SCOPE OF ACCREDITATION   View and Print Scope of Accreditation Printable version

Note: Not all of the columns of the scope of accreditation displayed include data.
The only data displayed is that deemed relevant and necessary for the clear description of the activities and services covered by the scope of accreditation.
Grey text appearing in a SoA is additional freetext providing further refinement or information on the data in the preceding line entry.



ISO/IEC 17025 (2017)
Calibration
SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
Optical metrology - Optical measuring equipment
Laser energy meters; Laser power meters;
Energy; Power;
Comparison with a reference meter; Direct determination by optical methods;
CAPABILITY

Optical Pulse Energy;
Optical Power;
With Calibration and Measurement Capability of -
0.02% at selected visible and near infrared laser lines measured using cryogenic radiometry;
0.36% at power levels from 1.0 μW to 1.0 mW;
0.7% at power levels above 1.0 mW up to 100 mW;
0.9% at power levels above 100 mW up to 1.0 W;
1.1% at power levels above 1.0 W up to 10 W;
1.3% at power levels above 10 W up to 20 W;
using laser line sources in the range 360 nm to 1 650 nm;



Filter radiometer;
Linearity; Responsivity;
Direct comparison against a reference standard; Direct measurement by superposition method;
CAPABILITY
Broadband responsivity;
0.2% for irradiances in the range 100 μ W/cm2 to 3 W/cm2;
for UV/visible/IR radiation

Spectral responsivity;
7.0% at 220 nm;
reducing linearly to 3.0% at 230 nm;
reducing linearly to 0.9% at 240 nm;
reducing linearly to 0.27% at 360 nm;
reducing linearly to 0.12% at 650 nm;
increasing linearly to 0.40% at 950 nm;
0.40% in the range 950 to 1 655 nm;
increasing linearly to 3.0% at 1 685 nm;
increasing linearly to 3.2% at 1 700 nm;
for incoherent sources;

Response linearity
0.01% for a 5:1 range of response;

Photodiodes; Radiometers;
Linearity; Responsivity;
Direct comparison against a reference standard; Direct determination by optical methods; Direct measurement by superposition method;
CAPABILITY
Spectral responsivity
With Calibration and Measurement Capability of -
7.0% at 220 nm;
reducing linearly to 3.0% at 230 nm;
reducing linearly to 0.9% at 240 nm;
reducing linearly to 0.27% at 360 nm;
reducing linearly to 0.12% at 650 nm;
increasing linearly to 0.40% at 950 nm;
0.40% in the range 950 to 1 655 nm;
increasing linearly to 3.0% at 1685 nm;
increasing linearly to 3.2% at 1 700 nm;
for incoherent sources;

Interferometers; Spectrometer arrays; Spectroradiometers; Spectrum analysers;
Wavelength;
Comparison with a reference meter; Direct measurement using a reference meter;
CAPABILITY
Measurement of wavelength of -
Optical spectrum analysers,
Interferometers,
Spectroradiometers,
Array spectrometers;
With Calibration and Measurement Capability of -
100 pm in the range 240 to 1700 nm;
200 pm in the range 1700 to 2500 nm;

Luminance meters;
Linearity; Luminance;
Comparison with a reference meter;
CAPABILITY

Luminance meters; Chromameters and telespectroradiometers;
With Calibration and Measurement Capability of -
6% at 1 cd/m²;
reducing to 1.8% at 10 cd/m²;
remaining at 1.8% at 100 cd/m²;
reducing to 0.55% at 1 000 cd/m²;
remaining at 0.55% at 10 000 cd/m²;
increasing to 1.0% at 30 000 cd/m²;

Bilirubin meters; Blue light radiometers; Ultraviolet radiometers; UVA meters; UVB meters; UVC meters;
Cosine correction; Irradiance; Linearity;
Direct comparison against a reference standard; Direct determination by optical methods;
CAPABILITY
With Calibration and Measurement Capability of -
Irradiance
5.6% for irradiance of 11 mW/cm²;
falling to 5.2% for irradiance of 1.4 mW/cm²;
remaining at 5.2% for irradiance of 100 μW/cm²;
for monochromatic sources near 254 nm;

5.2% for irradiance of 1 mW/cm²;
falling to 3.4% for irradiance of 25 μW/cm²;
rising to 4.2% for irradiance of 5 μW/cm²;
for monochromatic sources near 365 nm;

For significantly different source spectra, an additional uncertainty component which depends on the spectrum and may be in the order of 2% (k=1) is added to the uncertainty analysis;
5% for bilirubin treatment sources (blue light);

Linearity
0.01% for a 5:1 range of response;

Cosine correction
1% for field angles greater than 5 degrees;



Illuminance meters; Light meters; Luxmeters; Photometers; Photometric detectors;
Cosine correction; Illuminance; Linearity; Responsivity;
Direct comparison against a reference standard; Direct measurement by superposition method; Direct measurement using a reference standard;
CAPABILITY
With Calibration and Measurement Capability of -
Illuminance
0.5% at 1 mlx;
reducing to 0.39% at 1 lx;
0.39% from 1 lx to 50 lx;
Increasing to 8% at 200 klx;

Spectral responsivity;
7.0% at 220 nm;
reducing linearly to 3.0% at 230 nm;
reducing linearly to 0.9% at 240 nm;
reducing linearly to 0.27% at 360 nm;
reducing linearly to 0.12% at 650 nm;
increasing linearly to 0.40% at 950 nm;
0.40% in the range 950 to 1 655 nm;
increasing linearly to 3.0% at 1 685 nm;
increasing linearly to 3.2% at 1 700 nm;

Linearity
0.01% for a 5:1 range of response;

Cosine correction
0.1% for field angles greater than 20 degrees;

Infra-red radiometers; Thermal detectors; Thermopiles;
Irradiance;
Direct comparison against a reference standard;
Measurement of response to a black-body source at a radiance temperature of 800°C;

CAPABILITY
Measurement of response to a black-body source at a radiance temperature of 800°C;
With Calibration and Measurement Capability of -
0.2% irradiance level near 370 μW/cm2;

Optical metrology - Optical standards and reference equipment
Integrating sphere sources; Luminance standard sources;
Chromaticity; Correlated colour temperature; Distribution temperature; Illuminance; Luminance; Spectral radiance;
Direct comparison against a reference standard; Direct measurement using a reference meter; Measurement using a reference source and reference meter;
CAPABILITY
With Calibration and Measurement Capability of -
Illuminance:
0.39% from 1 mlx to 50 lx;
Increasing to 0.6% at 12 klx

Luminance:
6% at 1 cd/m²;
reducing to 1.8% at 10 cd/m²;
remaining at 1.8% at 100 cd/m²;
reducing to 0.55% at 1 000 cd/m²;
remaining at 0.55% at 10 000 cd/m²;
increasing to 1.0% at 30 000 cd/m²;

Spectral radiance:
10% at 240 nm;
reducing linearly to 6% at 245 nm;
reducing linearly to 3.4% at 250 nm;
reducing linearly to 2.0% at 270 nm;
reducing linearly to 1.7% at 290 nm;
reducing linearly to 1.0% at 390 nm;
reducing linearly to 0.8% at 555 nm;
increasing linearly to 3.7% at 2 500 nm;

Chromaticity:
0.00022 for CIE 1931 x, y coordinates for incandescent sources
0.00046 for CIE 1931 x, y coordinates for other sources

Distribution temperature:
(2.602 x 10⁻⁶ (DT)² - 1.625 x 10⁻³ (DT) + 1.950) K in the range 1 600 K to 3 400 K;

Correlated colour temperature
6 K for temperatures near 2 856 K;
3 to 50 K in the range 1700 to 11,000 K;



Luminous flux standard sources;
Luminous flux;
Direct measurement by goniophotometer;
CAPABILITY

Luminous flux:
With Calibration and Measurement Capability of -

Incandescent sources;
0.6% for values near 1 000 lm and 2 500 lm;

Other sources:
0.7% ;


LED sources; Solid state lighting (SSL) sources;
Chromaticity; Correlated colour temperature; Distribution temperature; Illuminance; Luminance; Luminous flux; Luminous intensity; Spectral irradiance; Spectral radiance;
Direct comparison against a reference standard; Direct measurement by goniophotometer; Direct measurement using a reference meter; Measurement using a reference source and reference meter;
CAPABILITY

Luminous flux
0.7%

Luminous intensity
0.34% for broad-band sources

Illuminance
0.39% from 1 mlx to 50 lx;
Increasing to 0.6% at 12 klx;

Luminance:
6% at 1 cd/m²;
reducing to 1.8% at 10 cd/m²;
remaining at 1.8% at 100 cd/m²;
reducing to 0.55% at 1 000 cd/m²;
remaining at 0.55% at 10 000 cd/m²;
increasing to 1.0% at 30 000 cd/m²;

Spectral radiance:
10% at 240 nm;
reducing linearly to 6% at 245 nm;
reducing linearly to 3.4% at 250 nm;
reducing linearly to 2.0% at 270 nm;
reducing linearly to 1.7% at 290 nm;
reducing linearly to 1.0% at 390 nm;
reducing linearly to 0.8% at 555 nm;
increasing linearly to 3.7% at 2 500 nm;

Spectral irradiance:
4.6% at 250 nm;
reducing linearly to 1.8% at 300 nm;
reducing linearly to 1.0% at 450 nm;
reducing linearly to 0.9% at 500 nm;
remaining at 0.9% at 600 nm;
increasing linearly to 2.3% at 1 600 nm;
increasing linearly to 3.1% at 1 800 nm;
increasing linearly to 4.2% at 2 500 nm;

Chromaticity
0.00046 for CIE 1931 x, y coordinates;

Distribution temperature:
(2.602 x 10⁻⁶ (DT)² - 1.625 x 10⁻³ (DT) + 1.950) K in the range 1 600 K to 3 400 K;

Correlated colour temperature
6 K for temperatures near 2 856 K;
3 to 50 K in the range 1700 to 11,000 K;



Broadband sources;
Broadband irradiance;
Direct measurement using a reference meter;
CAPABILITY
Broadband irradiance from Planckian Radiators and Blackbody Sources;
With Calibration and Measurement Capability of -
0.2% at irradiance levels near 370 μW/cm²

Incandescent lamps;
Chromaticity; Correlated colour temperature; Distribution temperature; Illuminance; Luminous flux; Luminous intensity; Spectral irradiance;
Direct comparison against a reference standard; Direct measurement by goniophotometer; Direct measurement using a reference meter; Measurement using a reference source and reference meter;
CAPABILITY
With Calibration and Measurement Capability of -
Luminous Flux
0.6% for values near 1 000 lm and 2 500 lm;
Illuminance
0.39% from 1 mlx to 50 lx;
Increasing to 0.6% at 12 klx;
Luminous Intensity
0.34% near 250 cd;
Spectral irradiance
4.6% at 250 nm;
reducing linearly to 1.8% at 300 nm;
reducing linearly to 1.0% at 450 nm;
reducing linearly to 0.9% at 500 nm;
remaining at 0.9% at 600 nm;
increasing linearly to 2.3% at 1 600 nm;
increasing linearly to 3.1% at 1 800 nm;
increasing linearly to 4.2% at 2 500 nm;
Chromaticity
0.00022 for CIE 1931 x, y coordinates
Distribution temperature
(2.602 x 10⁻⁶ (DT)² - 1.625 x 10⁻³ (DT) + 1.950) K in the range 1 600 K to 3 400 K;
Correlated Colour Temperature
6 K for temperatures near 2 856 K;
3 to 50 K in the range 1700 to 11,000 K;

Luminous intensity standard sources;
Luminous intensity;
Measurement using a reference source and reference meter;
CAPABILITY

Luminous intensity
With Calibration and Measurement Capability of -

Incandescent lamps and luminous intensity standard sources;
0.34% near 250 cd;

Other sources
0.34% for broad-band sources


Laser line sources; Laser sources;
Beam divergence; Pulse duration; Pulse energy; Radiant power - Laser line sources; Wavelength;
Direct comparison against a reference standard; Direct determination by optical methods; Direct measurement using a reference meter;
CAPABILITY
Including on site testing

Radiant power
With Calibration and Measurement Capability of -
0.36% at power levels from 1.0 μW to 1.0 mW;
0.7% at power levels above 1.0 mW up to 100 mW;
in the range 360 nm to 1 650 nm;
0.36% at power levels from 1.0 μW to 1.0 mW;
0.7% at power levels above 1.0 mW up to 100 mW;
0.9% at power levels above 100 mW up to 1.0 W;
1.1% at power levels above 1.0 W up to 10 W;
1.3% at power levels above 10 W up to 20 W;
for selected laser lines;
including on-site testing;

Pulse energy
With Calibration and Measurement Capability of -
0.36% for pulses with peak power levels from 1.0 μW to 1.0 mW;
0.7% for pulses with peak power levels above 1.0 mW up to 100 mW;
0.9% for pulses with peak power levels above 100 mW up to 1.0 W;
1.1% for pulses with peak power levels above 1.0 W up to 10 W;
1.3% for pulses with peak power levels above 10 W up to 20 W;

Wavelength
5 pm in the range 395 to 1 095 nm (vacuum wavelength) using a wavemeter;
100 pm in the range 600 to 1 600 nm using an optical spectrum analyser;

Pulse Duration
0.2 ns for pulses longer than 10 ns

Beam divergence
6.1% for divergence angles of 0.2 mrad;
reducing linearly to 1.5% for divergence angles of 0.6 mrad;
reducing linearly to 0.7% at 200 mrad;

Illuminance standard sources;
Chromaticity; Correlated colour temperature; Distribution temperature; Illuminance;
Direct comparison against a reference standard; Measurement using a reference source and reference meter;
CAPABILITY
With Calibration and Measurement Capability of -
Illuminance:
0.39% from 1 mlx to 50 lx;
Increasing to 0.6% at 12 klx;

Chromaticity:
0.00022 for CIE 1931 x, y coordinates for incandescent sources
0.00046 for CIE 1931 x, y coordinates for other sources

Distribution temperature:
(2.602 x 10⁻⁶ (DT)² - 1.625 x 10⁻³ (DT) + 1.950) K in the range 1 600 K to 3 400 K;

Correlated colour temperature
6 K for temperatures near 2 856 K;
3 to 50 K in the range 1700 to 11,000 K;

Discharge sources - Lamps;
Chromaticity; Correlated colour temperature; Spectral irradiance;
Direct measurement using a reference meter; Measurement using a reference source and reference meter;
CAPABILITY

With Calibration and Measurement Capability of -
Spectral irradiance
4.6% at 250 nm;
reducing linearly to 1.8% at 300 nm;
reducing linearly to 1.0% at 450 nm;
reducing linearly to 0.9% at 500 nm;
remaining at 0.9% at 600 nm;
increasing linearly to 2.3% at 1 600 nm;
increasing linearly to 3.1% at 1 800 nm;
increasing linearly to 4.2% at 2 500 nm;

Chromaticity
0.00046 for CIE 1931 x, y coordinates;

Correlated colour temperature
3 to 50 K in the range 1700 to 11,000 K;

Spectral irradiance standard sources;
Chromaticity; Correlated colour temperature; Distribution temperature; Spectral irradiance;
Measurement using a reference source and reference meter;
CAPABILITY
With Calibration and Measurement Capability of -
Spectral irradiance:
4.6% at 250 nm;
reducing linearly to 1.8% at 300 nm;
reducing linearly to 1.0% at 450 nm;
reducing linearly to 0.9% at 500 nm;
remaining at 0.9% at 600 nm;
increasing linearly to 2.3% at 1 600 nm;
increasing linearly to 3.1% at 1 800 nm;
increasing linearly to 4.2% at 2 500 nm;

Chromaticity:
0.00022 for CIE 1931 x, y coordinates for incandescent sources
0.00046 for CIE 1931 x, y coordinates for other sources

Distribution temperature:
(2.602 x 10⁻⁶ (DT)² - 1.625 x 10⁻³ (DT) + 1.950) K in the range 1 600 K to 3 400 K;

Correlated colour temperature
6 K for temperatures near 2 856 K;
3 to 50 K in the range 1700 to 11,000 K;

UVA sources; UVB sources; UVC sources;
Broadband irradiance; Spectral irradiance;
Direct measurement using a reference meter;
CAPABILITY
Broadband irradiance from UVA Sources, UVB Sources, UVC Sources, Blue Light Sources, Solar Simulators, Filtered Sources;
5.6% for irradiance of 11 mW/cm²;
falling to 5.2% for irradiance of 1.4 mW/cm²;
remaining at 5.2% for irradiance of 100 μW/cm²;
for monochromatic sources near 254 nm;

5.2% for irradiance of 1 mW/cm²;
falling to 3.4% for irradiance of 25 μW/cm²;
rising to 4.2% for irradiance of 5 μW/cm²;
for monochromatic sources near 365 nm;

For significantly different source spectra, an additional uncertainty component which depends on the spectrum and may be in the order of 2% (k=1) is added to the uncertainty analysis;

Spectral irradiance;
4.6% at 250 nm;
reducing linearly to 1.8% at 300 nm;
reducing linearly to 1.0% at 450 nm;

Spectral radiance standard sources;
Chromaticity; Correlated colour temperature; Distribution temperature; Spectral irradiance; Spectral radiance;
Direct measurement using a reference meter; Measurement using a reference source and reference meter;
CAPABILITY
With Calibration and Measurement Capability of -
Spectral radiance:
10% at 240 nm;
reducing linearly to 6% at 245 nm;
reducing linearly to 3.4% at 250 nm;
reducing linearly to 2.0% at 270 nm;
reducing linearly to 1.7% at 290 nm;
reducing linearly to 1.0% at 390 nm;
reducing linearly to 0.8% at 555 nm;
increasing linearly to 3.7% at 2 500 nm;

Spectral irradiance:
4.6% at 250 nm;
reducing linearly to 1.8% at 300 nm;
reducing linearly to 1.0% at 450 nm;
reducing linearly to 0.9% at 500 nm;
remaining at 0.9% at 600 nm;
increasing linearly to 2.3% at 1 600 nm;
increasing linearly to 3.1% at 1 800 nm;
increasing linearly to 4.2% at 2 500 nm;

Chromaticity:
0.00022 for CIE 1931 x, y coordinates for incandescent sources
0.00046 for CIE 1931 x, y coordinates for other sources

Distribution temperature:
(2.602 x 10⁻⁶ (DT)² - 1.625 x 10⁻³ (DT) + 1.950) K in the range 1 600 K to 3 400 K;

Correlated colour temperature
6 K for temperatures near 2 856 K;
3 to 50 K in the range 1700 to 11,000 K;

ISO/IEC 17025 (2017)
Manufactured Goods
SERVICEPRODUCTDETERMINANTTECHNIQUEPROCEDURELIMITATION / RANGE
Evaluation of consumer and industrial lighting, optical systems and eyewear
Laser and high radiance sources;
Beam divergence; Energy; Power; Pulse duration; Wavelength;
Measurement using a reference meter;
Measurements of Power, Pulse Energy, Wavelength, Pulse Duration and Beam Divergence of Laser and high radiance sources
Including on site testing
by the in-house methods of -
Test Method TM-36
Test Method TM-37
Test Method TM-41
Including on site testing;