Back to Search Results


Address Details:
36 Bradfield Road
LINDFIELD WEST, NSW 2070
AUSTRALIA Map
www.measurement.gov.au

Contact Details:
Ms Charmaine Kwan
+61(02) 84673836

Availability:
Services conditionally available to external clients

SCOPE OF ACCREDITATION   View and Print Scope of Accreditation Printable version

Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.



ISO/IEC 17025 (2017)
Calibration
SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
DC and low frequency electrical metrology - Electrical measurement and test equipment
Phase angle meters;
Phase angle;
Comparison with a reference standard;
10 Hz ≤f ≤ 500 kHz
0 s to 105 s
CAPABILITY

With Calibration and Measurement Capability of -
0.02° for 10 Hz ≤f ≤ 100 Hz
0.03° for 100 Hz <f ≤ 1 kHz
0.05° for 1 kHz <f ≤ 5 kHz
0.07° for 5 kHz <f ≤ 10 kHz
0.15° for 10 kHz <f ≤ 500 kHz
Speed and velocity - Speed measuring devices
LIDAR speed calibrators; RADAR speed calibrators; Time interval equipment;
Distance; Speed; Time interval;
Direct measurement against a reference standard;
0 km/h to 300 km/h
10 m to 2 km 
 0 s to 105 s
CAPABILITY
Calibrators for laser-based (LIDAR) speed and distance measuring instruments (LIDAR calibrators)
With Calibration and Measurement Capability of : 
0.2 km/h in the range 0 km/h to 300 km/h, and 
0.02 m in the range 10 m to 2 km 

Time Interval with least uncertainties of measurement of -
√[(200 ps)2 + ((1.3x10-11s½)√T)2], from 0 s to 105 s

(The uncertainty of measurement does not consider the DUT 
and follows the CCTF guidelines)
Time and frequency metrology - Frequency and time standards
Caesium standard; Frequency standards; GPS disciplined oscillators; Rubidium standard;
Coordinated universal time (UTC);
Measurement against reference standard;
CAPABILITY

Frequency standards including Caesium Standards, Rubidium Standards, and GPS Disciplined Oscillators
Measurement of Coordinated Universal Time (UTC)
with Calibration and Measurement Capability of -
50 ns

The uncertainty of measurement does not consider the DUT 
and follows the CCTF guidelines

Measurement capability includes both laboratory based calibrations 
and those performed via remote monitoring




Frequency;
Measurement against reference standard;
CAPABILITY

General frequency sources, including Caesium Standards, Rubidium Standards, and GPS Disciplined Oscillators
with Calibration and Measurement Capability of -
2 in 1013 from 1 mHz to 27.5 GHz
(a minimum slew rate of 1V/μs is assumed at frequencies below 100 kHz

(The uncertainty of measurement does not consider the DUT 
and follows the CCTF guidelines)

Measurement capability involves both laboratory based calibrations 
and those performed via remote monitoring


Time interval;
Measurement against reference standard;
 
CAPABILITY
Frequency standards, including Caesium Standards, Rubidium Standards and GPS Disciplined Oscillators
Measurement of Time Interval with least uncertainties of measurement of -
√[(200 ps)2 + ((1.3x10-11s½)√T)2], from 0 s to 105 s

(The uncertainty of measurement does not consider the DUT 
and follows the CCTF guidelines)


Frequency;
Measurement against reference standard;
In-house (PM-TFR-T10)
CAPABILITY
Standard frequency sources including Caesium Standards, Rubidium Standards and GPS Disciplined Oscillators
Measurement of Frequency
with Calibration and Measurement Capability of -
2 in 1013 at 1 Hz, 5 MHz and 10 MHz
(a minimum slew rate of 1V/μs is assumed at frequencies below 100 kHz

(The uncertainty of measurement does not consider the DUT
and follows the CCTF guidelines)

Measurement capability involves both laboratory based calibrations
and those performed via remote monitoring

Laser length standards (interferometers);
Frequency;
Measurement against reference standard;
Measurement of optical frequency / wavelength of stabilised lasers used in length interferometry in accordance with Test Method PM-TFR-T08
182 THz to  565 THz
(vacuum wavelength 1650 nm to 530 nm)
0 s to 105 s
CAPABILITY
10 kHz from 182 THz to 565 THz
(vacuum wavelength 1650 nm to 530 nm)
Time and frequency metrology - Frequency, time and waveform measuring equipment
Clocks and timers; Counters; Frequency analysers; Frequency meters; Time interval meters;
Frequency; Time interval;
Measurement against reference standard;
CAPABILITY
The uncertainty of measurement does not consider the DUT 
and follows the CCTF guidelines

Frequency characteristics
Frequency with Calibration and Measurement Capability of -
2 in 1013 from 1 mHz to 27.5 GHz
(a minimum slew rate of 1V/μs is assumed at frequencies below 100 kHz

Time characteristics
Time Interval with Calibration and Measurement Capability of -
√[(200 ps)2 + ((1.3x10-11s½)√T)2], from 0 s to 105 s