV2.2 μV11 μV/V1.4 μV/V
220 mV to 10 V-2.9 μV/V0.8 μV/V



Relative Expanded Uncertainty / (μV/V)
VoltageReference standard: CalibratorReference standard: JAVS
10 V to 22 V2.31.1
22 V to 220 V2.11.2
220 V to 1100 V2.81.2


AC Voltage


Expanded Uncertainty / (μV/V)

Digital
Sampling
Voltmeter
Digital
Sampling
Voltmeter
Voltage Measurement Standard
Frequency1 Hz to
< 5 Hz
5 Hz to
< 10 Hz
10 Hz to
< 20 Hz
20 Hz to
< 30 Hz
30 Hz to
< 40 Hz
40 Hz to
< 57 Hz
57 Hz to
< 100 Hz
100 Hz to
< 400 Hz
400 Hz to
< 1 kHz
1 kHz> 1 kHz to
5 kHz
0.6 mV to < 1 mV--250200210190180190170180170
1 mV to < 2.2 mV--120909011090110908090
2.2 mV to < 6 mV14044404070444040404041
6 mV to < 22 mV450110595038813952423838
22 mV to < 60 mV12049443431272729322727
60 mV to < 220 mV3207213011098979696969696
0.22 V to < 0.6 V8226462920131211111115
0.6 V to 1 V2506133201391010889
> 1 V to 6 V6820332113101010101010
> 6V to 22 V26054332114131213131212
> 22 V to 60 V6716412420181718172019
> 60 V to 100 V3720362318181818181718
> 100 V to 250 V7540352318181818181818
> 250 V to 300 V7540---202020202020
> 300 V to 750 V7540---202020202024
> 750 V to 1000 V-----212121212126



Expanded Uncertainty / (μV/V)

Voltage Measurement Standard
Frequency> 5 kHz to
10 kHz
> 10 kHz to
20 kHz
> 20 kHz to 30 kHz> 30 kHz to 50 kHz> 50 kHz to 100 kHz> 100 kHz to 200 kHz> 200 kHz to 300 kHz> 300 kHz to 500 kHz> 500 kHz to 800 kHz> 800 kHz to 1 MHz
0.6 mV to < 1 mV17016015025069011001300140019002000
1 mV to < 2.2 mV
808090120140310310490710690
2.2 mV to < 6 mV4040404153110140230270300
6 mV to < 22 mV38403845476984150210230
22 mV to < 60 mV2734343444546789160180
60 mV to < 220 mV98989899110110120140200220
0.22 V to < 0.6 V131418202226335882110
0.6 V to 1 V9101214182227417499
> 1 V to 6 V101212141926314995140
> 6 V to 22 V1214141624323451110160
> 22 V to 60 V19302643354354---
> 60 V to 100 V181818212842----
> 100 V to 250 V1818182129-----
> 250 V to 300 V2020232333-----
> 300 V to 750 V2424243466-----
> 750 V to 1000 V262632-------


DC Current


Relative Expanded Uncertainty / (μA/A)
50 μA to < 100 μA30
100 μA to < 1 mA14
1 mA to < 10 mA12
10 mA to < 100 mA7
100 mA to < 1A13
1 A to 10 A12


AC Current


Expanded Uncertainty / (μA/A)

10 Hz to < 57 Hz57 Hz> 57 Hz to 400 Hz> 400 Hz to 1 kHz> 1 kHz to 5 kHz> 5 kHz to 10 kHz> 10 kHz to 30 kHz
10 μA to < 0.1 mA230909092140600-
0.1 mA to < 1 mA12045454767300620
1 mA to < 10 mA12044444663300150
10 mA to < 100 mA12044444563300170
100 mA6526262640180130
> 100 mA to 1 A1003956200250300-
> 1 A to 10 A20078120400---
> 10 A to 20 A250100150500---

DC and low frequency electrical metrology - Electrical measurement and test equipment
Ammeters; Digital multimeters (DMM); Electricity meters; Ohm meters; Power factor meters; Varmeters; Voltmeters; Watt meters;
AC current; AC voltage; Apparent power; DC current; DC voltage; Power factor; Resistance;
Comparison with a primary realisation; Comparison with a reference standard;
DC Voltage from 1 mV up to 1100 V
AC Voltage from 1 mV to 1000 V and 1 Hz to 1 MHz
DC Current from 10 pA up to 20 A
AC Current from 10 μA to 20 A and 10 Hz to 30 kHz
Resistance from 0 mΩ to 1 GΩ
Wattmeters and Electricity Meters - sinewave from 1 to 1000 V and 0.01 to 200 A and 47 Hz to 60 Hz
Wattmeters and Electricity Meters - with distorted voltage or current from 1 to 240 V and 0.001 A to 10 A and 47 Hz to 60 Hz
fundamental frequency and voltage or current harmonics up to 200 kHz
Varmeters from 60 to 240 V and 0.1 to 120 A at 47 to 60 Hz
Power Factor Meters from 60 to 480 V and 0.1 A to 100 A and 47 to 60 Hz
CAPABILITY
with Calibration and Measurement Capability of -
DC Voltage

Relative Expanded Uncertainty / (μV/V)
VoltageReference standard: CalibratorReference standard: JAVS
1 mV to 10 mV50050
10 mV to 100 mV508.6
100 mV to 1 V80.5
1 V to 10 V30.1
10 V to 100 V20.1
100 V to 1000 V51.0

AC Voltage
For Digital Multimeter

Relative Expanded Uncertainty / (μV/V)

Digital
Sampling
Voltmeter
Digital
Sampling
Voltmeter
Voltage Measurement Standard
Frequency1 Hz to < 5 Hz5 Hz to < 10 Hz10 Hz to < 20 Hz20 Hz to < 30 Hz30 Hz to < 40 Hz40 Hz to < 57 Hz57 Hz to < 100 Hz100 Hz to < 400 Hz400 Hz to < 1 kHz1 kHz
3 mV to < 10 mV6030790690670670670620620620
10 mV to < 30 mV6020670560540530530460460460
30 mV to < 100 mV6020440430420420420380380380
100 mV to < 0.3 V6020130100808080808080
0.3 V to < 1 V602010070505050505050
1 V to 3 V60208060404040404040
> 3 V to 10 V60208060404040404040
> 10 V to 30 V60208060404040404040
> 30 V to 100 V60208060404040404040
> 100 V to 700 V60208060606040404040
> 700 V to 1000 V--10080707070707070


Relative Expanded Uncertainty / (μV/V)

Voltage Measurement Standard
Frequency> 1 kHz to 5 kHz> 5 kHz to 10 kHz
> 10 kHz to 20 kHz> 20 kHz to 30 kHz> 30 kHz to 50 kHz> 50 kHz to 100 kHz> 100 kHz to 200 kHz> 200 kHz to 300 kHz> 300 kHz to 500 kHz> 500 kHz to 1 MHz
3 mV to < 10 mV
620620620620720750850170024005000
10 mV to < 30 mV460460470480600640750160024005000
30 mV to < 100 mV380380380390470500620110012002500
100 Mv to < 0.3 V8080901001902704504806701500
0.3 V to < 1 V505060801102104204506501400
1 V to 3 V405060701102104104506501400
> 3 V to 10 V405060701102104104506501400
> 10 V to 30 V405060701102104104506501400
> 30 V to 100 V40506070110210410450640-
> 100 V to 700 V40506080110210410450--
> 700 V to 1000 V707080130160290----

 DC Current (including Picoameters)

Relative Expanded Uncertainty / (μA/A)
100 μA120
10 μA210
1 μA210
100 nA530
10 nA530
1 nA530
100 pA700
10 pA1400


Relative Expanded Uncertainty / (μA/A)
100 μA to < 1 mA14
1 mA to < 10 mA12
10 mA to 100 mA7
> 100 mA to 1 A13
> 1 A to 10 A12
> 10 A to 20 A26

AC Current

Expanded Uncertainty / (μA/A)

10 Hz to < 57 Hz57 Hz to < 400 Hz400 Hz to < 1 kHz1 kHz> 1 kHz to 5 kHz> 5 kHz to 10 kHz> 10 kHz to 20 kHz> 20 kHz to 30 kHz
10 μA to < 50 μA740430450440----
50 μA to < 100 μA470140140140----
0.1 mA to < 1 mA12059596988110340620
1 mA to < 10 mA1205959597387100150
10 mA to < 100 mA1205858586271120170
100 mA86292929294583130
> 100 mA to 1 A683030304583120200
> 1 A to 10 A110727295300500700-
> 10 A to 20 A1701101101504507501100-

Ohmmeters0.2 μΩ from 0 mΩ to 100 mΩ
2 μΩ/Ω from 100 mΩ to 1 Ω 
As per following Table from 1 Ω to 1 G Ω

Relative Expanded Uncertainty / (uΩ/Ω)
1 Ω to < 10 Ω2
10 Ω to 10 kΩ1
> 10 kΩ to 100 kΩ2
> 100 kΩ to 1 MΩ3
> 1 MΩ to 100 MΩ10
> 100 MΩ to 1 GΩ100

Wattmeters and Electricity Meters
Mains Frequency Power

Expanded Uncertainty / (μW/(VA))VoltageFrequencyPower Factor
0.01 A to 0.1 A201 V to 1000 V47 Hz to 60 Hz1 to 0 L/C
> 0.1 A to 100 A
> 100 A to 200 A
10
30
1 V to 1000 V
1 V to 1000 V
47 Hz to 60 Hz
47 Hz to 60 Hz
1 to 0 L/C
1 to 0 L/C

Power with Distorted Voltage or Current

Expanded Uncertainty / (μW/(VA))VoltageFrequencyVoltage or Current Harmonics
maximum frequency / kHz
Power Factor
0.001 A to 0.01 A10060 V to 240 V47 Hz to 60 Hz2001 to 0 L/C
> 0.01 A to 10 A
5060 V to 240 V47 Hz to 60 Hz
2001 to 0 L/C

Varmeters
0.004% of Apparent Power from 60 to 240 V and 0.1 to 120 A at 47 to 60 Hz

Power Factor Meters
0.00005 power factor from 60 to 480 V and 0.1 A to 100 A and 47 to 60 Hz

Multi-purpose AC measurement and transfer standard;
AC current; AC voltage; DC current; DC voltage;
Comparison with a primary realisation; Comparison with a reference standard;
DC Voltage from 1 mV up to 1100 V
AC Voltage from 1 mV to 1000 V and 1 Hz to 1 MHz
DC Current from 10 pA up to 20 A
AC Current from 10 μA to 20 A and 10 Hz to 30 kHz
CAPABILITY
with Calibration and Measurement Capability of -

DC Voltage

Relative Expanded Uncertainty / (μV/V)
VoltageReference standard: CalibratorReference standard: JAVS
1 mV to 10 mV50050
10 mV to 100 mV508.6
100 mV to 1 V80.5
1 V to 10 V30.1
10 V to 100 V20.1
100 V to 1000 V51.0

AC Voltage
For AC Measurement Standard


Relative Expanded Uncertainty / (μV/V)
Frequency10 Hz to
< 20 Hz
20 Hz to
< 30 Hz
30 Hz to
< 40 Hz
40 Hz to
< 57 Hz
57 Hz to
< 100 Hz
100 Hz to
< 400 Hz
400 Hz to
< 1 kHz
1 kHz> 1 kHz
to 5 kHz
> 5 kHz
to 10 kHz
1 mV to < 2 mV200150120100100100100100100100
2 mV to < 6 mV1301007070505050505050
6 mV to < 20 mV90704535353535353535
20 mV to < 60 mV70504035353535353535
60 mV to < 200 mV65453535353535353535
200 mV to < 600 mV60363020202020202020
600 mV to < 2 V
10988888888
2V
9877777777
> 2 V to 6 V
11999999999
> 6 V to 20 V
111099999999
> 20 V to 60 V
15121111111111111111
> 60 V to 200 V
20151515151515151515
> 200 V to 600 V20181717171717172121
> 600 V to 1000 V41272323232323232325
 

Relative Expanded Uncertainty / (μV/V)
Frequency> 10 kHz to
20 kHz
> 20 kHz to
30 kHz
> 30 kHz to
50 kHz
> 50 kHz to
100 kHz
> 100 kHz to
200 kHz
> 200 kHz to
300 kHz
> 300 kHz to
500 kHz
> 500 kHz to
800 kHz
> 800 kHz to
1 MHz
1 mV to < 2 mV1001001002008001100120015001600
2 mV to < 6 mV505050130180200220550800
6 mV to < 20 mV3535357080110200400600
20 mV to < 60 mV353535556590140260300
60 mV to < 200 mV353535506080130200240
200 mV to < 600 mV242832364470130190220
600 mV to < 2 V
888121626344050
2V
777101422293443
> 2 V to 6 V
999141827395060
> 6 V to 20 V
9111215232654--
> 20 V to 60 V
11121417-----
> 60 V to 200 V
38383838-----
> 200 V to 600 V22273545-----
> 600 V to 1000 V28334057-----
 
DC Current (including Picoameters)


Relative Expanded Uncertainty / (μA/A)
100 μA120
10 μA210
1 μA210
100 nA530
10 nA530
1 nA530
100 pA700
10 pA1400


Relative Expanded Uncertainty / (μA/A)
100 μA to < 1 mA14
1 mA to < 10 mA12
10 mA to 100 mA7
> 100 mA to 1 A13
> 1 A to 10 A12
> 10 A to 20 A26

AC Current

Expanded Uncertainty / (μA/A)

10 Hz to < 57 Hz57 Hz to < 400 Hz400 Hz to < 1 kHz1 kHz> 1 kHz to 5 kHz> 5 kHz to 10 kHz> 10 kHz to 20 kHz> 20 kHz to 30 kHz
10 μA to < 50 μA740430450440----
50 μA to < 100 μA470140140140----
0.1 mA to < 1 mA12059596988110340620
1 mA to < 10 mA1205959597387100150
10 mA to < 100 mA1205858586271120170
100 mA86292929294583130
> 100 mA to 1 A683030304583120200
> 1 A to 10 A110727295300500700-
> 10 A to 20 A1701101101504507501100-
DC and low frequency electrical metrology - Electrical standards
Current shunts;
Current AC-DC difference; Resistance; Voltage AC-DC difference;
Comparison with a primary realisation; Comparison with a reference standard;
0.0001 to 1000 Ω
Current 0.1 mA to 100 A
Frequency 10 Hz to 200 kHz
CAPABILITY
with Calibration and Measurement Capability of -


Relative Expanded Uncertainty / (μΩ/Ω)

10 Hz to < 20 Hz20 Hz to < 30 Hz30 Hz to < 40 Hz40 Hz to < 57 Hz57 Hz to < 100 Hz100 Hz to < 400 Hz400 Hz to < 1 kHz1 kHz
0.1 mA to < 1 mA9952403636364451
1 mA to < 10 mA9952393534343941
10 mA to < 30 mA65544444
30 mA to < 50 mA86544444
50 mA to < 100 mA97544444
100 mA97654444
> 100 mA to 400 mA108655544
> 400 mA to 1.5 A118655555
> 1.5 A to 5 A119755555
> 5 A to 20 A1412111010101010
> 20 A to 50 A1916151414141413
> 50 A to 100 A2218171616151414



Relative Expanded Uncertainty / (μΩ/Ω)

> 1 kHz to 5 kHz> 5 kHz to 10 kHz> 10 kHz to 20 kHz> 20 kHz to 30 kHz> 30 kHz to 50 kHz> 50 kHz to 100 kHz> 100 kHz to 200 kHz
0.1 mA to < 1 mA73120-----
1 mA to < 10 mA4965-----
10 mA to < 30 mA44567810
30 mA to < 50 mA45567911
50 mA to < 100 mA456781013
100 mA456791217
> 100 mA to 400 mA5668101218
> 400 mA to 1.5 A5678101319
> 1.5 A to 5 A56791321-
> 5 A to 20 A101416212632-
> 20 A to 50 A162129374661-
> 50 A to 100 A182638526584-



Phase displacement;
Comparison with a primary realisation; Comparison with a reference standard;
-0.1 rad to 0.1 rad
Current 0.01 A to 100 A
Frequency 40 Hz to 200 kHz
CAPABILITY
with Calibration and Measurement Capability of -


Expanded Uncertainty / (μrad)

40 Hz to 1 kHz1 kHz to 10 kHz> 10 kHz to 20 kHz> 20 kHz to 50 kHz> 50 kHz to 100 kHz> 100 kHz to 200 kHz
0.01 A to 0.5 A38164181200
> 0.5 A to 1 A410194795230
> 1 A to 2 A4112153110250
> 2 A to 5 A5122358120280
> 5 A to 10 A6132563130300
> 10 A to 20 A6163075150
> 20 A to 40 A6183586170
> 40 A to 100 A7203996190


Voltage standards - E.M.F. reference devices;
Voltage;
Comparison with a primary realisation; Comparison with a reference standard;
at 1 V, 1.018 V and 10 V
CAPABILITY
with Calibration and Measurement Capability of -


Relative Expanded Uncertainty / (nV/V)
1 V40
1.018 V40
10 V20


Potential dividers; Voltage ratio boxes;
Voltage ratio;
Comparison with a primary realisation; Comparison with a reference standard;
For ratios of 0.001 to 0.5 with maximum voltage of 1000 V DC
CAPABILITY
with Calibration and Measurement Capability of -
0.5 μV/V

10050
20016


AC/DC transfer instruments;
Voltage AC-DC difference;
Comparison with a reference standard;
1 mV to 1000 V
10 Hz to 1 MHz
CAPABILITY
with Calibration and Measurement Capability of -


Expanded Uncertainty / (μV/V)

10 Hz to < 20 Hz20 Hz to < 30 Hz30 Hz to < 40 Hz40 Hz to < 57 Hz57 Hz to < 100 Hz100 Hz to < 400 Hz400 Hz to < 1 kHz1 kHz> 1 kHz to 5 kHz> 5 kHz to 10 kHz
1 mV to < 2 mV260220200170170170170170180180
2 mV to < 5 mV110857777777777777878
5 mV to < 10 mV77585353535353535454
10 mV to < 20 mV53413838383838384040
20 mV to < 30 mV46333131313131313333
30 mV to < 50 mV38282424232323242424
50 mV to < 100 mV30201614141414141516
100 mV to < 200 mV28201613131313131515
200 mV to < 300 mV28201613121212121414
300 mV to < 0.5 V28201613121212121414
0.5 V to < 1 V4.82.51.61.31.21.21.21.21.41.4
1 V to < 2 V5.22.61.71.31.31.21.21.21.41.4
2 V to < 3 V5.12.61.61.31.31.21.21.21.41.4
3 V to 4 V5.22.61.71.31.31.21.21.21.41.4
> 4 V to 6 V7555555444
> 6 V to 10 V7666666555
> 10 V to 12 V8766666666
> 12 V to 20 V8777777666
> 20 V to 30 V10988888888
> 30 V to 40 V10999999888
> 40 V to 60 V1110999910999
> 60 V to 100 V111010101010109910
> 100 V to 120 V12111010101010101010
> 120 V to 200 V12111111111111101011
> 200 V to 300 V14121212121212111112
> 300 V to 400 V14131212121212121212
> 400 V to 600 V15141414141414131316
> 600 V to 1000 V17161616161616151618



Expanded Uncertainty / (μV/V)

> 10 kHz to 20 kHz> 20 kHz to 30 kHz> 30 kHz to 50 kHz> 50 kHz to 100 kHz> 100 kHz to 200 kHz> 200 kHz to 300 kHz> 300 kHz to 500 kHz> 500 kHz to 800 kHz> 800 kHz to 1 MHz
1 mV to 2 mV180180180190220270310360420
2 mV to < 5 mV78798088100160220280310
5 mV to < 10 mV5456576875120180220260
10 mV to < 20 mV404143566370110170200
20 mV to < 30 mV33333645536689130150
30 mV to < 50 mV25252631415780140150
50 mV to < 100 mV16172026375271120140
100 mV to < 200 mV15162026385473129140
200 mV to < 300 mV15161925375167110120
300 mV to < 0.5 V15161925375167110120
0.5 V to < 1 V1.62.23.05.26.88.310.613.817.6
1 V to < 2 V1.62.23.05.26.88.310.613.817.6
2 V to < 3 V1.62.12.85.06.89.211.615.819.5
3 V to 4 V1.62.12.85.06.89.211.615.819.5
> 4 V to 6 V45581115233037
> 6 V to 10 V55691217253239
> 10 V to 12 V66691318263340
> 12 V to 20 V677101419273442
> 20 V to 30 V9111317222837--
> 30 V to 40 V9111418233039--
> 40 V to 60 V101114192531---
> 60 V to 100 V1012152026----
> 100 V to 120 V11121521-----
> 120 V to 200 V11131621-----
> 200 V to 300 V12141723-----
> 300 V to 400 V12151825-----
> 400 V to 600 V17222839-----
> 600 V to 1000 V20263649-----




Current AC-DC difference;
Comparison with a reference standard;
from 0.0001 A to 100 A
from 10 Hz to 200 kHz
CAPABILITY
with Calibration and Measurement Capability of -


Expanded Uncertainty / (μA/A)

10 Hz to < 20 Hz20 Hz to < 30 Hz30 Hz to < 40 Hz40 Hz to < 57 Hz57 Hz to < 100 Hz100 Hz to < 400 Hz400 Hz to < 1 kHz1 kHz
0.1 mA to < 1 mA9952403636364451
1 mA to < 10 mA9952393534343941
10 mA to < 30 mA65544444
30 mA to < 50 mA86544444
50 mA to < 100 mA97544444
100 mA97654444
> 100 mA to 400 mA108655544
> 400 mA to 1.5 A118655555
> 1.5 A to 5 A119755555
> 5 A to 20 A1412111010101010
> 20 A to 50 A1916151414141413
> 50 A to 100 A2218171616151414



Expanded Uncertainty / (μA/A)

> 1 kHz to 5 kHz> 5 kHz to 10 kHz> 10 kHz to 20 kHz> 20 kHz to 30 kHz> 30 kHz to 50 kHz> 50 kHz to 100 kHz> 100 kHz to 200 kHz
0.1 mA to < 1 mA73120-----
1 mA to < 10 mA4965-----
10 mA to < 30 mA44567810
30 mA to < 50 mA45567911
50 mA to < 100 mA456781013
100 mA456791217
> 100 mA to 400 mA5668101218
> 400 mA to 1.5 A5678101319
> 1.5 A to 5 A56791321-
> 5 A to 20 A101416212632-
> 20 A to 50 A162129374661-
> 50 A to 100 A182638526584-

Magnetism - Magnetic instruments and equipment
Helmholtz coils; Magnets; Solenoids;
Magnetic field strength;
Direct measurement against a reference standard;
10 mT to 2 T
CAPABILITY
with Calibration and Measurement Capability of -
0.3%

Flux meters;
Magnetic flux;
Direct measurement against a reference standard;
10 μWb to 10 mWb
CAPABILITY
with Calibration and Measurement Capability of -
1%

Flux density meters;
Magnetic flux density;
Direct measurement against a reference standard;
0 to 2 T on DC
0 to 0.005 T on 50 Hz and 60 Hz
CAPABILITY
with Calibration and Measurement Capability of -

Transverse Probes, Reference Magnets


Magnetic Flux Density
(T)
Expanded Uncertainty
0 to < 0.05
0.05 to 2
0.1 mT + 0.5 %
0.15 %

Axial Probes

Magnetic Flux Density
(T)
Expanded Uncertainty
0 to 0.0950.1 mT + 0.5 %

Time and frequency metrology - Frequency, time and waveform measuring equipment
Clocks and timers; Counters; Distortion analysers; Electrical power calibrators; Frequency analysers; Frequency meters; Oscilloscopes; Power analysers; Time interval meters;
Attenuation; Current; Current harmonics; Distortion; Frequency; Phase angle; Voltage; Voltage flicker; Voltage harmonics;
Measurement against reference standard;
Frequency Characteristics
frequency from 47 to 53 Hz and 1 to 240 V
rate of change of frequency from 47 to 53 Hz and 1 to 240 V


Distortion
Total harmonic distortion up to 99th harmonic
from 0.01 V to 240 V, 0.005 A to 20 A and 47 to 63 Hz


Other Characteristics
 Voltage Harmonics up to 99th, amplitude and phase, from 1 to 240 V and 47 to 53 Hz
 Current Harmonics up to 99th, amplitude and phase, from 0.005 to 20 A and 47 to 53 Hz
 Voltage Flicker from 1 to 240 V and 47 to 53 Hz
 Total Vector Error from 1 to 240 V and 47 to 53 Hz
 Total Vector Error from 0.005 to 20 A and 47 to 53 Hz
 Phase angle from -180 ° to 180 °, 0.01 to 1000 V, 0.005 to 200 A and 40 to 1000 Hz
CAPABILITY
with Calibration and Measurement Capability of -
Frequency Characteristics
0.0034% for frequency from 47 to 53 Hz and 1 to 240 V
0.0016 Hz/s for rate of change of frequency from 47 to 53 Hz and 1 to 240 V


Distortion

VoltageHarmonic NumberExpanded Uncertainty
(% of fundamental)
Reference Standard
0.01 V to 240 V1 to 390.001Josephson Arbitraty Wavefom Synthesizer
1 V to 240 V1 to 990.005Digital Sampling Voltmeter


Voltage Harmonics  (Meter and Source)

DistortionExpanded Uncertainty (degrees)
Harmonic Number
1 - 911 - 1921 - 39
5 % to < 10 %0.00160.00210.0099
10 % to < 20 %0.00140.00190.0097
20 % to < 30 %0.00120.00180.0095
30 % to < 40 %0.00110.00180.0094

Current Harmonics (Meter and Source)

DistortionExpanded Uncertainty (degrees)
Harmonic Number
1 - 911 - 1921 - 39
5 % to < 10 %0.00260.00310.0109
10 % to < 20 %0.00240.00290.0107
20 % to < 30 %0.00220.00280.0105
30 % to < 40 %0.00210.00280.0104

Flicker Meters
0.01% of fundamental for Voltage Flicker from 1 to 240 V and 47 to 53 Hz


Phasor Measurement Units
0.001% for Total Vector Error from 1 to 240 V and 47 to 53 Hz
0.001% for Total Vector Error from 0.005 to 20 A and 47 to 53 Hz
0.0022 ° for phase angle from 0 ° to 90 °, 1 to 240 V, 0.005 to 20 A and 47 to 53 Hz

Phase Sources, Phase Standards, Power Calibrators

FrequencyExpanded Uncertainty
(Hz)(Degrees)
40 to 600.0005
> 60 to 4000.001
> 400 to 10000.002

Phase Meters, Power Analysers

FrequencyExpanded Uncertainty
(Hz)(Degrees)
40 to 600.0005
> 60 to 4000.001
> 400 to 10000.002