SCOPE OF ACCREDITATION   View and Print Scope of Accreditation Printable version

Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.


  • ISO/IEC 17025 (2005) - The uncertainty of measurement is reported as an expanded uncertainty having a level of confidence of 95% unless stated otherwise
    Calibration
  • SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
    Acoustical metrology - Acoustic equipment
    Attenuators;
    Attenuation;

    Stepped 600 Ω resistive attenuators
    for attenuation range up to 110 dB
    CAPABILITY
    with calibration and measurement capability of -
    0.003 dB for attenuation equal to and less than 40 dB
    0.006 dB for attenuation from 40 dB to 60 dB
    0.008 dB for attenuation from 60 dB to 70 dB
    0.015 dB for attenuation from 70 dB to 80 dB
    0.05 dB for attenuation from 80 dB to 90 dB
    0.15 dB for attenuation from 90 dB to 100 dB
    0.45 dB for attenuation from 100 dB to 110 dB

    Band pass filters;
    Attenuation;
    Direct measurement by electrical input;
    By methods of -
    AS/NZS 4476, IEC 1260
    For attenuation range -
    up to 80 dB
    CAPABILITY
    with calibration and measurement capability of -
    0.1 dB for attenuation equal to and less than 40 dB
    0.2 dB for attenuation from above 40 dB to 60 dB
    0.3 dB for attenuation from above 60 dB to 80 dB

    Sound level meters;
    Sound pressure level;
    By methods of AS 1259.1; By methods of AS 1259.2; By methods of IEC 651; By methods of IEC 804; By methods of IEC 61672-3;
    For frequency range -
    31.5 Hz to 20 kHz
    CAPABILITY
    with calibration and measurement capability of -
    Electrical tests,
    0.08 dB
    Acoustic tests,
    0.2 dB from 31.5 Hz to below 10 kHz
    0.3 dB from 10 kHz to below 12.5 kHz
    0.4 dB from 12.5 kHz to 20 kHz
    Acoustical metrology - Acoustic standards and reference equipment
    Acoustic calibrators - Multi frequency; Acoustic calibrators - Single frequency; Pistonphones; Reference sound sources;
    Frequency; Pressure sensitivity;
    By methods of AS/IEC 60942; By methods of IEC 60942;
    for frequency range -
    31.5 Hz to 16 kHz
    CAPABILITY
    with calibration and measurement capability of -
    0.09 dB at 31.5 Hz to below 63 Hz
    0.08 dB from 63 Hz to below 125 Hz
    0.07 dB from 125 Hz to below 1 kHz
    0.06 dB at 1 kHz
    0.07 dB from 1 kHz to below 4 kHz
    0.08 dB from 4 kHz to below 8 kHz
    0.09 dB from 8 kHz to below 12.5 kHz
    0.15 dB from 12.5 kHz to below 16 kHz
    0.23 dB at 16 kHz


    Reference microphones;
    Frequency response;
    Direct measurement by active coupler; Direct measurement by electrostatic actuator;
    For frequency range -
    1 Hz to 20 kHz
    CAPABILITY
    Frequency response by electrostatic actuator method over the
    frequency range 20 Hz to 20 kHz
    with calibration and measurement capability of-
    0.1 dB for frequency response by actuator up to 10 kHz for type WS1
    microphones and to 20 kHz for type WS2 microphones
    Frequency response for low frequency microphones in the range 1 Hz
    to 250 Hz using a large volume active coupler
    with calibration and measurement capability of-
    0.4 dB from 1 Hz to 200 Hz
    1.0 dB from above 200 Hz to 250 Hz


    Pressure sensitivity;
    By methods of IEC 60942; Pressure field comparison measurement; Reciprocity method;
    For frequency range 20 Hz to 31.5 kHz
    CAPABILITY
    Pressure sensitivity by reciprocity method according to IEC 61094-2
    between 20 Hz and 10 kHz for LS1 microphones
    0.1 dB from 20 Hz to 25 Hz 
    with calibration and measurement capability of -
    0.08 dB from above 25 Hz to 50 Hz
    0.04 dB from above 50 Hz to 5 kHz
    0.06 dB from above 5 kHz to 8 kHz
    0.08 dB from above 8 kHz to 10 kHz
    and between 20 Hz and 31.5 kHz for LS2 microphones
    with calibration and measurement capability of-
    0.1 dB from 20 Hz to 25 Hz
    0.08 dB from above 25 Hz to 50 Hz
    0.06 dB from above 50 Hz to 100 Hz
    0.05 dB from above 100 Hz to below 500 Hz
    0.04 dB from 500 Hz to 8 kHz
    0.05 dB from above 8 kHz to 10 kHz
    0.06 dB from above 10 kHz to 12.5 kHz
    0.08 dB from above 12.5 kHz to 16 kHz
    0.10 dB from above 16 kHz to 20 kHz
    0.15 dB from above 20 kHz to 25 kHz
    0.25 dB from above 25 kHz to 31.5 kHz
    Pressure sensitivity by modified reciprocity method at 250 Hz, 500 Hz or 1 kHz
    with calibration and measurement capability of-
    0.05 dB for microphone classes: LS1
    0.06 dB for microphone classes: WS1
    Pressure sensitivity of LS2 and WS2 type microphones by comparison
    using an active coupler in the frequency range 20 Hz to 25 kHz
    with calibration and measurement capability of -
    0.13 dB from 20 Hz to 50 Hz
    0.08 dB from above 50 to 4 kHz
    0.10 dB from above 4 kHz to 15 kHz
    0.13 dB from above 15 kHz to 20 kHz
    0.18 dB from above 20 kHz to 23 kHz
    0.26 dB from above 23 kHz to 25 kHz
    and for other sizes in the range 20 Hz to 1 kHz
    with Calibration and Measurement Capability of -
    0.15 dB from 20 Hz to 25 Hz
    0.10 dB from above 25 to 50 Hz
    0.07 dB from above 50 Hz to 1 kHz


    Free field response;
    Substitution comparison;
    For frequency range -
    31.5 Hz to 20 kHz
    CAPABILITY
    Free field response by substitution comparison
    with calibration and measurement capability of -
    0.2 dB from 31.5 Hz to below 10 kHz
    0.3 dB from 10 kHz to below 12.5 kHz
    0.4 dB from 12.5 kHz to 20 kHz
    Ultrasonics - Ultrasonic measuring equipment
    Ultrasonic transducers;
    Power;
    Direct measurement of radiation force;
    IEC 61161
    From 10 mW to 20 W
    CAPABILITY
    Calibration of power output of ultrasonic transducers in liquids- radiation force
    method according to IEC 61161
    across the frequency range from 0.5 MHz to 25 MHz
    with least uncertainties of measurement of -
    6% of reading from 10 mW to 100 mW
    3% of reading from 100 mW to 1 W
    5% of reading from 1 W to 20 W


    Ultrasonic power meters;
    Power;
    Direct measurement of radiation force;
    IEC 61161
    From 10 mW to 20 W
    CAPABILITY
    Calibration of ultrasonic force (power) balances by transfer from calibrated ultrasonic transducer according to IEC 61161
    across the frequency range from 0.5 MHz to 25 MHz
    with least uncertainties of measurement of -
    8% of reading from 10 mW to 100 mW
    5% of reading from 100 mW to 1 W
    7% of reading from 1 W to 20 W

    Vibration metrology - Vibration equipment
    Geophones;
    Acceleration sensitivity; Displacement sensitivity; Phase sensitivity; Velocity sensitivity;
    Comparison measurement against reference transducer; Comparison with a reference standard;
    for frequency range 0.5 Hz to 250 Hz
    CAPABILITY
    Geophones, seismometers and similar low-frequency transducers without external reference surfaces
    with least calibration and measurement capability of -
    Magnitude:
    1% of reading for voltage sensitivity, current sensitivity, charge sensitivity or displayed units from 0.5 Hz to 250 Hz
    Phase:
    0.7° from 0.5 Hz to 250 Hz

    Accelerometers; Shock measuring systems;
    Acceleration sensitivity;
    Comparison measurement against reference transducer;
    For frequency range 80 Hz to 8 kHz and acceleration up to 5000 m/s2
    CAPABILITY
    Calibration and verification of shock measuring systems which include accelerometers, for compliance to requirements of SAE j211b, SAE j211/1 and ISO 6487 channel frequency class 60 Hz, 180 Hz, 600 Hz or 1000 Hz
    with calibration and measurement capability of -
    Magnitude:
    0.4% of reading from 0.5 Hz to below 5 kHz
    0.5% of reading from 5 kHz to below 6.3 kHz
    0.6% of reading from 6.3 kHz to 10 kHz
    Phase:
    0.7° from 0.5 Hz to below 630 Hz
    0.8° from 630 Hz to below 2.5 kHz
    0.9° from 2.5 kHz to below 5 kHz
    1.2° from 5 kHz to 10 kHz
    In the acceleration range (or equivalent) 0.01 m/s2 up to 10 m/s2 below 10 Hz, up to 50 m/s2 from 10 Hz, up to 100 m/s2 from 40 Hz, up to 1000 m/s2 from 160 Hz, and up to 5000 m/s2 at approximately 650 Hz

    Accelerometers; Machine condition monitoring equipment; Velocity transducers; Vibration analysers; Vibration measuring systems;
    Acceleration sensitivity; Displacement sensitivity; Velocity sensitivity;

    by method ISO 16061-11
    For frequency range 0 Hz to 10 kHz
    and acceleration range 0.01 m/s2 to 1000 m/s2
    CAPABILITY
    Absolute calibration of complex vibration amplitude sensitivity by interferometry from 0.5 Hz to 10 kHz by the methods of ISO 16063-11
    Transducer response in voltage sensitivity, current sensitivity, charge sensitivity or displayed units
    with calibration and measurement capability of -
    Magnitude:
    0.3% of reading from 0.5 Hz to below 9 kHz
    0.5% of reading from 9 kHz to 10 kHz
    Phase:
    0.5° from 0.5 Hz to below 5 kHz
    1° from 5 kHz to 10 kHz
    In the acceleration range (or equivalent) 0.01 m/s2 up to 10 m/s2 below 10 Hz, up to 50 m/s2 from 10 Hz, up to 100 m/s2 from 40 Hz, up to 1000 m/s2 from 160 Hz, and up to 5000 m/s2 at approximately 650 Hz
    Extended lower velocity range down to 1 mm/s from 0.5 Hz to below 2 Hz
    Vibration transducers of mass up to 5000 grams for frequencies of 6 kHz and below, and up to 100 grams for frequencies of 10 kHz and below
    by comparison to the methods of ISO 16063-21
    with calibration and measurement capability of -
    Magnitude:
    0.4% of reading from 0.5 Hz to below 5 kHz
    0.5% of reading from 5 kHz to below 6.3 kHz
    0.6% of reading from 6.3 kHz to 10 kHz
    Phase:
    0.7° from 0.5 Hz to below 630 Hz
    0.8° from 630 Hz to below 2.5 kHz
    0.9° from 2.5 kHz to below 5 kHz
    1.2° from 5 kHz to 10 kHz
    In the acceleration range (or equivalent) 0.01 m/s2 up to 10 m/s2 below 10 Hz, up to 50 m/s2 from 10 Hz, up to 100 m/s2 from 40 Hz, up to 1000 m/s2 from 160 Hz, and up to 5000 m/s2 at approximately 650 Hz
    Very low frequency calibration with an acceleration of up to 9.8 m/s2
    with least uncertainties of measurement of -
    0.2% of reading at zero Hertz (DC)
    0.2% of reading from 0.01 Hz to below 5 Hz
    0.3% of reading from 5 Hz to below 10 Hz
    0.4% of reading at 10 Hz
    Calibration of laser vibrometers, including laser Doppler vibrometers (LDV), for complex vibration sensitivity by the methods of ISO 16063-41
    with least calibration and measurement capability of -
    Magnitude:
    0.3% of reading from 0.5 Hz to below 5 kHz
    0.4% of reading from 5 kHz to below 6.3 kHz
    0.5% of reading from 6.3 kHz to 10 kHz
    Phase:
    0.5° from 0.5 Hz to below 5 kHz
    1° from 5 kHz to 10 kHz


    Artificial mastoids;
    Force sensitivity; Phase sensitivity; Velocity sensitivity;
    Comparison with a reference standard; Direct measurement by electrical input;
    IEC 60318-6
    For frequency range 80 Hz to 8 kHz
    CAPABILITY
    with calibration and measurement capability of -
    For mechanical impedance level, and force sensitivity level
    Impedance magnitude:
    0.4 dB re 1 N/(m∙s-1) from 125 Hz to 4 kHz
    0.5 dB re 1 N/(m∙s-1) from above 4 kHz to 8 kHz
    Impedance phase sensitivity:
    0.7º for phase angle at 250 Hz
    Force sensitivity:
    0.4 dB re 1 V/N from 125 Hz to 1.5 kHz
    0.5 dB re 1 V/N from above 1.5 kHz to 8 kHz

    Artificial mastoids; Mechanical impedance transducers;
    Acceleration sensitivity; Force sensitivity; Phase sensitivity;
    Comparison with a reference standard; Direct measurement by electrical input;
    for acceleration range 10 m/sto 100 m/s2
    and forces up to 3 N
    CAPABILITY
    Calibration by interferometry of mechanical impedance transducers, including impedance heads for artificial mastoids, within the acceleration range 10 m/s2 and 100 m/s2 and forces up to 3 N
    with calibration and measurement capability of -
    Acceleration sensitivity:
    0.3% of reading from 80 Hz to below 5 kHz
    0.4% of reading from 5 kHz to below 6.3 kHz
    0.5% of reading from 6.3 kHz to 8 kHz
    Dynamic force sensitivity:
    0.5% of reading from 80 Hz to below 5 kHz
    0.7% of reading from 5 kHz to 8 kHz
    Phase:
    0.5º from 80 Hz to below 5 kHz
    1º from 5 kHz to 8 kHz

    Charge amplifiers; Machine condition monitoring equipment; Vibration analysers; Vibration filters; Vibration measuring systems;
    Electrical response;
    Direct measurement by electrical input;
    for Frequency DC to 20 kHz
    and 10 mV to 10 V
    CAPABILITY
    Calibration by electrical signal input
    with calibration and measurement capability of -
    Magnitude:
    0.1% of reading in the frequency range DC to 20 kHz from 10 mV to 10 V
    Phase:
    0.5° from DC to below 5 kHz
    1° from 5 kHz to 20 kHz
    Vibration metrology - Vibration standards and reference equipment
    Vibration calibrators;
    Acceleration; Frequency;
    Comparison measurement against reference transducer;
    for frequency range 0.5 Hz to 10 kHz
    CAPABILITY
    Calibration of vibration calibrators, shakers and exciters for vibration amplitude and frequency
    with least calibration and measurement capability of -
    Magnitude:
    0.4% of reading from 0.5 Hz to below 5 kHz
    0.5% of reading from 5 kHz to below 6.3 kHz
    0.6% of reading from 6.3 kHz to 10 kHz
    Frequency:
    0.01 Hz from 0.5 Hz to 10 kHz

    Reference accelerometers;
    Acceleration; Acceleration to voltage or charge sensitivity; Frequency;
    Comparison measurement against reference transducer; Direct measurement by interferometer;
    by method ISO 16061-11
    For frequency range 0 Hz to 10 kHz
    and acceleration range 0.01 m/s2 to 1000 m/s2
    CAPABILITY
    Absolute calibration of complex vibration amplitude sensitivity by interferometry from 0.5 Hz to 10 kHz by the methods of ISO 16063-11
    Transducer response in voltage sensitivity, current sensitivity, charge sensitivity or displayed units
    with calibration and measurement capability of -
    Magnitude:
    0.3% of reading from 0.5 Hz to below 9 kHz
    0.5% of reading from 9 kHz to 10 kHz
    Phase:
    0.5° from 0.5 Hz to below 5 kHz
    1° from 5 kHz to 10 kHz
    In the acceleration range (or equivalent) 0.01 m/s2 up to 10 m/s2 below 10 Hz, up to 50 m/s2 from 10 Hz, up to 100 m/s2 from 40 Hz, up to 1000 m/s2 from 160 Hz, and up to 5000 m/s2 at approximately 650 Hz
    Extended lower velocity range down to 1 mm/s from 0.5 Hz to below 2 Hz
    Vibration transducers of mass up to 5000 grams for frequencies of 6 kHz and below, and up to 100 grams for frequencies of 10 kHz and below
    by comparison to the methods of ISO 16063-21
    with calibration and measurement capability of -
    Magnitude:
    0.4% of reading from 0.5 Hz to below 5 kHz
    0.5% of reading from 5 kHz to below 6.3 kHz
    0.6% of reading from 6.3 kHz to 10 kHz
    Phase:
    0.7° from 0.5 Hz to below 630 Hz
    0.8° from 630 Hz to below 2.5 kHz
    0.9° from 2.5 kHz to below 5 kHz
    1.2° from 5 kHz to 10 kHz
    In the acceleration range (or equivalent) 0.01 m/s2 up to 10 m/s2 below 10 Hz, up to 50 m/s2 from 10 Hz, up to 100 m/s2 from 40 Hz, up to 1000 m/s2 from 160 Hz, and up to 5000 m/s2 at approximately 650 Hz
    Very low frequency calibration with an acceleration of up to 9.8 m/s2
    with least uncertainties of measurement of -
    0.2% of reading at zero Hertz (DC)
    0.2% of reading from 0.01 Hz to below 5 Hz
    0.3% of reading from 5 Hz to below 10 Hz
    0.4% of reading at 10 Hz