| SERVICE | PRODUCT | DETERMINANT | TECHNIQUE | PROCEDURE | LIMITATION / RANGE |
Temperature metrology - Ancillary temperature measuring equipment
| Resistance bridges;
| Resistance to temperature conversion;
| Direct measurement by electrical input; | | from 0 Ω to 300 Ω
|
|
CAPABILITY Calibration of AC bridges for temperature measurement with least uncertainties of measurement of - 1 mΩ for linearity from 0 Ω to 300 Ω 0.1 μΩ/Ω for linearity from 16 Ω to 130 Ω
|
| Type R extension wire; Type S extension wire;
| Volt to temperature conversion;
| Direct measurement by electrical input; | | from 0 °C to 50 °C
|
|
CAPABILITY 0.22 μV
|
| Digital voltmeters (DVM) and digital multimeters (DMM); Multi-channel thermocouple data recorders; Process calibrators;
| Volt to temperature conversion;
| Direct measurement by electrical input; | | 0 °C to 1100 °C |
|
CAPABILITY Calibration of thermocouple indicators (sumilation only) with least uncertainties of measurement of - Type T 0.09 °C from 0 °C to 500 °C Type K, J, and N 0.09 °C from 0 °C to 1100 °C Type S and R 0.12 °C from 0 °C to 1100 °C |
Temperature metrology - Humidity measuring equipment
| Dew point meters; Hygrometers; Psychrometers;
| Dew point; Temperature;
| To be determined; | | From -30 °C to 80 °C
|
|
CAPABILITY Calibration of dew point meters with least uncertainties of measurement of - 0.1 K for range of dew point from −30 °C to 80 °C Calibration of temperature sensors in dew point meters with least uncertainties of measurement of - from 0.28 K at −20 °C to 0.06 K at 14 °C in air 0.06 K from 14 °C to 32 °C in air from 0.06 K at 32 °C to 0.37 K at 80 °C in air
|
| Relative humidity sensors;
| Relative humidity (RH); Temperature;
| To be determined; | | from 5% to 95% RH From 0 °C to 80 °C
|
|
CAPABILITY Calibration of relative humidity sensors with least uncertainties of measurement of -
| Relative Humidity (RH)
| 5%
| 50%
| 95%
| 0 °C
| 0.20
| 0.87
| 1.39
| 23 °C
| 0.19
| 0.64
| 1.10
| 70 °C
| 0.18
| 0.50
| 0.83
| 80 °C
| 0.18
| 0.47
| 0.79
|
with a minimum frost point of −30 °C Calibration of temperature sensors for RH sensors and electronic psychrometers with least uncertainties of measurement of - from 0.28 K at −20 °C to 0.06 K at 14 °C in air 0.06 K from 14 °C to 32 °C in air from 0.06 K at 32 °C to 0.37 K at 80 °C in air
|
Temperature metrology - Non-contact temperature measuring equipment
| Thermal imaging systems;
| Temperature;
| To be determined; | | from -40 °C to 1100 °C
|
|
CAPABILITY Calibration of thermal imaging cameras with least uncertainties of measurement of - from 0.1 °C at −40 °C to 0.5 °C at 250 °C from 0.5 °C at 250 °C to 2 °C at 1100 °C
|
| Clinical infrared thermometers; Radiation pyrometers;
| Temperature;
| To be determined; | | from -40 °C to 2900 °C
|
|
CAPABILITY Radiation pyrometers from 0.1 °C at −40 °C to 0.3 °C at 250 °C from 0.5 °C at 250 °C to 2 °C at 1100 °C from 2 °C at 1100 °C to 4 °C at 2900 °C
Calibration of ear thermometers in the range 25 °C to 45 °C with a least uncertainty of measurement of - 0.1 °C
|
Temperature metrology - Temperature measuring equipment
| Liquid in glass (LIG) thermometers;
| Temperature;
| Measurement against reference standard; | | from −200 °C to 300 °C
|
|
CAPABILITY 0.3 °C from −200 °C to −50 °C 25 mK from −50 °C to −10 °C 5 mK from −10 °C to 50 °C 3 mK at 0 °C 15 mK from 50 °C to 100 °C 40 mK from 100 °C to 250 °C 0.15 °C from 250 °C to 300 °C
|
| Base metal thermocouples; Surface probes;
| mV over oC range;
| Direct temperature and electrical measurement against reference standard; | | from −200 °C to 1100 °C
|
|
CAPABILITY Calibration of Base Metal Thermocouples with least uncertainties of measurement of - Type T 1.0 μV from −200 °C to 0 °C (1.0 + 0.01% E) μV from 0 °C to 200 °C Type K (1.0 + 0.036% E) μV from −200 °C to 200 °C (1.0 + 0.055% E) μV from 200 °C to 400 °C (4.14 + 0.06% E) μV from 400 °C to 1100 °C where E is the thermocouple voltage
Calibration of surface temperature sensors with least uncertainties of measurement of - from 0.21 °C at 20 °C to 0.70 °C at 350 °C
|
| Digital temperature measuring systems;
| Temperature;
| Measurement against reference standard; | | from -200 °C to 1100 °C
|
|
CAPABILITY Thermocouple Measuring Systems Type T 1.0 μV from −200 °C to 0 °C (1.0 + 0.01% E) μV from 0 °C to 200 °C Type K (1.0 + 0.036% E) μV from −200 °C to 200 °C (1.0 + 0.055% E) μV from 200 °C to 400 °C (4.14 + 0.06% E) μV from 400 °C to 1100 °C where E is the thermocouple voltage
Calibration of surface temperature measuring system with least uncertainties of measurement of - from 0.21 °C at 20 °C to 0.70 °C at 350 °C
Metallic resistance Measuring Systems Calibration of standard PRTs using temperature fixed points with least uncertainties of measurement of - from 1 mK at −189.34 °C to 0.5 mK at −38.83 °C 0.5 mK from −38.83 °C to 419.53 °C from 0.5 mK at 419.53 °C to 1.5 mK at 660.32 °C from 1.5 mK at 660.32 °C to 13 mK at 961.78 °C
Calibration of standard PRTs and industrial PRTs by comparison against standard PRTs with least uncertainties of measurement of - 5 mK at −200 °C 5 mK at −80 °C 3 mK at 0 °C 3 mK at 250 °C 6 mK at 550 °C 13 mK at 960 °C |
Temperature metrology - Temperature standards and reference equipment
| Platinum-gold (Pt-Au) thermocouples; Rare metal thermocouples;
| mV per oC;
| To be determined; | | from 0 °C to 1560 °C
|
|
CAPABILITY Calibration of Rare Metal Thermocouple with least uncertainties of measurement of - (0.31 + 0.011% E) μV from 0 °C to 1100 °C (0.31 + 0.014% E) μV from 0 °C to 1560 °C (0.008% E) μV at 200 °C for homogeneity where E is the thermocouple voltage
Calibration of Pt-Au thermocouples with least uncertainties of measurement of - 19 mK from 0 °C to 50 °C 14 mK from 50 °C to 420 °C 26 mK from 420 °C to 960 °C
|
| Platinum resistance thermometers (PRTs) reference standards;
| Temperature;
| To be determined; | | from −200 °C to 961.78 °C
|
|
CAPABILITY Calibration of standard PRTs using temperature fixed points with least uncertainties of measurement of - from 1 mK at −189.34 °C to 0.5 mK at −38.83 °C 0.5 mK from −38.83 °C to 419.53 °C from 0.5 mK at 419.53 °C to 1.5 mK at 660.32 °C from 1.5 mK at 660.32 °C to 13 mK at 961.78 °C
Calibration of standard PRTs and industrial PRTs by comparison against standard PRTs with least uncertainties of measurement of - 5 mK at −200 °C 5 mK at −80 °C 3 mK at 0 °C 3 mK at 250 °C 6 mK at 550 °C 13 mK at 960 °C
|
| Black body sources;
| Temperature;
| To be determined; | | from −40 °C to 2000 °C
|
|
CAPABILITY from 0.4 °C at −40 °C to 0.3 °C at 50 °C from 0.3 °C at 50 °C to 1.4 °C at 1000 °C from 1.4 °C at 1000 °C to 2.9 °C at 2000 °C
Calibration of ear thermometer blackbodies in the range 25 °C to 45 °C with a least uncertainty of measurement of - 0.025 °C
|
| Dry block calibrators;
| Temperature;
| To be determined; | | from −50 °C to 650 °C
|
|
CAPABILITY from 0.15 °C at −50 °C to 0.05 °C at 50 °C from 0.05 °C at 50 °C to 0.15 °C at 650 °C
|
| Temperature reference standards - Fixed points;
| Silver (Ag) freezing point ; Aluminum (Al) freezing point; Argon (Ar) triple point; Cadmium freezing point; Gallium melting point; Indium (In) freezing point; Mercury (Hg) triple point; Tin (Sn) freezing point; Water triple point; Zinc freezing point;
| Direct measurement against a reference standard; Direct measurement against argon triple point; Direct measurement against H2O triple point; Direct measurement against mercury triple point; | | |
|
CAPABILITY 1 mK at Ar triple point 0.3 mK at Hg triple point 0.1 mK at water triple point 0.3 mK at Ga melting point 0.5 mK at In freezing point 0.3 mK at Sn freezing point 0.4 mK at Cd freezing point 0.5 mK at Zn freezing point 1.5 mK at Al freezing point 10 mK at Ag freezing point
|